Omron HBP-1300 specifications Setup, For cleaning, see

Page 11

Note:

Setup

Read and understand the manual for each optional accessory. This manual does not contain cautionary information for optional accessory.

Exercise caution with the cables and arrange so that the patient does not become entangled or bound.

Before use / during use

• Check the following after turning on the power:

-No smoke, abnormal odor, or abnormal sound is emitted.

-Press each button and verify that it operates.

-For functions that cause icons to light or flash, verify that the icons light or flash.

-Measurement can be performed normally, and measurement error is within the tolerance value.

• If the screen is not displayed normally, do not use the unit.

• When recycling or disposing of parts (including batteries) of the unit, follow local government rules and regulations.

Cleaning

For cleaning, see page 24.

Optional rechargeable battery

To prevent accidents, keep batteries out of reach of infants and small children.

If you sense that something is wrong with a battery, immediately move it to a safe location and contact the administrator responsible for the unit or OMRON HEALTHCARE Customer Service.

If the battery voltage is low, operation by battery may not be possible.

Non-Invasive Blood Pressure (NIBP) measurement

If the patient has acute inflammation, a pyogenic ailment, or an external wound at the location where the cuff is to be wrapped, follow the instructions of the doctor.

Non-Invasive Blood Pressure measurement (NIBP) is performed by compressing the upper arm. Some people may experience intense pain, or transient spotting caused by subcutaneous hemorrhaging (bruising) may appear. The spotting will disappear with time; however, it may be appropriate to inform patients for whom this may be a concern that spotting sometimes occurs, and if necessary, refrain from measurement.

To measure correctly, it is recommended that the patient relax and not talk during measurement.

To measure correctly, it is recommended that the patient rest quietly for 5 minutes before measurement.

Introduction

9

Image 11
Contents HBP-1300 Page Exemptions IntroductionIntended Use General Information Usage warnings and cautions SetupBefore use / during use Page Maintenance and inspection CleaningOptional rechargeable battery Non-Invasive Blood Pressure Nibp measurement For cleaning, see Components of the Product Using the UnitOptional Accessories Main unit Standard accessoriesCuff Hose 1.5m Features of the ProductKey features, benefits, look Cuff Hose 3.5 mFront and back of unit Features / Functions of UnitMeaning of the Symbols LCD Display Using the UnitOnly when the optional battery is installed Auto Power Off Installing the Optional Battery PackBattery life Low batteryAC power Connecting the AC AdapterBattery charging Battery levelPress the or button to change the setting System SettingsMenu Mode Press the Mode buttonExample Utility ModePress the or button to change the auto power off setting Check the Pressure AccuracyNon-Invasive Pressure Measurement Principles Non-Invasive Blood Pressure Nibp MeasurementSelecting the cuff Connecting the cuffAvailable as an optional accessory Cuff Selection and ConnectionAlign the artery mark Index Artery with the brachial artery Failure Measurement results are displayedBlood pressure measurement is performed once Use a stethoscope to perform measurement Taking the Measurement in Auscultation ModeIrregular pulse wave detection function Body movement detection functionIrregular pulse wave icon Appears Stopping the MeasurementDisplaying Last Reading Accessory Care MaintenanceMaintenance Inspection and Safety Management Device MaintenanceCheck before Use Before turning on the powerWhen turning on the power After turning on the powerTroubleshooting Measurement value is questionable StethoscopeAbnormal measurement value Error Priority Description Points to check List of Error CodesDescription Other problemsError Priority Cause Solution Code DisposalCall toll free Limited WarrantyBackup SpecificationsFactory Default Settings SpecificationsNon-Invasive Blood Pressure Nibp Power supplyEnvironmental Conditions Emission Test Compliance Electromagnetic Environment Manufacturers DeclarationElectromagnetic Immunity IEC60601-1-2 Recommend separation distance IEC60601-1-2 Electromagnetic Test levelPage Separation distance according to frequency of transmitter Recommended Separation DistancesPage West Field Court, Lake Forest, IL 60045, USA Made in China For Customer Service Call Toll FreeDistributed by