Non-Invasive Blood Pressure (NIBP) Measurement

Non-Invasive Pressure Measurement Principles

Oscillometric method

The beat in the pulsation generated by the contraction of the heart is captured as the pressure inside the cuff to measure the blood pressure. If the cuff wrapped around the upper arm is pressurized sufficiently, the blood flow stops, but the beat of the pulsation is present and the pressure inside the cuff receives this and oscillates. Next, as the pressure inside the cuff gradually decreases, the oscillation of the pressure within the cuff gradually increases and reaches a peak. As the pressure within the cuff decreases further, the oscillation decreases from its peak.

The pressure within the cuff and the relationship with the increase and decrease of the oscillation within the cuff in this series of processes are stored into memory, calculations are carried out, and the blood pressure value is determined.

The pressure within the cuff when the oscillation increases drastically is the systolic pressure and the pressure within the cuff when the oscillation decreases drastically is the diastolic pressure. Also, the pressure within the cuff when the oscillation peaks is taken as the average pulsation pressure.

The oscillometric method does not determine the blood pressure value instantly like a microphone type automatic blood pressure gauge with the auscultation method, but rather determines it from the series of change curves as explained above. Therefore, it is not easily affected by external noise, an electric scalpel or other electro surgical instruments.

KOROTKOV SOUNDS

RADIAL PULSE

18

Image 20
Contents HBP-1300 Page Exemptions IntroductionIntended Use General Information Setup Usage warnings and cautionsBefore use / during use Page Cleaning Maintenance and inspectionOptional rechargeable battery Non-Invasive Blood Pressure Nibp measurement For cleaning, see Using the Unit Optional AccessoriesMain unit Standard accessories Components of the ProductFeatures of the Product Key features, benefits, lookCuff Hose 3.5 m Cuff Hose 1.5mFront and back of unit Features / Functions of UnitMeaning of the Symbols LCD Display Using the UnitOnly when the optional battery is installed Installing the Optional Battery Pack Battery lifeLow battery Auto Power OffConnecting the AC Adapter Battery chargingBattery level AC powerSystem Settings Menu ModePress the Mode button Press the or button to change the settingUtility Mode Press the or button to change the auto power off settingCheck the Pressure Accuracy ExampleNon-Invasive Blood Pressure Nibp Measurement Non-Invasive Pressure Measurement PrinciplesConnecting the cuff Available as an optional accessoryCuff Selection and Connection Selecting the cuffAlign the artery mark Index Artery with the brachial artery Failure Measurement results are displayedBlood pressure measurement is performed once Taking the Measurement in Auscultation Mode Irregular pulse wave detection functionBody movement detection function Use a stethoscope to perform measurementIrregular pulse wave icon Appears Stopping the MeasurementDisplaying Last Reading Maintenance Maintenance Inspection and Safety ManagementDevice Maintenance Accessory CareBefore turning on the power When turning on the powerAfter turning on the power Check before UseTroubleshooting Measurement value is questionable StethoscopeAbnormal measurement value List of Error Codes Error Priority Description Points to checkOther problems Error Priority Cause Solution CodeDisposal DescriptionLimited Warranty Call toll freeSpecifications Factory Default SettingsSpecifications BackupNon-Invasive Blood Pressure Nibp Power supplyEnvironmental Conditions Emission Test Compliance Electromagnetic Environment Manufacturers DeclarationElectromagnetic Immunity IEC60601-1-2 IEC60601-1-2 Electromagnetic Test level Recommend separation distancePage Recommended Separation Distances Separation distance according to frequency of transmitterPage West Field Court, Lake Forest, IL 60045, USA Made in China For Customer Service Call Toll FreeDistributed by