Omron HBP-1300 IEC60601-1-2 Electromagnetic Test level, Recommend separation distance

Page 36

Immunity test

IEC60601-1-2

IEC60601-1-2

 

 

Electromagnetic

test level

test level

 

environment - guidance

 

 

 

 

 

 

 

 

 

 

Portable and mobile RF communications

 

 

 

equipment should be used no closer to any part

 

 

 

of the HBP-1300, including cables, than the

 

 

 

recommended separation distance calculated

 

 

 

from the equation applicable to the frequency of

 

 

 

the transmitter.

 

 

 

Recommend separation distance

 

 

 

d = 1.2

 

 

3 Vrms

 

P

 

 

 

d = 1.2

P

80 MHz to 800 MHz

Conducted RF

150 kHz to 80

3 Vrms

d = 2.3

P

800 MHz to 2.5 GHz

IEC 61000-4-6

MHz

 

 

 

 

 

80% AM (2Hz)

 

where P is the maximum output power rating

 

 

 

of the transmitter in watts (W) according to he

 

 

 

transmitter manufacturer and d is the recom-

 

3 Vrms

 

mended separation distance in meters (m).

 

 

 

 

 

Radiated RF

80 MHz to 2.5

3 V/m

Field strengths from fixed RF transmitters as

IEC 61000-4-3

GHz

 

determined by an electromagnetic site

 

80% AM (2Hz)

 

surveya, should be less than the compliance

 

 

 

level in each frequency rangeb.

 

 

 

Interference may occur in the vicinity of

 

 

 

equipment marked with the following symbol:

 

 

 

 

 

 

Note1: At 80 MHz and 800 MHz, the higher frequency range applies.

Note2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects, and people.

aField strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the HBP-1300 is used exceeds the applicable RF compliance level above, the HBP-1300 should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as reorienting or relocating the HBP-1300.

bOver the frequency range 150 kHz to 80MHz, field strengths should be less than 3 V/m.

34

Image 36
Contents HBP-1300 Page Introduction Intended UseExemptions General Information Setup Usage warnings and cautionsBefore use / during use Page Cleaning Maintenance and inspectionOptional rechargeable battery Non-Invasive Blood Pressure Nibp measurement For cleaning, see Using the Unit Optional AccessoriesMain unit Standard accessories Components of the ProductFeatures of the Product Key features, benefits, lookCuff Hose 3.5 m Cuff Hose 1.5mFeatures / Functions of Unit Meaning of the SymbolsFront and back of unit Using the Unit Only when the optional battery is installedLCD Display Installing the Optional Battery Pack Battery lifeLow battery Auto Power OffConnecting the AC Adapter Battery chargingBattery level AC powerSystem Settings Menu ModePress the Mode button Press the or button to change the settingUtility Mode Press the or button to change the auto power off settingCheck the Pressure Accuracy ExampleNon-Invasive Blood Pressure Nibp Measurement Non-Invasive Pressure Measurement PrinciplesConnecting the cuff Available as an optional accessoryCuff Selection and Connection Selecting the cuffAlign the artery mark Index Artery with the brachial artery Measurement results are displayed Blood pressure measurement is performed onceFailure Taking the Measurement in Auscultation Mode Irregular pulse wave detection functionBody movement detection function Use a stethoscope to perform measurementStopping the Measurement Displaying Last ReadingIrregular pulse wave icon Appears Maintenance Maintenance Inspection and Safety ManagementDevice Maintenance Accessory CareBefore turning on the power When turning on the powerAfter turning on the power Check before UseTroubleshooting Stethoscope Abnormal measurement valueMeasurement value is questionable List of Error Codes Error Priority Description Points to checkOther problems Error Priority Cause Solution CodeDisposal DescriptionLimited Warranty Call toll freeSpecifications Factory Default SettingsSpecifications BackupPower supply Environmental ConditionsNon-Invasive Blood Pressure Nibp Manufacturers Declaration Electromagnetic Immunity IEC60601-1-2Emission Test Compliance Electromagnetic Environment IEC60601-1-2 Electromagnetic Test level Recommend separation distancePage Recommended Separation Distances Separation distance according to frequency of transmitterPage For Customer Service Call Toll Free Distributed byWest Field Court, Lake Forest, IL 60045, USA Made in China