Omron HBP-1300 Stethoscope, Abnormal measurement value, Measurement value is questionable

Page 29

Abnormal measurement value

Cause / solution

The causes below are possible. Check the patient by palpation and then repeat measurement.

Body movement (chills or other trembling)

Arrhythmia.

Noise in the cuff

-A nearby person touched the patient.

-Cardiac massage was being performed.

The measurement value is questionable

Cause

 

Solution

 

 

 

Deflates quickly

 

Check for a loose cuff connection.

 

 

 

Stethoscope

Simultaneously perform measurement with a stethoscope.

 

Place the stethoscope and listen while viewing the pressure display of the

 

manometer.

Blood pressure may vary widely due to physiological effects.

The causes below are possible.

Emotional excitement or agitation

-Pain due to cuff wrapping

-White coat hypertension

Cuff size or wrapping method not correct

Cuff wrapping position on upper arm not at the same height as the heart

Patient's blood pressure not stable due to pulsus alternans, respiratory changes, or other reason

Incorrect cuff size used.

Measure circumference or patient's arm and ensure

correct sized cuff is used.

 

 

 

Cuff wrapped over thick clothing.

Ensure cuff is applied to a bare arm, or very thin

clothing.

 

 

 

Patient not seated properly.

Ensure patient is seated, feet flat on the floor, cuff at

heart level.

 

 

 

 

Ensure before measurement taken, patient has not

Patient ate, drank, or exerted themselves recently.

had food, caffeinated or alcohol beverages, or

 

exerted/exercised in the last 30 minutes.

Maintenance

27

Image 29
Contents HBP-1300 Page Exemptions IntroductionIntended Use General Information Usage warnings and cautions SetupBefore use / during use Page Maintenance and inspection CleaningOptional rechargeable battery Non-Invasive Blood Pressure Nibp measurement For cleaning, see Optional Accessories Using the UnitMain unit Standard accessories Components of the ProductKey features, benefits, look Features of the ProductCuff Hose 3.5 m Cuff Hose 1.5mFront and back of unit Features / Functions of UnitMeaning of the Symbols LCD Display Using the UnitOnly when the optional battery is installed Battery life Installing the Optional Battery PackLow battery Auto Power OffBattery charging Connecting the AC AdapterBattery level AC powerMenu Mode System SettingsPress the Mode button Press the or button to change the settingPress the or button to change the auto power off setting Utility ModeCheck the Pressure Accuracy ExampleNon-Invasive Pressure Measurement Principles Non-Invasive Blood Pressure Nibp MeasurementAvailable as an optional accessory Connecting the cuffCuff Selection and Connection Selecting the cuffAlign the artery mark Index Artery with the brachial artery Failure Measurement results are displayedBlood pressure measurement is performed once Irregular pulse wave detection function Taking the Measurement in Auscultation ModeBody movement detection function Use a stethoscope to perform measurementIrregular pulse wave icon Appears Stopping the MeasurementDisplaying Last Reading Maintenance Inspection and Safety Management MaintenanceDevice Maintenance Accessory CareWhen turning on the power Before turning on the powerAfter turning on the power Check before UseTroubleshooting Measurement value is questionable StethoscopeAbnormal measurement value Error Priority Description Points to check List of Error CodesError Priority Cause Solution Code Other problemsDisposal DescriptionCall toll free Limited WarrantyFactory Default Settings SpecificationsSpecifications BackupNon-Invasive Blood Pressure Nibp Power supplyEnvironmental Conditions Emission Test Compliance Electromagnetic Environment Manufacturers DeclarationElectromagnetic Immunity IEC60601-1-2 Recommend separation distance IEC60601-1-2 Electromagnetic Test levelPage Separation distance according to frequency of transmitter Recommended Separation DistancesPage West Field Court, Lake Forest, IL 60045, USA Made in China For Customer Service Call Toll FreeDistributed by
Related manuals
Manual 38 pages 5.6 Kb