Omron HBP-1300 Measurement results are displayed, Blood pressure measurement is performed once

Page 23

Caution

Make sure the cuff is wrapped in the correct arm position and is at the same height as the heart.

A difference of 4 inches (10 cm) in height may cause a variation in the blood pressure value of up to 7 - 8 mmHg.

Note:

If measurement is difficult due to arrhythmia, use a different blood pressure measurement method.

If the patient has acute inflammation, a pyogenic ailment, or an external wound at the location where the cuff is to be wrapped, follow the instructions of the doctor.

Non-Invasive Blood Pressure (NIBP) measurement is performed by compressing the upper arm. Some people may experience intense pain, or transient spotting caused by subcutaneous hemorrhaging may appear. The spotting will disappear over time, however, if it is possible that this will disturb the patient, try the following technique:

-Wrap a thin towel or cloth (one layer) under the cuff.

If the towel or cloth is too thick, there will be insufficient cuff compression and the blood pressure value will measure high.

If the patient moves or the cuff is touched, this may be falsely detected as a pulse and over-inflation will occur.

Do not inflate the cuff when it is not wrapped on the upper arm. This may damage the cuff.

Taking the Measurement in "Manual Mode"

1. Press the [START/STOP] button.

Blood pressure measurement is performed once.

2. The measurement results are displayed.

If a measurement value is outside the corresponding range below, the value will flash. SYS: 59 mmHg or less, or 251 mmHg or higher.

DIA: 39 mmHg or less, or 201 mmHg or higher.

PULSE: 39 bpm or less, or 201 bpm or higher.

■ Normal measurement

■ Measurement error /

 

failure

Caution

The accuracy of a flashing measurement value that is outside the measurement range is not guaranteed. Always check the patient's condition before deciding what steps to take.

Note:

If inflation is insufficient, inflation may restart while measurement is in progress.

Non-Invasive Blood Pressure (NIBP) Measurement

21

Image 23
Contents HBP-1300 Page Exemptions IntroductionIntended Use General Information Usage warnings and cautions SetupBefore use / during use Page Maintenance and inspection CleaningOptional rechargeable battery Non-Invasive Blood Pressure Nibp measurement For cleaning, see Components of the Product Using the UnitOptional Accessories Main unit Standard accessoriesCuff Hose 1.5m Features of the ProductKey features, benefits, look Cuff Hose 3.5 mFront and back of unit Features / Functions of UnitMeaning of the Symbols LCD Display Using the UnitOnly when the optional battery is installed Auto Power Off Installing the Optional Battery PackBattery life Low batteryAC power Connecting the AC AdapterBattery charging Battery levelPress the or button to change the setting System SettingsMenu Mode Press the Mode buttonExample Utility ModePress the or button to change the auto power off setting Check the Pressure AccuracyNon-Invasive Pressure Measurement Principles Non-Invasive Blood Pressure Nibp MeasurementSelecting the cuff Connecting the cuffAvailable as an optional accessory Cuff Selection and ConnectionAlign the artery mark Index Artery with the brachial artery Failure Measurement results are displayedBlood pressure measurement is performed once Use a stethoscope to perform measurement Taking the Measurement in Auscultation ModeIrregular pulse wave detection function Body movement detection functionIrregular pulse wave icon Appears Stopping the MeasurementDisplaying Last Reading Accessory Care MaintenanceMaintenance Inspection and Safety Management Device MaintenanceCheck before Use Before turning on the powerWhen turning on the power After turning on the powerTroubleshooting Measurement value is questionable StethoscopeAbnormal measurement value Error Priority Description Points to check List of Error CodesDescription Other problemsError Priority Cause Solution Code DisposalCall toll free Limited WarrantyBackup SpecificationsFactory Default Settings SpecificationsNon-Invasive Blood Pressure Nibp Power supplyEnvironmental Conditions Emission Test Compliance Electromagnetic Environment Manufacturers DeclarationElectromagnetic Immunity IEC60601-1-2 Recommend separation distance IEC60601-1-2 Electromagnetic Test levelPage Separation distance according to frequency of transmitter Recommended Separation DistancesPage West Field Court, Lake Forest, IL 60045, USA Made in China For Customer Service Call Toll FreeDistributed by