Before using the unit, verify that none of the following apply to the patient:

-Poor peripheral circulation, noticeably low blood pressure, or low body temperature (there will be low blood flow to the measurement position)

-The patient uses an artificial heart and lung (there will be no pulse)

-An SpO2 sensor and the cuff are attached to the same arm

-The patient has an aneurysm

-The patient has frequent arrhythmia

-Body motions such as convulsions, arterial pulsations, or trembling (cardiac massage in progress, minute continuous vibrations, rheumatism, etc.)

Before use, visually inspect the unit to make sure there are no deformations due to falling, and that there is no dirt or moisture on the unit.

When the unit has not been used for an extended period of time, always verify that it operates normally and safely before use.

Do not use in a location where the unit may easily fall. In the event that the unit falls, verify that it operates normally and safely.

Introduction

5

Image 7
Contents HBP-1300 Page Intended Use IntroductionExemptions General Information Usage warnings and cautions SetupBefore use / during use Page Maintenance and inspection CleaningOptional rechargeable battery Non-Invasive Blood Pressure Nibp measurement For cleaning, see Components of the Product Using the UnitOptional Accessories Main unit Standard accessoriesCuff Hose 1.5m Features of the ProductKey features, benefits, look Cuff Hose 3.5 mMeaning of the Symbols Features / Functions of UnitFront and back of unit Only when the optional battery is installed Using the UnitLCD Display Auto Power Off Installing the Optional Battery PackBattery life Low batteryAC power Connecting the AC AdapterBattery charging Battery levelPress the or button to change the setting System SettingsMenu Mode Press the Mode buttonExample Utility ModePress the or button to change the auto power off setting Check the Pressure AccuracyNon-Invasive Pressure Measurement Principles Non-Invasive Blood Pressure Nibp MeasurementSelecting the cuff Connecting the cuffAvailable as an optional accessory Cuff Selection and ConnectionAlign the artery mark Index Artery with the brachial artery Blood pressure measurement is performed once Measurement results are displayedFailure Use a stethoscope to perform measurement Taking the Measurement in Auscultation ModeIrregular pulse wave detection function Body movement detection functionDisplaying Last Reading Stopping the MeasurementIrregular pulse wave icon Appears Accessory Care MaintenanceMaintenance Inspection and Safety Management Device MaintenanceCheck before Use Before turning on the powerWhen turning on the power After turning on the powerTroubleshooting Abnormal measurement value StethoscopeMeasurement value is questionable Error Priority Description Points to check List of Error CodesDescription Other problemsError Priority Cause Solution Code DisposalCall toll free Limited WarrantyBackup SpecificationsFactory Default Settings SpecificationsEnvironmental Conditions Power supplyNon-Invasive Blood Pressure Nibp Electromagnetic Immunity IEC60601-1-2 Manufacturers DeclarationEmission Test Compliance Electromagnetic Environment Recommend separation distance IEC60601-1-2 Electromagnetic Test levelPage Separation distance according to frequency of transmitter Recommended Separation DistancesPage Distributed by For Customer Service Call Toll FreeWest Field Court, Lake Forest, IL 60045, USA Made in China