Maintenance

Maintenance Inspection and Safety Management

The HBP-1300 must be maintained to ensure functionality and to secure the safety of patients and operators.

Daily checks and maintenance should be performed by the operator. (page 25)

In addition, qualified personnel are necessary to maintain the performance and the safety, and to conduct periodic inspections. We recommend that the verification test be performed at least once a year. (page 17)

Managing Supplies

Keep the following supplies on hand.

• GS CUFF - All sizes

Device Maintenance

Cleaning and disinfecting should be performed in accordance with your facility's infection control practice and OSHA regulations.

Surface cleaning

Wipe with a cloth that has been moistened with isopropyl alcohol diluted to 50 v/v%, or ethyl alcohol (disinfection alcohol) diluted to 76.9 - 81.4 v/v% and wrung out.

Do not wipe the DC jack or allow it to become wet.

Removing dust

Use a moistened cotton bud to remove dust that has accumulated on the vent ports.

Service

The device requires no routine service other than cleaning, and visually checking the cuffs, tubing, etc.

Caution

Do not sterilize by autoclave or gas sterilization (EOG, formaldehyde gas, high-concentration ozone, etc.).

If using an antiseptic solution for cleaning, follow the instructions of the manufacturer.

Accessory Care

Non-Invasive Blood Pressure Measurement (NIBP)

Cuff

Wipe clean on the surface of the cuff with a cloth moistened with a 70 v/v% dilution of isopropyl alcohol, or a 76.9 to 81.4 v/v% dilution of disinfection ethanol (ethyl alcohol).

Do not allow any liquids inside the cuff. If a liquid gets in the cuff, dry the inside well.

24

Image 26
Contents HBP-1300 Page Exemptions IntroductionIntended Use General Information Setup Usage warnings and cautionsBefore use / during use Page Cleaning Maintenance and inspectionOptional rechargeable battery Non-Invasive Blood Pressure Nibp measurement For cleaning, see Main unit Standard accessories Using the UnitOptional Accessories Components of the ProductCuff Hose 3.5 m Features of the ProductKey features, benefits, look Cuff Hose 1.5mFront and back of unit Features / Functions of UnitMeaning of the Symbols LCD Display Using the UnitOnly when the optional battery is installed Low battery Installing the Optional Battery PackBattery life Auto Power OffBattery level Connecting the AC AdapterBattery charging AC powerPress the Mode button System SettingsMenu Mode Press the or button to change the settingCheck the Pressure Accuracy Utility ModePress the or button to change the auto power off setting ExampleNon-Invasive Blood Pressure Nibp Measurement Non-Invasive Pressure Measurement PrinciplesCuff Selection and Connection Connecting the cuffAvailable as an optional accessory Selecting the cuffAlign the artery mark Index Artery with the brachial artery Failure Measurement results are displayedBlood pressure measurement is performed once Body movement detection function Taking the Measurement in Auscultation ModeIrregular pulse wave detection function Use a stethoscope to perform measurementIrregular pulse wave icon Appears Stopping the MeasurementDisplaying Last Reading Device Maintenance MaintenanceMaintenance Inspection and Safety Management Accessory CareAfter turning on the power Before turning on the powerWhen turning on the power Check before UseTroubleshooting Measurement value is questionable StethoscopeAbnormal measurement value List of Error Codes Error Priority Description Points to checkDisposal Other problemsError Priority Cause Solution Code DescriptionLimited Warranty Call toll freeSpecifications SpecificationsFactory Default Settings BackupNon-Invasive Blood Pressure Nibp Power supplyEnvironmental Conditions Emission Test Compliance Electromagnetic Environment Manufacturers DeclarationElectromagnetic Immunity IEC60601-1-2 IEC60601-1-2 Electromagnetic Test level Recommend separation distancePage Recommended Separation Distances Separation distance according to frequency of transmitterPage West Field Court, Lake Forest, IL 60045, USA Made in China For Customer Service Call Toll FreeDistributed by