Omron HBP-1300 specifications Troubleshooting

Page 28

Troubleshooting

The power does not turn on

Cause

Solution

 

 

If the unit is being powered by the battery, the battery

Insert battery or replace with a new battery (page 14).

is not installed or the charge is depleted.

 

Internal part failure

Disconnect the AC adapter, remove the battery, and

contact OMRON HEALTHCARE Customer Service.

 

 

 

• Check if the AC adapter is disconnected or the connection is loose.

• Check if the AC adapter or battery has failed.

 

 

 

 

 

The unit display does not operate

 

Cause / solution

 

Stop using the unit and contact OMRON HEALTHCARE Customer Service.

The unit becomes hot

Cause

Solution

 

 

An object is on top of the unit or right next to the unit.

Keep the area around the unit free of objects.

 

 

If the unit becomes too hot to be touched, there may be a problem in the unit. Turn off the unit power, disconnect the AC adapter, remove the battery, and contact OMRON HEALTHCARE Customer Service at 1-877-216-1336.

The unit is connected to a power outlet, but it runs on the the battery pack

Cause / solution

If AC power cannot be supplied, the unit will operate by battery only.

Check if the AC adapter is properly connected to the unit.

Check if AC adapter is connected to a power outlet.

Check that the electrical outlet is working by connecting a different device to the same power outlet.

The cuff does not inflate when the [START/STOP] button is pressed

Cause

Solution

 

 

Loose cuff tube connection.

Check the connection.

 

 

There is an air leak in the cuff.

Replace the cuff.

 

 

If pressure is displayed, the cuff tube is bent.

Make sure no part of the cuff tube is bent.

 

 

 

 

Measurement was not possible

 

Cause / solution

 

Check the patient by palpation or other method.

After checking the patient, check the error code and see "List of Error Codes" (page 28) for Non Invasive Blood Pressure (NIBP) measurement.

26

Image 28
Contents HBP-1300 Page Intended Use IntroductionExemptions General Information Setup Usage warnings and cautionsBefore use / during use Page Cleaning Maintenance and inspectionOptional rechargeable battery Non-Invasive Blood Pressure Nibp measurement For cleaning, see Using the Unit Optional AccessoriesMain unit Standard accessories Components of the ProductFeatures of the Product Key features, benefits, lookCuff Hose 3.5 m Cuff Hose 1.5mMeaning of the Symbols Features / Functions of UnitFront and back of unit Only when the optional battery is installed Using the UnitLCD Display Installing the Optional Battery Pack Battery lifeLow battery Auto Power OffConnecting the AC Adapter Battery chargingBattery level AC powerSystem Settings Menu ModePress the Mode button Press the or button to change the settingUtility Mode Press the or button to change the auto power off settingCheck the Pressure Accuracy ExampleNon-Invasive Blood Pressure Nibp Measurement Non-Invasive Pressure Measurement PrinciplesConnecting the cuff Available as an optional accessoryCuff Selection and Connection Selecting the cuffAlign the artery mark Index Artery with the brachial artery Blood pressure measurement is performed once Measurement results are displayedFailure Taking the Measurement in Auscultation Mode Irregular pulse wave detection functionBody movement detection function Use a stethoscope to perform measurementDisplaying Last Reading Stopping the MeasurementIrregular pulse wave icon Appears Maintenance Maintenance Inspection and Safety ManagementDevice Maintenance Accessory CareBefore turning on the power When turning on the powerAfter turning on the power Check before UseTroubleshooting Abnormal measurement value StethoscopeMeasurement value is questionable List of Error Codes Error Priority Description Points to checkOther problems Error Priority Cause Solution CodeDisposal DescriptionLimited Warranty Call toll freeSpecifications Factory Default SettingsSpecifications BackupEnvironmental Conditions Power supplyNon-Invasive Blood Pressure Nibp Electromagnetic Immunity IEC60601-1-2 Manufacturers DeclarationEmission Test Compliance Electromagnetic Environment IEC60601-1-2 Electromagnetic Test level Recommend separation distancePage Recommended Separation Distances Separation distance according to frequency of transmitterPage Distributed by For Customer Service Call Toll FreeWest Field Court, Lake Forest, IL 60045, USA Made in China