Power supply
AC adapter | Input voltage range: AC 100 V to 240 V |
| Rated Current: 1.6 A |
| Frequency: 50/60 Hz |
| Output voltage range: DC 6 V ±5% |
|
|
Optional rechargeable battery | Type: 3.6 V, 1900 mAh |
| Number of operation cycles when fully charged: 300 |
| Measurement conditions |
| • New battery fully charged |
| • Ambient temperature of 23°C (73.4°F) |
| • Using |
| • SYS 120 / DIA 80 / PR 60 (Inflation setting: AUTO) |
| • One |
|
|
Environmental Conditions
Operational temperature and | Temperature range: 5 to 40°C (41 to 104°F) |
humidity | Humidity range: 15 to 85%RH (not condensed) |
| Atmospheric pressure: 700 to 1060hPa |
|
|
Storage and transportation | Temperature range: |
| Humidity range: 10 to 95%RH (not condensed) |
| Atmospheric pressure: 500 to 1060hPa |
|
|
EMC: Reference standard | |
| Medical electrical equipment - |
| Part |
| - Collateral standard: Electromagnetic compatibility - Requirements and |
| tests. |
|
|
Non-Invasive Blood Pressure (NIBP)
Measurement technology | Oscillometric |
|
|
Measurement method | Dynamic Linear Deflation method |
|
|
Pressure display range | 0 to 300 mmHg |
|
|
Pressure display accuracy | Within ±3 mmHg or 2 % |
|
|
NIBP measurement range | SYS 60 to 250 mmHg |
| DIA 40 to 200 mmHg |
| Pulse rate 40 to 200 /min |
|
|
NIBP accuracy* | Maximum mean error within ±5mmHg |
| Maximum standard deviation within ±8mmHg |
|
|
Pulse rate accuracy | Within ±5 % of reading |
|
|
Reference standard: | IEC |
| ANSI/AAMI |
|
*Comparison with auscultation method performed by a trained professional. DIA determined by the auscultation method is "K5".
32