Power supply

AC adapter

Input voltage range: AC 100 V to 240 V

 

Rated Current: 1.6 A

 

Frequency: 50/60 Hz

 

Output voltage range: DC 6 V ±5%

 

 

Optional rechargeable battery

Type: 3.6 V, 1900 mAh

 

Number of operation cycles when fully charged: 300

 

Measurement conditions

 

• New battery fully charged

 

• Ambient temperature of 23°C (73.4°F)

 

• Using M-size cuff

 

• SYS 120 / DIA 80 / PR 60 (Inflation setting: AUTO)

 

• One 5-minute cycle consisting of "cuff measurement time + wait time"

 

 

Environmental Conditions

Operational temperature and

Temperature range: 5 to 40°C (41 to 104°F)

humidity

Humidity range: 15 to 85%RH (not condensed)

 

Atmospheric pressure: 700 to 1060hPa

 

 

Storage and transportation

Temperature range: -20 to 60°C (-4 to 140°F)

 

Humidity range: 10 to 95%RH (not condensed)

 

Atmospheric pressure: 500 to 1060hPa

 

 

EMC: Reference standard

IEC60601-1-2:2007

 

Medical electrical equipment -

 

Part 1-2: General requirements for basic safety and essential performance

 

- Collateral standard: Electromagnetic compatibility - Requirements and

 

tests.

 

 

Non-Invasive Blood Pressure (NIBP)

Measurement technology

Oscillometric

 

 

Measurement method

Dynamic Linear Deflation method

 

 

Pressure display range

0 to 300 mmHg

 

 

Pressure display accuracy

Within ±3 mmHg or 2 %

 

 

NIBP measurement range

SYS 60 to 250 mmHg

 

DIA 40 to 200 mmHg

 

Pulse rate 40 to 200 /min

 

 

NIBP accuracy*

Maximum mean error within ±5mmHg

 

Maximum standard deviation within ±8mmHg

 

 

Pulse rate accuracy

Within ±5 % of reading

 

 

Reference standard:

IEC 80601-2-30:2009

 

ANSI/AAMI SP-10:2002+A1:2003+A2: 2006/(R)2008

 

ISO81060-2:2009

*Comparison with auscultation method performed by a trained professional. DIA determined by the auscultation method is "K5".

32

Image 34
Contents HBP-1300 Page Intended Use IntroductionExemptions General Information Setup Usage warnings and cautionsBefore use / during use Page Cleaning Maintenance and inspectionOptional rechargeable battery Non-Invasive Blood Pressure Nibp measurement For cleaning, see Main unit Standard accessories Using the UnitOptional Accessories Components of the ProductCuff Hose 3.5 m Features of the ProductKey features, benefits, look Cuff Hose 1.5mMeaning of the Symbols Features / Functions of UnitFront and back of unit Only when the optional battery is installed Using the UnitLCD Display Low battery Installing the Optional Battery PackBattery life Auto Power OffBattery level Connecting the AC AdapterBattery charging AC powerPress the Mode button System SettingsMenu Mode Press the or button to change the settingCheck the Pressure Accuracy Utility ModePress the or button to change the auto power off setting ExampleNon-Invasive Blood Pressure Nibp Measurement Non-Invasive Pressure Measurement PrinciplesCuff Selection and Connection Connecting the cuffAvailable as an optional accessory Selecting the cuffAlign the artery mark Index Artery with the brachial artery Blood pressure measurement is performed once Measurement results are displayedFailure Body movement detection function Taking the Measurement in Auscultation ModeIrregular pulse wave detection function Use a stethoscope to perform measurementDisplaying Last Reading Stopping the MeasurementIrregular pulse wave icon Appears Device Maintenance MaintenanceMaintenance Inspection and Safety Management Accessory CareAfter turning on the power Before turning on the powerWhen turning on the power Check before UseTroubleshooting Abnormal measurement value StethoscopeMeasurement value is questionable List of Error Codes Error Priority Description Points to checkDisposal Other problemsError Priority Cause Solution Code DescriptionLimited Warranty Call toll freeSpecifications SpecificationsFactory Default Settings BackupEnvironmental Conditions Power supplyNon-Invasive Blood Pressure Nibp Electromagnetic Immunity IEC60601-1-2 Manufacturers DeclarationEmission Test Compliance Electromagnetic Environment IEC60601-1-2 Electromagnetic Test level Recommend separation distancePage Recommended Separation Distances Separation distance according to frequency of transmitterPage Distributed by For Customer Service Call Toll FreeWest Field Court, Lake Forest, IL 60045, USA Made in China