Omron HBP-1300 Utility Mode, Press the or button to change the auto power off setting, Example

Page 19

Utility Mode

"Utility Mode" allows you to configure the "Auto Power Off" and "Pressure Accuracy Confirmation" setting.

1. Confirm that the device is switched off.

2. Hold down the [MODE] button until the "Auto Power Off" setting screen appears.

The "Initial Inflation Pressure Value" setting screen appears,

 

and changes to the "Auto Power Off" setting screen appears.

"5" or "10" appears.

3. Press the [ ] or [ ] button to change the auto power off setting.

The "5 min." or "10 min." setting is entered.

When the unit is powered by the battery pack and is not used for the set time ("5 min." or "10 min."), the power automatically turns off to save battery power.

If an alarm has occurred, the power does not turn off automatically.

4. When you have completed the "Auto Power Off" setting, press the [MODE] button.

The "Pressure Accuracy Confirmation" screen appears. "0" appears.

5. Check the Pressure Accuracy.

Change the pressure externally.

"0" appears.

Compare with the displayed value and verify that there is no problem.

 

6. When you have completed "Pressure Accuracy Confirmation", press the [START/STOP] button.

The power turns off.

Example:

1. Connect the blood pressure monitor, the standard pressure gauge, and the cuff and rubber ball.

For details on the standard pressure gauge, contact OMRON HEALTHCARE Customer Service.

2.Check the blood pressure value of the blood pressure monitor and the blood pressure value of the standard pressure gauge.

Note:

The standard is within ± 3 mmHg or 2 % of manometer reading.

To rapidly deflate the cuff, press the [] button.

To repeat "Pressure Accuracy Confirmation", turn off the power and repeat the procedure from step 1 of "Utility Mode".

System Settings

17

Image 19
Contents HBP-1300 Page Intended Use IntroductionExemptions General Information Usage warnings and cautions SetupBefore use / during use Page Maintenance and inspection CleaningOptional rechargeable battery Non-Invasive Blood Pressure Nibp measurement For cleaning, see Components of the Product Using the UnitOptional Accessories Main unit Standard accessoriesCuff Hose 1.5m Features of the ProductKey features, benefits, look Cuff Hose 3.5 mMeaning of the Symbols Features / Functions of UnitFront and back of unit Only when the optional battery is installed Using the UnitLCD Display Auto Power Off Installing the Optional Battery PackBattery life Low batteryAC power Connecting the AC AdapterBattery charging Battery levelPress the or button to change the setting System SettingsMenu Mode Press the Mode buttonExample Utility ModePress the or button to change the auto power off setting Check the Pressure AccuracyNon-Invasive Pressure Measurement Principles Non-Invasive Blood Pressure Nibp MeasurementSelecting the cuff Connecting the cuffAvailable as an optional accessory Cuff Selection and ConnectionAlign the artery mark Index Artery with the brachial artery Blood pressure measurement is performed once Measurement results are displayedFailure Use a stethoscope to perform measurement Taking the Measurement in Auscultation ModeIrregular pulse wave detection function Body movement detection functionDisplaying Last Reading Stopping the MeasurementIrregular pulse wave icon Appears Accessory Care MaintenanceMaintenance Inspection and Safety Management Device MaintenanceCheck before Use Before turning on the powerWhen turning on the power After turning on the powerTroubleshooting Abnormal measurement value StethoscopeMeasurement value is questionable Error Priority Description Points to check List of Error CodesDescription Other problemsError Priority Cause Solution Code DisposalCall toll free Limited WarrantyBackup SpecificationsFactory Default Settings SpecificationsEnvironmental Conditions Power supplyNon-Invasive Blood Pressure Nibp Electromagnetic Immunity IEC60601-1-2 Manufacturers DeclarationEmission Test Compliance Electromagnetic Environment Recommend separation distance IEC60601-1-2 Electromagnetic Test levelPage Separation distance according to frequency of transmitter Recommended Separation DistancesPage Distributed by For Customer Service Call Toll FreeWest Field Court, Lake Forest, IL 60045, USA Made in China