Omron HBP-1300 specifications Stopping the Measurement, Displaying Last Reading

Page 25

Stopping the Measurement

To stop measurement while measurement is in progress, press [START/STOP] button.

Displaying Last Reading

The previous measurement value (systolic blood pressure, diastolic blood pressure, and pulse rate) and whether or not an irregular pulse wave was detected can be displayed. In the case of auscultation measurement, the auscultation icon appears.

Press the [ ] or [ ] button to display the previous data. This function is also available when the device is switched off.

If an irregular pulse wave or

 

 

 

 

 

 

body movement was detected,

 

 

Lights while the previous

the irregular pulse wave icon

 

 

appears.

 

 

data are being displayed.

 

 

 

 

When the data was measured in "Auscultation Mode", the auscultation icon appears.

If held down at least 3 seconds when a measurement result is displayed, the measured data is cleared without being stored.

Note:

When the monitor is left idle for one minute, the backlight will disappear.

Non-Invasive Blood Pressure (NIBP) Measurement

23

Image 25
Contents HBP-1300 Page Intended Use IntroductionExemptions General Information Usage warnings and cautions SetupBefore use / during use Page Maintenance and inspection CleaningOptional rechargeable battery Non-Invasive Blood Pressure Nibp measurement For cleaning, see Optional Accessories Using the UnitMain unit Standard accessories Components of the ProductKey features, benefits, look Features of the ProductCuff Hose 3.5 m Cuff Hose 1.5mMeaning of the Symbols Features / Functions of UnitFront and back of unit Only when the optional battery is installed Using the UnitLCD Display Battery life Installing the Optional Battery PackLow battery Auto Power OffBattery charging Connecting the AC AdapterBattery level AC powerMenu Mode System SettingsPress the Mode button Press the or button to change the settingPress the or button to change the auto power off setting Utility ModeCheck the Pressure Accuracy ExampleNon-Invasive Pressure Measurement Principles Non-Invasive Blood Pressure Nibp MeasurementAvailable as an optional accessory Connecting the cuffCuff Selection and Connection Selecting the cuffAlign the artery mark Index Artery with the brachial artery Blood pressure measurement is performed once Measurement results are displayedFailure Irregular pulse wave detection function Taking the Measurement in Auscultation ModeBody movement detection function Use a stethoscope to perform measurementDisplaying Last Reading Stopping the MeasurementIrregular pulse wave icon Appears Maintenance Inspection and Safety Management MaintenanceDevice Maintenance Accessory CareWhen turning on the power Before turning on the powerAfter turning on the power Check before UseTroubleshooting Abnormal measurement value StethoscopeMeasurement value is questionable Error Priority Description Points to check List of Error CodesError Priority Cause Solution Code Other problemsDisposal DescriptionCall toll free Limited WarrantyFactory Default Settings SpecificationsSpecifications BackupEnvironmental Conditions Power supplyNon-Invasive Blood Pressure Nibp Electromagnetic Immunity IEC60601-1-2 Manufacturers DeclarationEmission Test Compliance Electromagnetic Environment Recommend separation distance IEC60601-1-2 Electromagnetic Test levelPage Separation distance according to frequency of transmitter Recommended Separation DistancesPage Distributed by For Customer Service Call Toll FreeWest Field Court, Lake Forest, IL 60045, USA Made in China