Omron HBP-1300 Taking the Measurement in Auscultation Mode, Body movement detection function

Page 24

Irregular pulse wave detection function

If the pulse wave interval becomes irregular during measurement, the irregular pulse wave detection icon will light to notify you.

Body movement detection function

If body movement is detected during measurement, deflation stops for 5 seconds. The irregular pulse wave icon appears in the measurement result display.

Deflation stopped

After 5 seconds, measurement resumes, and an attempt is made to complete measurement in one cycle.

Taking the Measurement in "Auscultation Mode"

Use a stethoscope to perform measurement.

When set to "ON", SYS and DIA can be recorded using auscultation measurement. To determine SYS and DIA during measurement, press the [MODE] button.

SYS is registered the first time and DIA is registered the second time you press the [MODE] button during deflation.

After DIA is determined, the cuff rapidly deflates and SYS and DIA are displayed as the measurement results.

The pulse rate does not appear in the measurement result display.

During deflation, the [ ] button can be held down to re-inflate, or the [ ] button can be held down to deflate faster.

Stored data that was measured in "Auscultation Mode" is displayed when the auscultation icon is lit.

During re-inflation

Note:

The body movement detection function is disabled while "Auscultation Mode" is in use. For Auscultation Mode settings, see page 16.

22

Image 24
Contents HBP-1300 Page Introduction Intended UseExemptions General Information Setup Usage warnings and cautionsBefore use / during use Page Cleaning Maintenance and inspectionOptional rechargeable battery Non-Invasive Blood Pressure Nibp measurement For cleaning, see Using the Unit Optional AccessoriesMain unit Standard accessories Components of the ProductFeatures of the Product Key features, benefits, lookCuff Hose 3.5 m Cuff Hose 1.5mFeatures / Functions of Unit Meaning of the SymbolsFront and back of unit Using the Unit Only when the optional battery is installedLCD Display Installing the Optional Battery Pack Battery lifeLow battery Auto Power OffConnecting the AC Adapter Battery chargingBattery level AC powerSystem Settings Menu ModePress the Mode button Press the or button to change the settingUtility Mode Press the or button to change the auto power off settingCheck the Pressure Accuracy ExampleNon-Invasive Blood Pressure Nibp Measurement Non-Invasive Pressure Measurement Principles Connecting the cuff Available as an optional accessory Cuff Selection and Connection Selecting the cuffAlign the artery mark Index Artery with the brachial artery Measurement results are displayed Blood pressure measurement is performed onceFailure Taking the Measurement in Auscultation Mode Irregular pulse wave detection functionBody movement detection function Use a stethoscope to perform measurementStopping the Measurement Displaying Last ReadingIrregular pulse wave icon Appears Maintenance Maintenance Inspection and Safety ManagementDevice Maintenance Accessory CareBefore turning on the power When turning on the powerAfter turning on the power Check before UseTroubleshooting Stethoscope Abnormal measurement valueMeasurement value is questionable List of Error Codes Error Priority Description Points to checkOther problems Error Priority Cause Solution CodeDisposal DescriptionLimited Warranty Call toll freeSpecifications Factory Default SettingsSpecifications BackupPower supply Environmental ConditionsNon-Invasive Blood Pressure Nibp Manufacturers Declaration Electromagnetic Immunity IEC60601-1-2Emission Test Compliance Electromagnetic Environment IEC60601-1-2 Electromagnetic Test level Recommend separation distancePage Recommended Separation Distances Separation distance according to frequency of transmitterPage For Customer Service Call Toll Free Distributed byWest Field Court, Lake Forest, IL 60045, USA Made in China