Troubleshooting
Problem | Probable Cause | Corrective Action | |
Nothing appears in the | Batteries are drained. | Replace all batteries with | |
display, after I press |
| new ones. Alkaline batteries | |
the START button. |
| recommended. | |
|
|
| |
| Battery terminals are not in | Reinstall the batteries with negative | |
| the correct position. | and positive terminals matching | |
|
| those indicated in the battery | |
|
| compartment. | |
|
|
| |
The cuff does not inflate. | Battery voltage is too low. Low | Replace all batteries with new ones. | |
. | battery symbol blinks. [If the | Alkaline batteries recommended. | |
batteries are drained completely, |
| ||
|
| ||
| the mark does not appear.] |
| |
|
|
| |
The unit does not | The cuff is not fastened properly. | Fasten the cuff correctly. | |
measure. Readings are |
|
| |
too high or too low. |
|
| |
You moved your arm or body | Make sure you remain very still and | ||
| |||
| during the measurement. | quiet during the measurement. | |
|
|
| |
| The cuff position is not correct. | Sit comfortably and still. Make sure | |
|
| the cuff is at the same level as your | |
|
| heart. | |
|
|
| |
| You are using the wrong | See Pg. | |
| size cuff. | Cuff." | |
|
|
| |
The value is different | The healthcare professional may | See Pg. | |
from that measured at a | be using a different sized cuff. | Readings Are Different Between | |
clinic or doctor’s office |
| Home and at the Doctor’s Office." | |
|
|
| |
| Your measurements may | See Pg. | |
| be elevated by white coat | Blood Pressure at Home?" | |
| hypertension |
| |
|
| ||
ActiLink/computer | • Is your computer on and your ActiLink plugged in? | ||
software is not receiving | • Move the blood pressure monitor closer to the ActiLink with a clear | ||
new measurements | |||
| • Replace all batteries with new ones | ||
|
|
|
NOTE: If the actions described above do not solve the problem, call
A&D Medical