TROUBLESHOOTING
Refer to the following instructions to check and correct any failure before contacting service personnel.
SYMPTOM |
| POSSIBLE CAUSE |
|
| CORRECTIVE ACTION |
|
|
|
|
|
|
Scale does not turn on | 1. | Dead Battery | 1. |
| Replace batteries |
| 2. | Faulty electrical outlet | 2. Use a different outlet | ||
| 3. | Bad power supply | 3. |
| Replace adapter |
Questionable weight or | 1. | External object | 1. |
| Remove interfering object |
the scale does not zero |
| interfering with the scale |
|
| from the scale |
| 2. The display did not show | 2. | Ask the patient to get off | ||
|
| ”0.0” before weighing |
|
| the scale, zero the scale |
|
|
|
|
| and begin weighing |
|
|
|
|
| process again |
| 3. Scale is not placed on a | 3. | Place the scale on a level | ||
|
| level floor |
|
| floor and begin weighing |
|
|
|
|
| process again |
| 4. | Scale is out of calibration | 4. | Check weight with known | |
|
|
|
|
| weight value |
| 5. | Improper Tare | 5. | Place the item to be tared | |
|
|
|
|
| on the scale. Press |
|
|
|
|
| REWEIGH. Once the |
|
|
|
|
| weight of the item is |
|
|
|
|
| displayed, press TARE. |
|
|
|
|
| Place the patient and the |
|
|
|
|
| tared item back on the |
|
|
|
|
| scale. Press REWEIGH |
|
|
|
|
| again. |
Weighing is performed but | The patient is not sitting still | Ask the patient to be still or | |||
the display shows “weigh” |
|
| you can change to live weight | ||
and “reweigh” every few |
|
| setting | ||
seconds; the weighing |
|
|
|
|
|
process takes too long and |
|
|
|
|
|
no weight is displayed. |
|
|
|
|
|
The display shows | The load on the scale | Remove the excess weight | |||
“Overload” message | exceeds the capacity | and use the scale according to | |||
| (800 Lbs) | its limits | |||
The display shows | The batteries are empty | Replace batteries according to | |||
“LOW BAT” message |
|
| instructions | ||
The display shows | There is a problem with one | Check load cell cable | |||
“Load Cell Error” message | or more load cells or the load | connection at the display and | |||
| cell cable is disconnected. | platform assembly ports. If the | |||
|
|
| problem is not corrected, refer | ||
|
|
| to qualified service personnel | ||
|
|
| to replace the defective load | ||
|
|
| cell |
|