Health O Meter 2600KL Tare Weight, Data Management, Automatic Tare, Removing the Tare

Page 12

OPERATION INSTRUCTIONS

02TARE WEIGHT

NOTE: Due to the scale’s sensitivity, we recommend using the REWEIGH function prior to setting the TARE weight, in order to eliminate any operator interference with the item to be tared out.

There are two ways to manually enter a tare weight (such as the weight of a wheelchair, shoes, etc): press MENU (1) and select option 02-Tare Weight or press TARE (9). If there is no weight on the scale platform (value displayed is zero and there is no tare) and the user presses the TARE button (9), the Tare Weight window appears and prompts the user to enter the TARE value and to press ENTER. NOTE: The tare weight must be entered using the following increments: 4oz, 0.2lb, ¼lb. The value entered will display in minus (-). After the TARE has been entered, the scale goes back to normal operation. This TARE value is stored in memory until changed or cleared.

Important: The TARE weight cannot exceed 200 lb.

The number that is to be changed will flash and will move from the left to the right after the appropriate number was entered or by using the c key (left) and the d key (right).

TARE display indicates the weight that has been tared out

Figure 9. Tare Weight

Automatic Tare

The user can set a tare weight by pressing the TARE button (9) while there is a weight on the scale platform. The display will come to zero and TARE will be displayed to indicate there is a tare value in the memory (as displayed in Figure 9 above).

Removing the Tare

Additional pressing of the TARE button will delete the tare value from memory, TARE will disappear from the display and the scale will resume normal operation.

03DATA MANAGEMENT

The scale manages patient data including patient ID, weight, height, tare and BMI. The value is stored in memory or transferred to PC. This function is performed by opening a new data file.

OPENING A NEW DATA FILE:

1.Press the ID button (7).

2.Using the keypad, type in the identification number.

3.Press ENTER.

This scale will also calculate Body Mass Index (BMI).

CALCULATING BMI:

1.Place the patient on the scale platform.

2.Press the BMI button (2).

11

Image 12
Contents PRO Plustm Electronic Folding Wheelchair Scale Table of Contents Specifications Scale SpecificationsGeneral Assembly Instructions Assembly Instructions Ramp Moving Instructions SET UP Scale Opening and Folding ProceduresPreparing the Scale for Use Replacing BatteriesOperation Instructions Description Function Weight Display Mode MenuOnly applies to pound values, not the metric values Tare Weight Data ManagementAutomatic Tare Removing the TareEnter Scale Settings Operation Instructions System Settings System Test User Settings Operation Instructions Maintenance MaintenanceCleaning Disconnect the scale from the AC power sourceTroubleshooting Symptom Possible Cause Corrective ActionCalibration Path Operator Action DisplayPage Parts List Description QtyWarranty Limited WarrantyWhat does the Warranty Cover? To get Warranty ServiceTable DES Matieres Precaution ET Prevention Specifications DE LA BalanceGeneralites Mode Dassemblage Mode Dassemblage Rampe daccès Instructions de retrait Installation Balance Instructions douverture et de pliageRemplacement des piles Préparation de la balance avant usageMode DE Fonctionnement Touche Description Fonction Mode Daffichage DU Poids Menu PrincipalPoids DE LA Tare Gestion DES DonneesTarage Automatique Retrait de la TareMode DE Fonctionnement Reglages DE LA Balance Mode DE Fonctionnement Reglages DU Systeme Test DE Systeme Reglages Utilisateur Mode DE Fonctionnement Nettoyage Déconnectez la balance du secteur CASymptome Cause Possible Action Corrective Localisation DES PannesCalibrage Action réalisée par lopérateur Message affichéPage Liste DES Pieces Pièce N Code Description QuantitéPour obtenir un service de garantie GarantieÍndice Precaución Y Advertencia EspecificacionesEspecificaciones DE LA Balanza Instrucciones DE Montaje Tapa TraseraInstrucciones DE Montaje Rampa Instrucciones de Movimiento Configuración Balanza Procedimientos de Apertura y PlegamientoPreparando la Balanza para ser Usada Cambiando las BateríasInstrucciones DE Operación Ítem Descripción Función Menú Modo DE Presentación DEL PesoAdministración DE Datos Peso TaraTara Automática Removiendo la TaraCalculando EL BMI Configuraciones DE LA Balanza Instrucciones DE Operación Configuraciones DEL Sistema Prueba DEL Sistema Configuraciones DEL Usuario Instrucciones DE Operación Mantenimiento MantenimientoLimpieza Desconecte la balanza de la fuente de energía CASolución DE Problemas Síntoma Posible Causa Acción CorrectivaTrayectoria DE Calibración Acción del Operador DisplayPage Lista DE Partes No. Llave Parte No Descripción Cd3Garantía Garantía Limitada¿Qué Cubre la Garantía?