Reference Manual
Rosemount 4500
AMS
Re-mapping
Fast Keys | 1, 4, 3, 6, 4 |
|
|
Right click on the device and select “Device Configuration” then select “SV Config” from the menu.
1.Click Next after setting the control loop to manual.
2.Enter desired scaled variable units in “Enter SV units” box and click Next.
3.Select scaled data option: Linear and click Next.
4.Enter pressure value position 1 and click Next.
5.Enter scaled variable position 1 and click Next.
6.Enter pressure value position 2 and click Next.
7.Enter scaled variable position 2 and click Next.
8.Enter linear offset and click Next. Skip to Step 12.
9.Select Next to acknowledge that “Pressure and Scaled Variable values for position 0 set to 0.”
10.Enter pressure value position 1 and click Next.
11.Enter scaled variable position 1 and click Next.
12.Select Next to acknowledge that the loop can be returned to automatic control.
13.Select Finish to acknowledge the method is complete.
The
Primary variable = Pressure
Secondary variable = Temperature
Tertiary variable = Scaled Variable
NOTE
The variable assigned as the primary variable drives the
The scaled variable can be remapped as the primary variable if desired.
HART Communicator
From the HOME screen, enter the fast key sequence for
1.Select OK after the control loop is set to manual (see “Setting the Loop to Manual” on page
2.Choose desired primary variable and select Enter.
3.Choose desired secondary variable and select Enter.
4.Select OK to acknowledge the tertiary variable setting.
5.Select OK to acknowledge that the loop can be returned to automatic control.