Rosemount 4500
Reference Manual
INSTALLATION
PROCEDURES
Mount the Transmitter
For dimensional drawing information refer to Appendix A: Reference Data.
Electronics Housing
Provide 0.75 inches (19 mm) of clearance for units without an LCD display. Three inches of clearance is required for cover removal if a meter is installed.
Cover Installation
Always ensure a proper seal by installing the electronics housing cover so the
The Rosemount 4500 is designed to be mounted directly to a process pipe or vessel using a standard sanitary fitting. The transmitter is available with either a 1.5 or
When installing the transmitter to the sanitary fitting it is important to use the proper sanitary clamp and gasket
Figure 2-3. Rosemount 4500
Mounting Configuration Using a
Sanitary Fitting
1.5 or
DETAIL OF SANITARY
CONNECTION
B
A
User
Supplied
Gasket and
Clamp
4500_04_AA.EPS, 4500_24_AA.EPS
Table
Description | Connection Size in (mm) | A | B |
|
|
|
|
11/2 inch | 1.50 (38) | 1.99 (50) | 1.71 (43) |
2 inch | 2.00 (51) | 2.52 (64) | 2.22 (56) |
Fractional Line Fitting | 1.50 (38) | 1.99 (50) | 1.50 (38) |
Table 2-2. Standard Sanitary Clamp Models
Clamp Model
13MHHM
13MHHS
psi @ 70 °F
(kPa @ 21 °C)
450 (3 103)
500 (3 448)
600 (4 138)
550 (3 793)
1500 (10 345)
1000 (6 896)
psi @ 250 °F
(kPa @ 121 °C)
250 (1 724)
250 (1 724)
300 (2 069)
275 (1 896)
1200 (8 276)
800 (5 517)
Recommended Torque
25
25
25
25
20
20