Range and Sensor Limits
Rosemount 4500 Sanitary Pressure Transmitter Range Limits
Units
psi
kPa
bar
kg/cm2
Range 1 AP | Range 1 GP | Range 2 | |||
min. | max. | min. | max. | min. | max. |
|
|
|
|
|
|
0.60 | 30 | 0.6 | 30 | 10 | 150 |
4.14 | 206.8 | 4.136 | 206.8 | 68.94 | 1034 |
0.041 | 2.068 | 0.04136 | 2.068 | 0.689 | 10.34 |
0.042 | 2.109 | 0.04218 | 2.109 | 0.703 | 10.54 |
|
|
|
|
|
|
Functional Specifications
Dynamic Performance
250 milliseconds (response time + dead time)
Ambient Temperature Effect per 50°F (28°C)
±0.2% CS ± 0.02% URL
Process Temperature Effect per 104°F (58°C)
±0.3% CS + 0.03% URL
Service
Liquid, gas, vapor, and steam applications
Zero and Span Adjustment
Zero and span values can be set anywhere within the range. Span must be greater than or equal to the minimum span.
Output
Digital process variable superimposed on
Power Supply
External power supply required. Standard transmitter
Load Limitations
Maximum loop resistance is determined by the voltage level of the external power supply, as described by:
Max. Loop Resistance = 43.5 (Power Supply Voltage – 11.25)
(Ohms) | 1355 |
|
|
|
|
|
|
|
|
|
|
|
| ||
1000 |
|
|
|
|
|
| |
|
|
|
|
|
|
| |
Load | 500 |
|
|
|
| Operating | |
|
|
|
| ||||
0 |
|
|
|
| |||
|
|
|
| Region | |||
|
|
|
|
|
| ||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| 11.25 | 20 | 30 |
42.4
Voltage (V dc)
Communication requires a minimum
loop resistance of 250 ohms.