Philips M3860A, M3861A manual Placing and Securing the HeartStart FR2+

Page 19

2-5

3.Press the upper Option button to activate the test.

4.OR make no selection for 10 seconds, and the selftest will start automatically if the FR2+ has been turned off for at least 5 minutes.

N O T E : If you connect defibrillator pads (that are applied to the patient) to the FR2+ during a battery insertion selftest, the selftest will stop and the FR2+ will go to its standby mode to be ready for use.

2

GOOD BATTERY

When the automatic part of the selftest is successfully completed, the screen displays a message that the test has passed, and then automatically starts the interactive part of the selftest. It is important to press the buttons and verify the indicators to ensure that the FR2+ will be ready for use.

When the entire selftest is complete, the FR2+ automatically turns off and returns to standby mode. In the standby mode, the Status Indicator displays a flashing black hourglass. This means that the FR2+ has passed its most recent self-test and is therefore ready for use, simply by pressing the On/Off button to turn it on.

P hili ps Me dical Sy stems

Placing and Securing the HeartStart FR2+

Place the HeartStart FR2+ Defibrillator in an accessible area with the Status Indicator easily visible. Useful accessories for placing and securing the FR2+ include a carrying case, which is suitable for use with a wall mount bracket or defibrillator cabinet. (See Appendix A for a list of accessories.)

N O T E : Do not store the FR2+ with the defibrillator pads attached. Do not open the pads package until ready for use.

With the battery installed and the FR2+ stored in appropriate environmental conditions, the FR2+ performs detailed periodic selftests to make sure that it remains ready for use. (See Appendix B for the environmental storage specifications.)

While the FR2+ is in the standby mode, the Status Indicator shows the flashing black hourglass unless the periodic selftests detect a problem. If a problem is detected, the Status Indicator will show a flashing red X or a solid red X and the FR2+ will beep (“chirp”) to alert you to the need for troubleshooting. (See Chapter 4, “Maintaining, Testing, and Troubleshooting Your HeartStart FR2+,” for instructions.)

Preparing Your HeartStart FR2+ for Use

Image 19
Contents S T R U C T I O N S F O R U S E A R T S T a R T F R 2 + D E F I B R I L L a T O RHeartStart FR2+ Defibrillator I C K R E F E R E N C E C a R D Intentionally left blank A R T S T a R T F R 2 + M3860A, M3861A Defibrillator Device Tracking Contents Setup and Advanced Mode Features Clinical and Safety ConsiderationsData Management and Review Appendices Accessories for the HeartStart FR2+Technical Specifications Glossary of Symbols and ControlsM3860A/M3861A Heartstart FR2+ Introduction to the HeartStart FR2+ What is the HeartStart FR2+?Model M3860A Model M3861A How Does the HeartStart FR2+ Work? When Is the HeartStart FR2+ Used?How Is the HeartStart FR2+ Supplied? M3860A/M3861A Heartstart FR2+ Preparing Your HeartStart FR2+ for Use Installing the BatteryOverview M3860A/M3861A Heartstart FR2+ Setting the Clock Running the Battery Insertion Selftest Press the upper Option button to activate the test Placing and Securing the HeartStart FR2+M3860A/M3861A Heartstart FR2+ Using Your HeartStart FR2+ Preparation ECG Analysis and Monitoring Shock Delivery ECG Display for Ongoing Observation M3860A/M3861A Heartstart FR2+ Maintenance Daily Monthly Maintenance task/response After Using the HeartStart FR2+Operator’s checklist Cleaning the HeartStart FR2+Date Scheduled frequency Operators ChecklistBattery insertion selftest TestingM3860A/M3861A Heartstart FR2+ Hili ps Me dical Sy stems Test Feature Test descriptionDevice history Periodic selftestsBattery History Recommended action during an emergency Troubleshooting GuideStatus indicator summary Status indicator MeaningStatus Indicator Flashing Black Hourglass Symptom Possible cause Recommended actionSymptom Possible cause Symptom General troubleshootingFR2+ has been physically damaged M3860A/M3861A Heartstart FR2+ Indications Clinical and Safety ConsiderationsSafety Considerations Clinical ConsiderationsGeneral dangers, warnings, and cautions Safety level Possible shock or fire hazard, or explosionSafety level Possible improper device performance Safety levels Possible ECG misinterpretation Safety level Possible shock hazardDefibrillation warnings and cautions Monitoring cautions Maintenance cautionsSafety level Possible fire or shock hazard Non-protocol parameters Setup and Advanced Mode FeaturesSetup Overview Parameter Settings Default DescriptionAutomatic protocol parameters NSA action Manual override parameters Parameter Settings DefaultUsing Setup Features Revising setup Reviewing current setupReceiving setup Reading setup Sending and Receiving Clock SettingsUsing Advanced Mode Features Using the manual analyze feature Using the manual charge feature M3860A only M3860A/M3861A Heartstart FR2+ Recording data on a data card Data Management and ReviewRecording Incident Data Recording data in internal memoryM3860A/M3861A Heartstart FR2+ Reviewing data from internal memory Reviewing Incident DataReviewing data from a data card Data Management and Review M3860A/M3861A Heartstart FR2+ HeartStart Accessories Accessories for the HeartStart FR2+Suggested Additional Items Technical Specifications Waveform Parameters Category Nominal specificationsEnergy dose ECG Analysis System Controls and indicators Accessories Specifications Optional M3848A FR2+ rechargeable battery Optional M3854A data card Optional M3873A/M3874A FR2+ ECG assessment M3860A/M3861A Heartstart FR2+ Instructions for Use Glossary of Symbols and ControlsSymboldescription Symbol DescriptionSymbol Accessories Symbols YYYY-MM M3863A standard battery M3848A FR2+ rechargeable battery M3849A charger For the M3848A FR2+ rechargeable battery Printed on recycled paper Electrical input Electrical output M3873A/M3874A FR2+ ECG assessment module Symboldescription M3855A charger For the M3864A training & administration pack Glossary of Terms Defibrillator pads with the HeartStart FR2+ See defibrillator pads Running HeartStart Event Review Web software Charger M3860A/M3861A Heartstart FR2+ Index ECG analysis system description D-2 Index M3860A/M3861A Heartstart FR2+ Index M3860A/M3861A Heartstart FR2+ Hilip s Me dical Sy ste ms Royal Philips Electronics Philips Medical Systems is part8 6 0 9 1 9 0 0