Philips M3861A, M3860A Using the manual analyze feature

Page 58

6-10

model used. For the M3860A, the user can select ANALYZE, CHARGE, or OFF. For the M3861A the user can select only ANALYZE or OFF. (Detailed directions for use are supplied with the training & administration pack.)

C A U T I O N : Alteration of the factory default setup of the FR2+ can affect its performance and should be performed under the authorization of your Medical Director. Modifications to device operation resulting from changes to the default settings should be specifically covered in user training.

C A U T I O N : The HeartStart FR2+ advanced mode’s MANUAL CHARGE feature is intended for use only by authorized operators who have been specifically trained in cardiac rhythm recognition and in defibrillation therapy using manual charge and shock delivery.

The ANALYZE feature is particularly useful for organizations that include responders who have Basic Life Support (BLS) training as well as more highly trained responders who may be certified in Advanced Life Support (ALS). In such situations, the Medical Director may set up a “tiered-response” system. The HeartStart FR2+ is specifically designed to provide different product features appropriate to each tier of responder.

In a scenario where a BLS responder is the first on the scene of an incident, he or she is trained to treat the patient immediately — for example, to check for breathing and responsiveness; to apply the defibrillator pads and connect them to the HeartStart FR2+; and to follow the voice and text prompts provided by the HeartStart FR2+ in its automated (AED) mode. When an ALS-trained responder arrives, the BLS responder “hands off” the patient’s care to the more highly trained responder.

Because these second-tier responders have advanced training and developed clinical skills, they may be authorized to access the advanced mode features of the HeartStart FR2+. These include user-initiated analysis and manual charge and disarm control.

Using the manual analyze feature

The Manual Analyze feature is available in both the M3860A and the

M3861A models, when enabled in setup.

To enter the advanced mode during use of an FR2+ that has this feature enabled, press both Option buttons simultaneously. This brings up a screen that includes a highlighted line at the bottom, labeled ANALYZE, with an arrowhead pointing to the lower Option button.

P hili ps Me dical Sy stems

M3860A/M3861A HEARTSTART FR2+

Image 58
Contents A R T S T a R T F R 2 + D E F I B R I L L a T O R S T R U C T I O N S F O R U S EHeartStart FR2+ Defibrillator I C K R E F E R E N C E C a R D Intentionally left blank A R T S T a R T F R 2 + M3860A, M3861A Defibrillator Device Tracking Contents Setup and Advanced Mode Features Clinical and Safety ConsiderationsData Management and Review Glossary of Symbols and Controls Accessories for the HeartStart FR2+Technical Specifications AppendicesM3860A/M3861A Heartstart FR2+ Introduction to the HeartStart FR2+ What is the HeartStart FR2+?Model M3860A Model M3861A When Is the HeartStart FR2+ Used? How Does the HeartStart FR2+ Work?How Is the HeartStart FR2+ Supplied? M3860A/M3861A Heartstart FR2+ Preparing Your HeartStart FR2+ for Use Installing the BatteryOverview M3860A/M3861A Heartstart FR2+ Setting the Clock Running the Battery Insertion Selftest Placing and Securing the HeartStart FR2+ Press the upper Option button to activate the testM3860A/M3861A Heartstart FR2+ Using Your HeartStart FR2+ Preparation ECG Analysis and Monitoring Shock Delivery ECG Display for Ongoing Observation M3860A/M3861A Heartstart FR2+ Maintenance After Using the HeartStart FR2+ Daily Monthly Maintenance task/responseCleaning the HeartStart FR2+ Operator’s checklistOperators Checklist Date Scheduled frequencyTesting Battery insertion selftestM3860A/M3861A Heartstart FR2+ Hili ps Me dical Sy stems Feature Test description TestPeriodic selftests Device historyBattery History Status indicator Meaning Troubleshooting GuideStatus indicator summary Recommended action during an emergencySymptom Possible cause Recommended action Status Indicator Flashing Black HourglassSymptom Possible cause General troubleshooting SymptomFR2+ has been physically damaged M3860A/M3861A Heartstart FR2+ Clinical Considerations Clinical and Safety ConsiderationsSafety Considerations IndicationsSafety level Possible shock or fire hazard, or explosion General dangers, warnings, and cautionsSafety level Possible improper device performance Safety levels Possible ECG misinterpretation Safety level Possible shock hazardDefibrillation warnings and cautions Maintenance cautions Monitoring cautionsSafety level Possible fire or shock hazard Parameter Settings Default Description Setup and Advanced Mode FeaturesSetup Overview Non-protocol parametersAutomatic protocol parameters NSA action Parameter Settings Default Manual override parametersUsing Setup Features Reviewing current setup Revising setup Receiving setup Sending and Receiving Clock Settings Reading setupUsing Advanced Mode Features Using the manual analyze feature Using the manual charge feature M3860A only M3860A/M3861A Heartstart FR2+ Recording data in internal memory Data Management and ReviewRecording Incident Data Recording data on a data cardM3860A/M3861A Heartstart FR2+ Reviewing Incident Data Reviewing data from internal memoryReviewing data from a data card Data Management and Review M3860A/M3861A Heartstart FR2+ Accessories for the HeartStart FR2+ HeartStart AccessoriesSuggested Additional Items Technical Specifications Category Nominal specifications Waveform ParametersEnergy dose ECG Analysis System Controls and indicators Accessories Specifications Optional M3848A FR2+ rechargeable battery Optional M3854A data card Optional M3873A/M3874A FR2+ ECG assessment M3860A/M3861A Heartstart FR2+ Symbol Description Glossary of Symbols and ControlsSymboldescription Instructions for UseSymbol Accessories Symbols YYYY-MM M3863A standard battery M3848A FR2+ rechargeable battery M3849A charger For the M3848A FR2+ rechargeable battery Printed on recycled paper Electrical input Electrical output M3873A/M3874A FR2+ ECG assessment module Symboldescription M3855A charger For the M3864A training & administration pack Glossary of Terms Defibrillator pads with the HeartStart FR2+ See defibrillator pads Running HeartStart Event Review Web software Charger M3860A/M3861A Heartstart FR2+ Index ECG analysis system description D-2 Index M3860A/M3861A Heartstart FR2+ Index M3860A/M3861A Heartstart FR2+ Hilip s Me dical Sy ste ms Royal Philips Electronics Philips Medical Systems is part8 6 0 9 1 9 0 0