Philips M3860A, M3861A manual Energy dose

Page 71

P hili ps Me dical Sy stems

 

 

 

 

B-3

B

 

 

 

 

 

 

 

 

 

category

 

nominal specifications

 

 

 

 

Energy

Using adult defibrillator pads: 150 J nominal into a 50 ohm load.

 

 

Using infant/child reduced-energy defibrillator pads: 50 J nominal into a 50 ohm

 

 

load. Sample pediatric energy doses:

 

 

 

 

 

age

energy dose

 

 

 

 

 

 

 

 

 

newborn

14 J/kg

 

 

 

1 year

5 J/kg

 

 

 

2 − 3 years

4 J/kg

 

 

 

4 − 5 years

3 J/kg

 

 

 

6 − 8 years

2 J/kg

 

 

Doses indicated are based on CDC growth charts for the 50th percentile weights

 

 

for boys.*

 

 

 

*National Center for Health Statistics in collaboration with the National Center for Chronic Disease Prevention and Health Promotion. CDC growth charts: weight-for-age percentiles, revised and corrected November 28, 2000. Atlanta, GA: Centers for Disease Control and Prevention © 2000.

Charge Control

Controlled by Patient Analysis System for automated operation. Can be

 

programmed for manual initiation using advanced mode of the M3860A.

 

 

Charge Time

< 10 seconds typical, including confirming analysis. Charge time increases near

from “Shock Advised”

end of battery service life.

 

 

Shock-to-Shock Cycle Time

< 20 seconds typical, including analysis, in AED mode.

 

 

“Charge Complete”

Shock button flashes, audio tone sounds.

Indicator

 

 

 

Disarm (AED mode)

Once charged, the HeartStart FR2+ will disarm if:

 

• patient’s heart rhythm changes to non-shockable rhythm, OR

 

• a shock is not delivered within 30 seconds after the FR2+ is armed, OR

 

• the PAUSE button (if enabled) is pressed, OR

 

• the On/Off button is pressed to turn off the FR2+, OR

• the defibrillator pads are removed from the patient or the pads connector is disconnected from the FR2+.

Technical Specifications

Image 71
Contents S T R U C T I O N S F O R U S E A R T S T a R T F R 2 + D E F I B R I L L a T O RHeartStart FR2+ Defibrillator I C K R E F E R E N C E C a R D Intentionally left blank A R T S T a R T F R 2 + M3860A, M3861A Defibrillator Device Tracking Contents Data Management and Review Clinical and Safety ConsiderationsSetup and Advanced Mode Features Appendices Accessories for the HeartStart FR2+Technical Specifications Glossary of Symbols and ControlsM3860A/M3861A Heartstart FR2+ Model M3860A Model M3861A What is the HeartStart FR2+?Introduction to the HeartStart FR2+ How Does the HeartStart FR2+ Work? When Is the HeartStart FR2+ Used?How Is the HeartStart FR2+ Supplied? M3860A/M3861A Heartstart FR2+ Overview Installing the BatteryPreparing Your HeartStart FR2+ for Use M3860A/M3861A Heartstart FR2+ Setting the Clock Running the Battery Insertion Selftest Press the upper Option button to activate the test Placing and Securing the HeartStart FR2+M3860A/M3861A Heartstart FR2+ Using Your HeartStart FR2+ Preparation ECG Analysis and Monitoring Shock Delivery ECG Display for Ongoing Observation M3860A/M3861A Heartstart FR2+ Maintenance Daily Monthly Maintenance task/response After Using the HeartStart FR2+Operator’s checklist Cleaning the HeartStart FR2+Date Scheduled frequency Operators ChecklistBattery insertion selftest TestingM3860A/M3861A Heartstart FR2+ Hili ps Me dical Sy stems Test Feature Test descriptionDevice history Periodic selftestsBattery History Recommended action during an emergency Troubleshooting GuideStatus indicator summary Status indicator MeaningStatus Indicator Flashing Black Hourglass Symptom Possible cause Recommended actionSymptom Possible cause Symptom General troubleshootingFR2+ has been physically damaged M3860A/M3861A Heartstart FR2+ Indications Clinical and Safety ConsiderationsSafety Considerations Clinical ConsiderationsGeneral dangers, warnings, and cautions Safety level Possible shock or fire hazard, or explosionSafety level Possible improper device performance Defibrillation warnings and cautions Safety level Possible shock hazardSafety levels Possible ECG misinterpretation Monitoring cautions Maintenance cautionsSafety level Possible fire or shock hazard Non-protocol parameters Setup and Advanced Mode FeaturesSetup Overview Parameter Settings Default DescriptionAutomatic protocol parameters NSA action Manual override parameters Parameter Settings DefaultUsing Setup Features Revising setup Reviewing current setupReceiving setup Reading setup Sending and Receiving Clock SettingsUsing Advanced Mode Features Using the manual analyze feature Using the manual charge feature M3860A only M3860A/M3861A Heartstart FR2+ Recording data on a data card Data Management and ReviewRecording Incident Data Recording data in internal memoryM3860A/M3861A Heartstart FR2+ Reviewing data from internal memory Reviewing Incident DataReviewing data from a data card Data Management and Review M3860A/M3861A Heartstart FR2+ HeartStart Accessories Accessories for the HeartStart FR2+Suggested Additional Items Technical Specifications Waveform Parameters Category Nominal specificationsEnergy dose ECG Analysis System Controls and indicators Accessories Specifications Optional M3848A FR2+ rechargeable battery Optional M3854A data card Optional M3873A/M3874A FR2+ ECG assessment M3860A/M3861A Heartstart FR2+ Instructions for Use Glossary of Symbols and ControlsSymboldescription Symbol DescriptionSymbol Accessories Symbols YYYY-MM M3863A standard battery M3848A FR2+ rechargeable battery M3849A charger For the M3848A FR2+ rechargeable battery Printed on recycled paper Electrical input Electrical output M3873A/M3874A FR2+ ECG assessment module Symboldescription M3855A charger For the M3864A training & administration pack Glossary of Terms Defibrillator pads with the HeartStart FR2+ See defibrillator pads Running HeartStart Event Review Web software Charger M3860A/M3861A Heartstart FR2+ Index ECG analysis system description D-2 Index M3860A/M3861A Heartstart FR2+ Index M3860A/M3861A Heartstart FR2+ Hilip s Me dical Sy ste ms 8 6 0 9 1 9 0 0 Philips Medical Systems is partRoyal Philips Electronics