2Preparing Your HeartStart FR2+ for Use
2
Overview
There are a few basic steps to preparing your HeartStart FR2+ Defibrillator for use:
Install data card (optional).
Install a battery.
Set the clock in the FR2+ (optional).
Run the battery insertion selftest.
Place the FR2+ with recommended accessories in a convenient location.
The instructions presented here briefly describe the normal sequence of preparation. It assumes that you are using a fresh battery, that the selftest passes, that you are not using a data card, and that the factory default settings will not be changed. Exceptions to this sequence are provided elsewhere in this manual.
Installing the Battery
The HeartStart FR2+ Defibrillator is shipped with a M3863A standard,
Before installing the battery, make sure the defibrillator pads are not connected to the FR2+. To install the battery:
1.Hold the battery by the latch end and slide it into the battery compartment at the top of the FR2+.
2.Slide the battery all the way into the opening, until the latch clicks into place. The latch will click into place only when the battery is inserted correctly.
C A U T I O N : Follow all instructions supplied with the HeartStart M3863A standard battery. Install the battery before the