Philips M3860A, M3861A Preparing Your HeartStart FR2+ for Use, Overview, Installing the Battery

Page 15

P hili ps Me dical Sy stems

2Preparing Your HeartStart FR2+ for Use

2

Overview

There are a few basic steps to preparing your HeartStart FR2+ Defibrillator for use:

￿Install data card (optional).

￿Install a battery.

￿Set the clock in the FR2+ (optional).

￿Run the battery insertion selftest.

￿Place the FR2+ with recommended accessories in a convenient location.

The instructions presented here briefly describe the normal sequence of preparation. It assumes that you are using a fresh battery, that the selftest passes, that you are not using a data card, and that the factory default settings will not be changed. Exceptions to this sequence are provided elsewhere in this manual.

Installing the Battery

The HeartStart FR2+ Defibrillator is shipped with a M3863A standard, long-life battery. The battery is enclosed in a gray plastic case. There is a yellow latch at one end that holds the battery in place when it is correctly installed in the FR2+. (The optional M3848A FR2+ rechargeable battery is enclosed in a blue plastic case and also has a yellow latch. Except where otherwise noted, the following information applies to both battery types.)

Before installing the battery, make sure the defibrillator pads are not connected to the FR2+. To install the battery:

1.Hold the battery by the latch end and slide it into the battery compartment at the top of the FR2+.

2.Slide the battery all the way into the opening, until the latch clicks into place. The latch will click into place only when the battery is inserted correctly.

C A U T I O N : Follow all instructions supplied with the HeartStart M3863A standard battery. Install the battery before the install-by date shown on the battery.

2-1

Image 15
Contents S T R U C T I O N S F O R U S E A R T S T a R T F R 2 + D E F I B R I L L a T O RHeartStart FR2+ Defibrillator I C K R E F E R E N C E C a R D Intentionally left blank A R T S T a R T F R 2 + M3860A, M3861A Defibrillator Device Tracking Contents Clinical and Safety Considerations Setup and Advanced Mode FeaturesData Management and Review Appendices Accessories for the HeartStart FR2+Technical Specifications Glossary of Symbols and ControlsM3860A/M3861A Heartstart FR2+ What is the HeartStart FR2+? Introduction to the HeartStart FR2+Model M3860A Model M3861A How Does the HeartStart FR2+ Work? When Is the HeartStart FR2+ Used?How Is the HeartStart FR2+ Supplied? M3860A/M3861A Heartstart FR2+ Installing the Battery Preparing Your HeartStart FR2+ for UseOverview M3860A/M3861A Heartstart FR2+ Setting the Clock Running the Battery Insertion Selftest Press the upper Option button to activate the test Placing and Securing the HeartStart FR2+M3860A/M3861A Heartstart FR2+ Using Your HeartStart FR2+ Preparation ECG Analysis and Monitoring Shock Delivery ECG Display for Ongoing Observation M3860A/M3861A Heartstart FR2+ Maintenance Daily Monthly Maintenance task/response After Using the HeartStart FR2+Operator’s checklist Cleaning the HeartStart FR2+Date Scheduled frequency Operators ChecklistBattery insertion selftest TestingM3860A/M3861A Heartstart FR2+ Hili ps Me dical Sy stems Test Feature Test descriptionDevice history Periodic selftestsBattery History Recommended action during an emergency Troubleshooting GuideStatus indicator summary Status indicator MeaningStatus Indicator Flashing Black Hourglass Symptom Possible cause Recommended actionSymptom Possible cause Symptom General troubleshootingFR2+ has been physically damaged M3860A/M3861A Heartstart FR2+ Indications Clinical and Safety ConsiderationsSafety Considerations Clinical ConsiderationsGeneral dangers, warnings, and cautions Safety level Possible shock or fire hazard, or explosionSafety level Possible improper device performance Safety level Possible shock hazard Safety levels Possible ECG misinterpretationDefibrillation warnings and cautions Monitoring cautions Maintenance cautionsSafety level Possible fire or shock hazard Non-protocol parameters Setup and Advanced Mode FeaturesSetup Overview Parameter Settings Default DescriptionAutomatic protocol parameters NSA action Manual override parameters Parameter Settings DefaultUsing Setup Features Revising setup Reviewing current setupReceiving setup Reading setup Sending and Receiving Clock SettingsUsing Advanced Mode Features Using the manual analyze feature Using the manual charge feature M3860A only M3860A/M3861A Heartstart FR2+ Recording data on a data card Data Management and ReviewRecording Incident Data Recording data in internal memoryM3860A/M3861A Heartstart FR2+ Reviewing data from internal memory Reviewing Incident DataReviewing data from a data card Data Management and Review M3860A/M3861A Heartstart FR2+ HeartStart Accessories Accessories for the HeartStart FR2+Suggested Additional Items Technical Specifications Waveform Parameters Category Nominal specificationsEnergy dose ECG Analysis System Controls and indicators Accessories Specifications Optional M3848A FR2+ rechargeable battery Optional M3854A data card Optional M3873A/M3874A FR2+ ECG assessment M3860A/M3861A Heartstart FR2+ Instructions for Use Glossary of Symbols and ControlsSymboldescription Symbol DescriptionSymbol Accessories Symbols YYYY-MM M3863A standard battery M3848A FR2+ rechargeable battery M3849A charger For the M3848A FR2+ rechargeable battery Printed on recycled paper Electrical input Electrical output M3873A/M3874A FR2+ ECG assessment module Symboldescription M3855A charger For the M3864A training & administration pack Glossary of Terms Defibrillator pads with the HeartStart FR2+ See defibrillator pads Running HeartStart Event Review Web software Charger M3860A/M3861A Heartstart FR2+ Index ECG analysis system description D-2 Index M3860A/M3861A Heartstart FR2+ Index M3860A/M3861A Heartstart FR2+ Hilip s Me dical Sy ste ms Philips Medical Systems is part Royal Philips Electronics8 6 0 9 1 9 0 0