6 | Setup and Advanced Mode Features |
| ||||
|
| Setup Overview |
| |||
|
| The “setup” of the HeartStart FR2+ Defibrillator is made up of several |
| |||
|
| programmable aspects, or parameters, of FR2+ operation. Some setup |
| |||
|
| parameters govern specific features that are not related to the patient care |
| |||
|
| protocol, some are used to define the automatic patient care protocol used |
| |||
|
| by the FR2+, and some provide options for manual override of the protocol |
| |||
|
| during use. |
|
| ||
|
| The FR2+ comes with a factory default setup designed to meet the needs of |
| |||
|
| most users. If desired, your Medical Director can revise the setup. Even if no |
| |||
|
| changes are made, however, it is a good idea to understand the setup of your |
| |||
|
| FR2+ and how the different parameter settings affect the way the device |
| |||
|
| works. |
|
|
| |
|
|
|
| |||
|
| The parameters listed in the following table enable features of FR2+ |
| |||
|
| operation that are not related to the patient treatment protocol. The table |
| |||
|
| describes each of these |
| |||
|
| for it, and identifies the default setting. |
| |||
|
|
|
|
|
|
|
parameter | settings | default | description |
| ||
|
|
|
|
|
|
|
speaker volume |
| 1, 2, 3, 4, |
| 8 | Sets volume of the FR2+’s speaker. 1 is lowest; 8 highest. |
|
|
| 5, 6, 7, 8 |
|
| The speaker is used for voice prompts and the | 6 |
|
|
|
|
| ||
|
|
|
|
|
| |
|
|
|
|
|
|
|
record voice |
| YES, NO |
| NO | Enables or disables the audio recording during incident. |
|
|
|
|
|
| Voice recording requires use of a data card. |
|
|
|
|
|
|
|
|
ECG display |
| ON, OFF |
| ON | Enables (ON) or disables (OFF) ECG display on the |
|
(M3860A only) |
|
|
|
| screen. |
|
|
|
|
|
| FR2+ rhythm analysis does not require ECG display to be |
|
|
|
|
|
| on. (ECG display cannot be changed from default OFF |
|
|
|
|
|
| for M3861A.) |
|
|
|
|
|
|
|
|
autosend PST |
| ON, OFF |
| OFF | Enables (ON) or disables (OFF) transmission of the |
|
|
|
|
|
| results of the FR2+’s periodic selftests (PST) from its |
|
|
|
|
|
| infrared communications port. |
|
|
|
|
|
|
|
|