Philips M3860A, M3861A manual Optional M3848A FR2+ rechargeable battery

Page 75

P hili ps Me dical Sy stems

 

B-7

B

 

(Optional) M3848A FR2+ rechargeable battery

 

 

 

 

 

category

nominal specifications

 

 

 

 

Battery Type

11.3 VDC, 6.5 Ah, lithium ion. Rechargeable cell using the M3849A charger.

 

 

 

 

Capacity

When freshly charged and used at 77° F (25° C) , provides a minimum of 80

 

 

(typically 100) shocks, or 3.5 hours (typically 5 hours) of ECG display time only,

 

 

before recharging is indicated.

 

 

 

 

Status Indicators

Good battery: bar graph on display screen indicating remaining power level.

 

 

Low battery: flashing red X on the front panel of the FR2+ (When low battery

 

 

indicator appears, there is still enough energy to deliver 9 shocks plus 15 minutes

 

 

of ECG display time).

 

 

Dead battery: solid red X on the front panel of the FR2+.

 

 

 

 

Storage Temperature

32° to 109° F (0° to 43° C).

 

 

 

 

Standby Life

6 months when installed fully charged in a defibrillator labeled FR2+.

 

(after installation)

 

 

 

(Optional) M3849A charger

 

 

 

 

 

 

category

nominal specifications

 

 

 

 

Application

For use with M3848A FR2+ rechargeable battery only.

 

 

 

 

Power Requirements

100 to 240 VAC, 47 to 63 Hz, 30 Watts

 

 

 

 

Environmental

32° to 122° F (0° to 50° C).

 

Requirements

0% to 95% relative humidity (non-condensing).

 

Conformance Testing

International: EN60335-1:1994 Class 1

 

 

North America: UL 1310 Class 2

 

 

M3870A and DP2/DP6 defibrillator pads

 

 

 

 

 

 

category

nominal specifications

 

 

 

 

Pads, Cable, and Connector Disposable and self-adhesive. DP2/DP6 adult defibrillator pads have a nominal active surface area of 100 cm2 each and are provided in a sealed package with an integrated 122 cm (48 inch), typical, cable and connector. M3870A FR2 infant/child reduced-energy defibrillator pads have a nominal active surface area of 44 cm2 each and are provided in a sealed package with an integrated 122 cm (48 inch), typical, cable and connector incorporating attenuating electronics.

Technical Specifications

Image 75
Contents S T R U C T I O N S F O R U S E A R T S T a R T F R 2 + D E F I B R I L L a T O RHeartStart FR2+ Defibrillator I C K R E F E R E N C E C a R D Intentionally left blank A R T S T a R T F R 2 + M3860A, M3861A Defibrillator Device Tracking Contents Clinical and Safety Considerations Setup and Advanced Mode FeaturesData Management and Review Appendices Accessories for the HeartStart FR2+Technical Specifications Glossary of Symbols and ControlsM3860A/M3861A Heartstart FR2+ What is the HeartStart FR2+? Introduction to the HeartStart FR2+Model M3860A Model M3861A How Does the HeartStart FR2+ Work? When Is the HeartStart FR2+ Used?How Is the HeartStart FR2+ Supplied? M3860A/M3861A Heartstart FR2+ Installing the Battery Preparing Your HeartStart FR2+ for UseOverview M3860A/M3861A Heartstart FR2+ Setting the Clock Running the Battery Insertion Selftest Press the upper Option button to activate the test Placing and Securing the HeartStart FR2+M3860A/M3861A Heartstart FR2+ Using Your HeartStart FR2+ Preparation ECG Analysis and Monitoring Shock Delivery ECG Display for Ongoing Observation M3860A/M3861A Heartstart FR2+ Maintenance Daily Monthly Maintenance task/response After Using the HeartStart FR2+Operator’s checklist Cleaning the HeartStart FR2+Date Scheduled frequency Operators ChecklistBattery insertion selftest TestingM3860A/M3861A Heartstart FR2+ Hili ps Me dical Sy stems Test Feature Test descriptionDevice history Periodic selftestsBattery History Recommended action during an emergency Troubleshooting GuideStatus indicator summary Status indicator MeaningStatus Indicator Flashing Black Hourglass Symptom Possible cause Recommended actionSymptom Possible cause Symptom General troubleshootingFR2+ has been physically damaged M3860A/M3861A Heartstart FR2+ Indications Clinical and Safety ConsiderationsSafety Considerations Clinical ConsiderationsGeneral dangers, warnings, and cautions Safety level Possible shock or fire hazard, or explosionSafety level Possible improper device performance Safety level Possible shock hazard Safety levels Possible ECG misinterpretationDefibrillation warnings and cautions Monitoring cautions Maintenance cautionsSafety level Possible fire or shock hazard Non-protocol parameters Setup and Advanced Mode FeaturesSetup Overview Parameter Settings Default DescriptionAutomatic protocol parameters NSA action Manual override parameters Parameter Settings DefaultUsing Setup Features Revising setup Reviewing current setupReceiving setup Reading setup Sending and Receiving Clock SettingsUsing Advanced Mode Features Using the manual analyze feature Using the manual charge feature M3860A only M3860A/M3861A Heartstart FR2+ Recording data on a data card Data Management and ReviewRecording Incident Data Recording data in internal memoryM3860A/M3861A Heartstart FR2+ Reviewing data from internal memory Reviewing Incident DataReviewing data from a data card Data Management and Review M3860A/M3861A Heartstart FR2+ HeartStart Accessories Accessories for the HeartStart FR2+Suggested Additional Items Technical Specifications Waveform Parameters Category Nominal specificationsEnergy dose ECG Analysis System Controls and indicators Accessories Specifications Optional M3848A FR2+ rechargeable battery Optional M3854A data card Optional M3873A/M3874A FR2+ ECG assessment M3860A/M3861A Heartstart FR2+ Instructions for Use Glossary of Symbols and ControlsSymboldescription Symbol DescriptionSymbol Accessories Symbols YYYY-MM M3863A standard battery M3848A FR2+ rechargeable battery M3849A charger For the M3848A FR2+ rechargeable battery Printed on recycled paper Electrical input Electrical output M3873A/M3874A FR2+ ECG assessment module Symboldescription M3855A charger For the M3864A training & administration pack Glossary of Terms Defibrillator pads with the HeartStart FR2+ See defibrillator pads Running HeartStart Event Review Web software Charger M3860A/M3861A Heartstart FR2+ Index ECG analysis system description D-2 Index M3860A/M3861A Heartstart FR2+ Index M3860A/M3861A Heartstart FR2+ Hilip s Me dical Sy ste ms Philips Medical Systems is part Royal Philips Electronics8 6 0 9 1 9 0 0