safety levels

possible burns and ineffective energy

 

 

 

 

WARNING

Do not allow the defibrillator pads to touch each other or other ECG

 

 

electrodes, lead wires, dressings, transdermal patches, etc. Such

 

 

contact can cause electrical arcing and patient skin burns during

 

 

defibrillation and may also divert the defibrillation current away from

 

 

the heart.

 

 

 

 

WARNING

During defibrillation, air pockets between the skin and defibrillator

 

pads can cause patient skin burns. To help prevent air pockets, make

 

sure defibrillator pads completely adhere to the skin. Do not use

 

dried out defibrillator pads.

 

 

 

 

 

 

safety level

possible patient injury

 

 

 

 

CAUTION

The HeartStart FR2+ advanced mode’s MANUAL CHARGE feature is

 

 

intended for use only by authorized operators who have been specifically

 

 

trained in cardiac rhythm recognition and in defibrillation therapy using

 

 

manual charge and shock delivery.

 

 

 

 

 

Monitoring cautions

 

5

 

 

 

safety level

possible misinterpretation of ECG recordings

 

 

 

 

CAUTION

The LCD screen on the HeartStart FR2+ model M3860A is intended only

 

 

for basic ECG rhythm identification. The frequency response of the

 

 

monitor screen is not intended to provide the resolution needed for

 

 

diagnostic and ST segment interpretation.

 

 

 

 

 

Maintenance cautions

 

 

safety level

possible fire or shock hazard

 

 

 

 

CAUTION

Electrical shock hazard. Dangerous high voltages and currents are

 

 

present. Do not open the HeartStart FR2+, remove its covers, or attempt

 

 

repair. There are no user-serviceable components in the HeartStart FR2+.

 

 

The HeartStart FR2+ should be returned to an authorized service center

 

 

for repair.

 

 

 

 

Clinical and Safety Considerations

Image 47
Contents S T R U C T I O N S F O R U S E A R T S T a R T F R 2 + D E F I B R I L L a T O RHeartStart FR2+ Defibrillator I C K R E F E R E N C E C a R D Intentionally left blank A R T S T a R T F R 2 + M3860A, M3861A Defibrillator Device Tracking Contents Data Management and Review Clinical and Safety ConsiderationsSetup and Advanced Mode Features Appendices Accessories for the HeartStart FR2+Technical Specifications Glossary of Symbols and ControlsM3860A/M3861A Heartstart FR2+ Model M3860A Model M3861A What is the HeartStart FR2+?Introduction to the HeartStart FR2+ How Does the HeartStart FR2+ Work? When Is the HeartStart FR2+ Used?How Is the HeartStart FR2+ Supplied? M3860A/M3861A Heartstart FR2+ Overview Installing the BatteryPreparing Your HeartStart FR2+ for Use M3860A/M3861A Heartstart FR2+ Setting the Clock Running the Battery Insertion Selftest Press the upper Option button to activate the test Placing and Securing the HeartStart FR2+M3860A/M3861A Heartstart FR2+ Using Your HeartStart FR2+ Preparation ECG Analysis and Monitoring Shock Delivery ECG Display for Ongoing Observation M3860A/M3861A Heartstart FR2+ Maintenance Daily Monthly Maintenance task/response After Using the HeartStart FR2+Operator’s checklist Cleaning the HeartStart FR2+Date Scheduled frequency Operators ChecklistBattery insertion selftest TestingM3860A/M3861A Heartstart FR2+ Hili ps Me dical Sy stems Test Feature Test descriptionDevice history Periodic selftestsBattery History Recommended action during an emergency Troubleshooting GuideStatus indicator summary Status indicator MeaningStatus Indicator Flashing Black Hourglass Symptom Possible cause Recommended actionSymptom Possible cause Symptom General troubleshootingFR2+ has been physically damaged M3860A/M3861A Heartstart FR2+ Indications Clinical and Safety ConsiderationsSafety Considerations Clinical ConsiderationsGeneral dangers, warnings, and cautions Safety level Possible shock or fire hazard, or explosionSafety level Possible improper device performance Defibrillation warnings and cautions Safety level Possible shock hazardSafety levels Possible ECG misinterpretation Monitoring cautions Maintenance cautionsSafety level Possible fire or shock hazard Non-protocol parameters Setup and Advanced Mode FeaturesSetup Overview Parameter Settings Default DescriptionAutomatic protocol parameters NSA action Manual override parameters Parameter Settings DefaultUsing Setup Features Revising setup Reviewing current setupReceiving setup Reading setup Sending and Receiving Clock SettingsUsing Advanced Mode Features Using the manual analyze feature Using the manual charge feature M3860A only M3860A/M3861A Heartstart FR2+ Recording data on a data card Data Management and ReviewRecording Incident Data Recording data in internal memoryM3860A/M3861A Heartstart FR2+ Reviewing data from internal memory Reviewing Incident DataReviewing data from a data card Data Management and Review M3860A/M3861A Heartstart FR2+ HeartStart Accessories Accessories for the HeartStart FR2+Suggested Additional Items Technical Specifications Waveform Parameters Category Nominal specificationsEnergy dose ECG Analysis System Controls and indicators Accessories Specifications Optional M3848A FR2+ rechargeable battery Optional M3854A data card Optional M3873A/M3874A FR2+ ECG assessment M3860A/M3861A Heartstart FR2+ Instructions for Use Glossary of Symbols and ControlsSymboldescription Symbol DescriptionSymbol Accessories Symbols YYYY-MM M3863A standard battery M3848A FR2+ rechargeable battery M3849A charger For the M3848A FR2+ rechargeable battery Printed on recycled paper Electrical input Electrical output M3873A/M3874A FR2+ ECG assessment module Symboldescription M3855A charger For the M3864A training & administration pack Glossary of Terms Defibrillator pads with the HeartStart FR2+ See defibrillator pads Running HeartStart Event Review Web software Charger M3860A/M3861A Heartstart FR2+ Index ECG analysis system description D-2 Index M3860A/M3861A Heartstart FR2+ Index M3860A/M3861A Heartstart FR2+ Hilip s Me dical Sy ste ms 8 6 0 9 1 9 0 0 Philips Medical Systems is partRoyal Philips Electronics