Philips M3861A, M3860A manual Sending and Receiving Clock Settings, Reading setup

Page 56

6-8

“receiving” FR2+ displays a SETUP COMPLETE message. Your HeartStart FR2+ immediately uses the new setup.

Receiving setup from a computer running HeartStart Event Review software is discussed in the directions for use provided with HeartStart Event Review software.

Reading setup

This method copies setup data from a data card to your HeartStart FR2+. To read the setup, follow these steps:

1. Insert the data card in the data card tray and install the loaded tray into the data card slot in the FR2+.

2. Select READ SETUP from the setup menu.

3.A new screen comes up. If the FR2+ cannot read the data card or

cannot find a valid setup on the data card, the screen displays a NO SETUP FILE error message. Otherwise, the FR2+ begins reading the setup information from the data card immediately.

4. If you select EXIT before the transfer is complete, the revised setup will

not be copied. When the transfer is finished, the screen displays a SETUP COMPLETE message. Your FR2+ immediately uses the revised setup.

Sending and Receiving Clock Settings

To synchronize the clock settings of your HeartStart FR2+ with the clock of another FR2+ or a computer running HeartStart Event Review software, you can use the infrared communications feature.

Instructions for synchronizing clock settings using a computer running HeartStart Event Review are provided with the HeartStart Event Review software.

To transfer clock settings from one FR2+ to another:

1. Remove and reinstall the battery of both FR2+ devices to bring up the first menu screen.

2. Select NEXT to go to the second menu screen.

3. Select CLOCK from the second menu screen. The CLOCK screen then comes up.

4. Locate the infrared communications port on each FR2+ and line them up with one another, so that the infrared “eye” in each one has an

P hili ps Me dical Sy stems

M3860A/M3861A HEARTSTART FR2+

Image 56
Contents A R T S T a R T F R 2 + D E F I B R I L L a T O R S T R U C T I O N S F O R U S EHeartStart FR2+ Defibrillator I C K R E F E R E N C E C a R D Intentionally left blank A R T S T a R T F R 2 + M3860A, M3861A Defibrillator Device Tracking Contents Data Management and Review Clinical and Safety ConsiderationsSetup and Advanced Mode Features Accessories for the HeartStart FR2+ Technical SpecificationsGlossary of Symbols and Controls AppendicesM3860A/M3861A Heartstart FR2+ Model M3860A Model M3861A What is the HeartStart FR2+?Introduction to the HeartStart FR2+ When Is the HeartStart FR2+ Used? How Does the HeartStart FR2+ Work?How Is the HeartStart FR2+ Supplied? M3860A/M3861A Heartstart FR2+ Overview Installing the BatteryPreparing Your HeartStart FR2+ for Use M3860A/M3861A Heartstart FR2+ Setting the Clock Running the Battery Insertion Selftest Placing and Securing the HeartStart FR2+ Press the upper Option button to activate the testM3860A/M3861A Heartstart FR2+ Using Your HeartStart FR2+ Preparation ECG Analysis and Monitoring Shock Delivery ECG Display for Ongoing Observation M3860A/M3861A Heartstart FR2+ Maintenance After Using the HeartStart FR2+ Daily Monthly Maintenance task/responseCleaning the HeartStart FR2+ Operator’s checklistOperators Checklist Date Scheduled frequencyTesting Battery insertion selftestM3860A/M3861A Heartstart FR2+ Hili ps Me dical Sy stems Feature Test description TestPeriodic selftests Device historyBattery History Troubleshooting Guide Status indicator summaryStatus indicator Meaning Recommended action during an emergencySymptom Possible cause Recommended action Status Indicator Flashing Black HourglassSymptom Possible cause General troubleshooting SymptomFR2+ has been physically damaged M3860A/M3861A Heartstart FR2+ Clinical and Safety Considerations Safety ConsiderationsClinical Considerations IndicationsSafety level Possible shock or fire hazard, or explosion General dangers, warnings, and cautionsSafety level Possible improper device performance Defibrillation warnings and cautions Safety level Possible shock hazardSafety levels Possible ECG misinterpretation Maintenance cautions Monitoring cautionsSafety level Possible fire or shock hazard Setup and Advanced Mode Features Setup OverviewParameter Settings Default Description Non-protocol parametersAutomatic protocol parameters NSA action Parameter Settings Default Manual override parametersUsing Setup Features Reviewing current setup Revising setupReceiving setup Sending and Receiving Clock Settings Reading setupUsing Advanced Mode Features Using the manual analyze feature Using the manual charge feature M3860A only M3860A/M3861A Heartstart FR2+ Data Management and Review Recording Incident DataRecording data in internal memory Recording data on a data cardM3860A/M3861A Heartstart FR2+ Reviewing Incident Data Reviewing data from internal memoryReviewing data from a data card Data Management and Review M3860A/M3861A Heartstart FR2+ Accessories for the HeartStart FR2+ HeartStart AccessoriesSuggested Additional Items Technical Specifications Category Nominal specifications Waveform ParametersEnergy dose ECG Analysis System Controls and indicators Accessories Specifications Optional M3848A FR2+ rechargeable battery Optional M3854A data card Optional M3873A/M3874A FR2+ ECG assessment M3860A/M3861A Heartstart FR2+ Glossary of Symbols and Controls SymboldescriptionSymbol Description Instructions for UseSymbol Accessories Symbols YYYY-MM M3863A standard battery M3848A FR2+ rechargeable battery M3849A charger For the M3848A FR2+ rechargeable battery Printed on recycled paper Electrical input Electrical output M3873A/M3874A FR2+ ECG assessment module Symboldescription M3855A charger For the M3864A training & administration pack Glossary of Terms Defibrillator pads with the HeartStart FR2+ See defibrillator pads Running HeartStart Event Review Web software Charger M3860A/M3861A Heartstart FR2+ Index ECG analysis system description D-2 Index M3860A/M3861A Heartstart FR2+ Index M3860A/M3861A Heartstart FR2+ Hilip s Me dical Sy ste ms 8 6 0 9 1 9 0 0 Philips Medical Systems is partRoyal Philips Electronics