The following table describes the parts of the FR2+ tested in the interactive

 

part of the selftest and any action you are asked to take.

feature

test description

speaker

Screen prompt: CHECK SPEAKER SOUND (2 beeps)

 

> Listen for the two beeps from the speaker.

lights

Screen prompt: CHECK SHOCK BUTTON LIGHT AND PADS CONNECTOR LIGHT

 

> Check that the lights come on.

option buttons

Screen prompt: PRESS THE OPTION BUTTONS

 

 

 

> Press the blue upper and lower Option buttons and listen for a beep to

 

 

 

confirm each press. Look at the screen to be sure the button presses

 

 

 

have been verified.

 

 

 

 

 

 

 

 

 

 

 

display screen

Screen prompt: CHECK DISPLAY. ADJUST CONTRAST IF NEEDED

 

> Check the test pattern displayed on the screen. Adjust the contrast if

 

desired using the Option buttons.

 

N O T E : Screen contrast can be adjusted at any time during the interactive

 

selftest by repeatedly pressing the appropriate Option button until

 

desired contrast is achieved.

shock button

Screen prompt: PRESS THE SHOCK BUTTON

 

> Press the Shock button and listen for a beep to confirm the press.

 

No shock will be delivered when you press the Shock button during the

 

test.

 

> Look at the screen to be sure the button press has been verified.

P hili ps Me dical Sy stems

M3860A/M3861A HEARTSTART FR2+

Image 34
Contents A R T S T a R T F R 2 + D E F I B R I L L a T O R S T R U C T I O N S F O R U S EHeartStart FR2+ Defibrillator I C K R E F E R E N C E C a R D Intentionally left blank A R T S T a R T F R 2 + M3860A, M3861A Defibrillator Device Tracking Contents Setup and Advanced Mode Features Clinical and Safety ConsiderationsData Management and Review Glossary of Symbols and Controls Accessories for the HeartStart FR2+Technical Specifications AppendicesM3860A/M3861A Heartstart FR2+ Introduction to the HeartStart FR2+ What is the HeartStart FR2+?Model M3860A Model M3861A When Is the HeartStart FR2+ Used? How Does the HeartStart FR2+ Work?How Is the HeartStart FR2+ Supplied? M3860A/M3861A Heartstart FR2+ Preparing Your HeartStart FR2+ for Use Installing the BatteryOverview M3860A/M3861A Heartstart FR2+ Setting the Clock Running the Battery Insertion Selftest Placing and Securing the HeartStart FR2+ Press the upper Option button to activate the testM3860A/M3861A Heartstart FR2+ Using Your HeartStart FR2+ Preparation ECG Analysis and Monitoring Shock Delivery ECG Display for Ongoing Observation M3860A/M3861A Heartstart FR2+ Maintenance After Using the HeartStart FR2+ Daily Monthly Maintenance task/responseCleaning the HeartStart FR2+ Operator’s checklistOperators Checklist Date Scheduled frequencyTesting Battery insertion selftestM3860A/M3861A Heartstart FR2+ Hili ps Me dical Sy stems Feature Test description TestPeriodic selftests Device historyBattery History Status indicator Meaning Troubleshooting GuideStatus indicator summary Recommended action during an emergencySymptom Possible cause Recommended action Status Indicator Flashing Black HourglassSymptom Possible cause General troubleshooting SymptomFR2+ has been physically damaged M3860A/M3861A Heartstart FR2+ Clinical Considerations Clinical and Safety ConsiderationsSafety Considerations IndicationsSafety level Possible shock or fire hazard, or explosion General dangers, warnings, and cautionsSafety level Possible improper device performance Safety levels Possible ECG misinterpretation Safety level Possible shock hazardDefibrillation warnings and cautions Maintenance cautions Monitoring cautionsSafety level Possible fire or shock hazard Parameter Settings Default Description Setup and Advanced Mode FeaturesSetup Overview Non-protocol parametersAutomatic protocol parameters NSA action Parameter Settings Default Manual override parametersUsing Setup Features Reviewing current setup Revising setupReceiving setup Sending and Receiving Clock Settings Reading setupUsing Advanced Mode Features Using the manual analyze feature Using the manual charge feature M3860A only M3860A/M3861A Heartstart FR2+ Recording data in internal memory Data Management and ReviewRecording Incident Data Recording data on a data cardM3860A/M3861A Heartstart FR2+ Reviewing Incident Data Reviewing data from internal memoryReviewing data from a data card Data Management and Review M3860A/M3861A Heartstart FR2+ Accessories for the HeartStart FR2+ HeartStart AccessoriesSuggested Additional Items Technical Specifications Category Nominal specifications Waveform ParametersEnergy dose ECG Analysis System Controls and indicators Accessories Specifications Optional M3848A FR2+ rechargeable battery Optional M3854A data card Optional M3873A/M3874A FR2+ ECG assessment M3860A/M3861A Heartstart FR2+ Symbol Description Glossary of Symbols and ControlsSymboldescription Instructions for UseSymbol Accessories Symbols YYYY-MM M3863A standard battery M3848A FR2+ rechargeable battery M3849A charger For the M3848A FR2+ rechargeable battery Printed on recycled paper Electrical input Electrical output M3873A/M3874A FR2+ ECG assessment module Symboldescription M3855A charger For the M3864A training & administration pack Glossary of Terms Defibrillator pads with the HeartStart FR2+ See defibrillator pads Running HeartStart Event Review Web software Charger M3860A/M3861A Heartstart FR2+ Index ECG analysis system description D-2 Index M3860A/M3861A Heartstart FR2+ Index M3860A/M3861A Heartstart FR2+ Hilip s Me dical Sy ste ms Royal Philips Electronics Philips Medical Systems is part8 6 0 9 1 9 0 0