Troubleshooting Guide

Status indicator summary

status indicator

meaning

 

 

Flashing black hourglass The FR2+ passed the battery insertion self-test or the last periodic self-test and is therefore ready for use.

Flashing red X

A self-test error has occurred or the battery is low or depleted.

 

accompanied by a chirping

 

 

sound.

 

 

 

 

 

Solid red X

The battery is completely depleted or a self-test failure occurred.

 

 

 

 

 

 

 

 

N O T E : Perform CPR (if needed) any time there is a delay before the FR2+

4

 

can be used.

 

 

 

Recommended action during an emergency

If the status indicator displays the flashing black hourglass, follow all voice and screen prompts.

The HeartStart FR2+ Defibrillator is designed to continue working even if the status indicator displays a flashing red X, although the device may not perform to all of its specifications. Voice and text prompts should be followed whenever they are given. If for any reason you cannot hear voice prompts during use of the defibrillator, periodically check the device screen for text prompts.

N O T E : After completing emergency use of the FR2+, if you are unable to clear the problem as described in this Troubleshooting section, and the Status Indicator does not show the flashing black hourglass, contact Philips Medical Systems for technical support.

In the unlikely event that the device becomes unresponsive during use:

1.cycle power (press the On/Off button once, wait one second, then press it again), or

2.remove and reinstall the battery (use a new M3863A FR2 standard battery, if available, or a charged M3848A FR2+ rechargeable battery).

Maintaining, Testing, and Troubleshooting Your HeartStart FR2+

Image 37
Contents S T R U C T I O N S F O R U S E A R T S T a R T F R 2 + D E F I B R I L L a T O RHeartStart FR2+ Defibrillator I C K R E F E R E N C E C a R D Intentionally left blank A R T S T a R T F R 2 + M3860A, M3861A Defibrillator Device Tracking Contents Setup and Advanced Mode Features Clinical and Safety ConsiderationsData Management and Review Technical Specifications Accessories for the HeartStart FR2+Glossary of Symbols and Controls AppendicesM3860A/M3861A Heartstart FR2+ Introduction to the HeartStart FR2+ What is the HeartStart FR2+?Model M3860A Model M3861A How Does the HeartStart FR2+ Work? When Is the HeartStart FR2+ Used?How Is the HeartStart FR2+ Supplied? M3860A/M3861A Heartstart FR2+ Preparing Your HeartStart FR2+ for Use Installing the BatteryOverview M3860A/M3861A Heartstart FR2+ Setting the Clock Running the Battery Insertion Selftest Press the upper Option button to activate the test Placing and Securing the HeartStart FR2+M3860A/M3861A Heartstart FR2+ Using Your HeartStart FR2+ Preparation ECG Analysis and Monitoring Shock Delivery ECG Display for Ongoing Observation M3860A/M3861A Heartstart FR2+ Maintenance Daily Monthly Maintenance task/response After Using the HeartStart FR2+Operator’s checklist Cleaning the HeartStart FR2+Date Scheduled frequency Operators ChecklistBattery insertion selftest TestingM3860A/M3861A Heartstart FR2+ Hili ps Me dical Sy stems Test Feature Test descriptionDevice history Periodic selftestsBattery History Status indicator summary Troubleshooting GuideStatus indicator Meaning Recommended action during an emergencyStatus Indicator Flashing Black Hourglass Symptom Possible cause Recommended actionSymptom Possible cause Symptom General troubleshootingFR2+ has been physically damaged M3860A/M3861A Heartstart FR2+ Safety Considerations Clinical and Safety ConsiderationsClinical Considerations IndicationsGeneral dangers, warnings, and cautions Safety level Possible shock or fire hazard, or explosionSafety level Possible improper device performance Safety levels Possible ECG misinterpretation Safety level Possible shock hazardDefibrillation warnings and cautions Monitoring cautions Maintenance cautionsSafety level Possible fire or shock hazard Setup Overview Setup and Advanced Mode FeaturesParameter Settings Default Description Non-protocol parametersAutomatic protocol parameters NSA action Manual override parameters Parameter Settings DefaultUsing Setup Features Revising setup Reviewing current setupReceiving setup Reading setup Sending and Receiving Clock SettingsUsing Advanced Mode Features Using the manual analyze feature Using the manual charge feature M3860A only M3860A/M3861A Heartstart FR2+ Recording Incident Data Data Management and ReviewRecording data in internal memory Recording data on a data cardM3860A/M3861A Heartstart FR2+ Reviewing data from internal memory Reviewing Incident DataReviewing data from a data card Data Management and Review M3860A/M3861A Heartstart FR2+ HeartStart Accessories Accessories for the HeartStart FR2+Suggested Additional Items Technical Specifications Waveform Parameters Category Nominal specificationsEnergy dose ECG Analysis System Controls and indicators Accessories Specifications Optional M3848A FR2+ rechargeable battery Optional M3854A data card Optional M3873A/M3874A FR2+ ECG assessment M3860A/M3861A Heartstart FR2+ Symboldescription Glossary of Symbols and ControlsSymbol Description Instructions for UseSymbol Accessories Symbols YYYY-MM M3863A standard battery M3848A FR2+ rechargeable battery M3849A charger For the M3848A FR2+ rechargeable battery Printed on recycled paper Electrical input Electrical output M3873A/M3874A FR2+ ECG assessment module Symboldescription M3855A charger For the M3864A training & administration pack Glossary of Terms Defibrillator pads with the HeartStart FR2+ See defibrillator pads Running HeartStart Event Review Web software Charger M3860A/M3861A Heartstart FR2+ Index ECG analysis system description D-2 Index M3860A/M3861A Heartstart FR2+ Index M3860A/M3861A Heartstart FR2+ Hilip s Me dical Sy ste ms Royal Philips Electronics Philips Medical Systems is part8 6 0 9 1 9 0 0