Step 1: Preparation

Press the On/Off button to turn on the HeartStart FR2+ Defibrillator. Follow the instructions provided by the FR2+ voice and screen prompts in the order indicated.

Remove clothing from the patient's upper body. Wipe moisture from the patient's skin and clip or shave excessive chest hair, if necessary.

If the patient appears to be under eight years of age or 55 lbs (25 Kg), use M3870A FR2 infant/child reduced-energy defibrillator pads, if available. If the child appears older/larger, use adult defibrillator pads. DO NOT DELAY

TREATMENT TO DETERMINE THE CHILDS EXACT AGE/WEIGHT.

Open the defibrillator pads package. Check to see that the pads and attached cable and connector are undamaged. Pull off the protective backing from the defibrillator pads and check that the gel has not dried out. If the pads are damaged or the gel has dried out, use a new set of pads.

Place each pad on the patient. The pads must be placed with the sticky side on the patient’s skin. I M P O R T A N T : Refer to the drawing on each pad for correct positioning. For adult patients, one pad goes just below the patient's right collarbone, and the other one goes over the patient’s ribs in line with the armpit and below the left breast. For children under eight years old, one pad is centered on the chest between the nipples, and the other on the back between the scapulae (shoulder bones).

Connect the pads to the FR2+. Insert the defibrillator pads connector firmly into the connector socket. A flashing light shows you where the socket is located, at the top left of the FR2+.

P hili ps Me dical Sy stems

M3860A/M3861A HEARTSTART FR2+

Image 22
Contents A R T S T a R T F R 2 + D E F I B R I L L a T O R S T R U C T I O N S F O R U S EHeartStart FR2+ Defibrillator I C K R E F E R E N C E C a R D Intentionally left blank A R T S T a R T F R 2 + M3860A, M3861A Defibrillator Device Tracking Contents Setup and Advanced Mode Features Clinical and Safety ConsiderationsData Management and Review Glossary of Symbols and Controls Accessories for the HeartStart FR2+Technical Specifications AppendicesM3860A/M3861A Heartstart FR2+ Introduction to the HeartStart FR2+ What is the HeartStart FR2+?Model M3860A Model M3861A When Is the HeartStart FR2+ Used? How Does the HeartStart FR2+ Work?How Is the HeartStart FR2+ Supplied? M3860A/M3861A Heartstart FR2+ Preparing Your HeartStart FR2+ for Use Installing the BatteryOverview M3860A/M3861A Heartstart FR2+ Setting the Clock Running the Battery Insertion Selftest Placing and Securing the HeartStart FR2+ Press the upper Option button to activate the testM3860A/M3861A Heartstart FR2+ Using Your HeartStart FR2+ Preparation ECG Analysis and Monitoring Shock Delivery ECG Display for Ongoing Observation M3860A/M3861A Heartstart FR2+ Maintenance After Using the HeartStart FR2+ Daily Monthly Maintenance task/responseCleaning the HeartStart FR2+ Operator’s checklistOperators Checklist Date Scheduled frequencyTesting Battery insertion selftestM3860A/M3861A Heartstart FR2+ Hili ps Me dical Sy stems Feature Test description TestPeriodic selftests Device historyBattery History Status indicator Meaning Troubleshooting GuideStatus indicator summary Recommended action during an emergencySymptom Possible cause Recommended action Status Indicator Flashing Black HourglassSymptom Possible cause General troubleshooting SymptomFR2+ has been physically damaged M3860A/M3861A Heartstart FR2+ Clinical Considerations Clinical and Safety ConsiderationsSafety Considerations IndicationsSafety level Possible shock or fire hazard, or explosion General dangers, warnings, and cautionsSafety level Possible improper device performance Safety levels Possible ECG misinterpretation Safety level Possible shock hazardDefibrillation warnings and cautions Maintenance cautions Monitoring cautionsSafety level Possible fire or shock hazard Parameter Settings Default Description Setup and Advanced Mode FeaturesSetup Overview Non-protocol parametersAutomatic protocol parameters NSA action Parameter Settings Default Manual override parametersUsing Setup Features Reviewing current setup Revising setupReceiving setup Sending and Receiving Clock Settings Reading setupUsing Advanced Mode Features Using the manual analyze feature Using the manual charge feature M3860A only M3860A/M3861A Heartstart FR2+ Recording data in internal memory Data Management and ReviewRecording Incident Data Recording data on a data cardM3860A/M3861A Heartstart FR2+ Reviewing Incident Data Reviewing data from internal memoryReviewing data from a data card Data Management and Review M3860A/M3861A Heartstart FR2+ Accessories for the HeartStart FR2+ HeartStart AccessoriesSuggested Additional Items Technical Specifications Category Nominal specifications Waveform ParametersEnergy dose ECG Analysis System Controls and indicators Accessories Specifications Optional M3848A FR2+ rechargeable battery Optional M3854A data card Optional M3873A/M3874A FR2+ ECG assessment M3860A/M3861A Heartstart FR2+ Symbol Description Glossary of Symbols and ControlsSymboldescription Instructions for UseSymbol Accessories Symbols YYYY-MM M3863A standard battery M3848A FR2+ rechargeable battery M3849A charger For the M3848A FR2+ rechargeable battery Printed on recycled paper Electrical input Electrical output M3873A/M3874A FR2+ ECG assessment module Symboldescription M3855A charger For the M3864A training & administration pack Glossary of Terms Defibrillator pads with the HeartStart FR2+ See defibrillator pads Running HeartStart Event Review Web software Charger M3860A/M3861A Heartstart FR2+ Index ECG analysis system description D-2 Index M3860A/M3861A Heartstart FR2+ Index M3860A/M3861A Heartstart FR2+ Hilip s Me dical Sy ste ms Royal Philips Electronics Philips Medical Systems is part8 6 0 9 1 9 0 0