Philips M3860A, M3861A manual Controls and indicators

Page 73

P hili ps Me dical Sy stems

 

B-5

B

 

 

 

 

 

category

nominal specifications

 

 

 

 

Pacemaker Detection

On detection of a pacemaker (in advanced mode or with M3873A/M3874A FR2+

 

 

ECG assessment module), provides screen display of PACEMAKER DETECTED

 

 

alert. M3860A includes pacemaker artifact in ECG display. In both models,

 

 

pacemaker artifact is removed from the signal for rhythm analysis.

 

 

Display

 

 

 

 

 

 

category

nominal specifications

 

 

 

 

Monitored ECG Lead

ECG information is received from adult defibrillator pads in anterior-anterior (Lead

 

II)position or from FR2 infant/child reduced-energy defibrillator pads in anterior-posterior position. (Displayed on M3860A only.)

ECG information can also be displayed in the M3860A using the FR2+ ECG assessment module.

 

NOTE: The ECG display provided by the FR2+ Defibrillator is not is not

 

intended to provide diagnostic or ST segment interpretation.

 

 

Display Range

Differential: ±2 mV full scale, nominal.

(M3860A only)

 

 

 

Screen Type

High-resolution liquid crystal display (LCD) with backlight.

 

 

Screen Dimensions

2.8" wide x 2.3" high (70 mm x 58 mm).

 

 

Sweep Speed

23 mm/s nominal.

(M3860A only)

 

 

 

ECG Display

3 second-segments displayed (M3860A only).

 

 

Frequency Response

Nondiagnostic rhythm monitor 1 Hz to 20 Hz (-3 dB), nominal.

(Bandwidth)

 

 

 

Sensitivity

1.16 cm/mV, nominal.

 

 

Heart Rate Displayed

30 to 300 bpm, updated each analysis period. Displayed (M3860A only) during

(Normal Sinus Rhythm)

monitoring and advanced modes.

 

Controls and indicators

 

 

 

category

nominal specifications

 

 

LCD Screen

High-resolution, backlighted LCD screen, displays text messages and (model

 

M3860A only) ECG.

 

 

Technical Specifications

Image 73
Contents S T R U C T I O N S F O R U S E A R T S T a R T F R 2 + D E F I B R I L L a T O RHeartStart FR2+ Defibrillator I C K R E F E R E N C E C a R D Intentionally left blank A R T S T a R T F R 2 + M3860A, M3861A Defibrillator Device Tracking Contents Setup and Advanced Mode Features Clinical and Safety ConsiderationsData Management and Review Technical Specifications Accessories for the HeartStart FR2+Glossary of Symbols and Controls AppendicesM3860A/M3861A Heartstart FR2+ Introduction to the HeartStart FR2+ What is the HeartStart FR2+?Model M3860A Model M3861A How Does the HeartStart FR2+ Work? When Is the HeartStart FR2+ Used?How Is the HeartStart FR2+ Supplied? M3860A/M3861A Heartstart FR2+ Preparing Your HeartStart FR2+ for Use Installing the BatteryOverview M3860A/M3861A Heartstart FR2+ Setting the Clock Running the Battery Insertion Selftest Press the upper Option button to activate the test Placing and Securing the HeartStart FR2+M3860A/M3861A Heartstart FR2+ Using Your HeartStart FR2+ Preparation ECG Analysis and Monitoring Shock Delivery ECG Display for Ongoing Observation M3860A/M3861A Heartstart FR2+ Maintenance Daily Monthly Maintenance task/response After Using the HeartStart FR2+Operator’s checklist Cleaning the HeartStart FR2+Date Scheduled frequency Operators ChecklistBattery insertion selftest TestingM3860A/M3861A Heartstart FR2+ Hili ps Me dical Sy stems Test Feature Test descriptionDevice history Periodic selftestsBattery History Status indicator summary Troubleshooting GuideStatus indicator Meaning Recommended action during an emergencyStatus Indicator Flashing Black Hourglass Symptom Possible cause Recommended actionSymptom Possible cause Symptom General troubleshootingFR2+ has been physically damaged M3860A/M3861A Heartstart FR2+ Safety Considerations Clinical and Safety ConsiderationsClinical Considerations IndicationsGeneral dangers, warnings, and cautions Safety level Possible shock or fire hazard, or explosionSafety level Possible improper device performance Safety levels Possible ECG misinterpretation Safety level Possible shock hazardDefibrillation warnings and cautions Monitoring cautions Maintenance cautionsSafety level Possible fire or shock hazard Setup Overview Setup and Advanced Mode FeaturesParameter Settings Default Description Non-protocol parametersAutomatic protocol parameters NSA action Manual override parameters Parameter Settings DefaultUsing Setup Features Revising setup Reviewing current setupReceiving setup Reading setup Sending and Receiving Clock SettingsUsing Advanced Mode Features Using the manual analyze feature Using the manual charge feature M3860A only M3860A/M3861A Heartstart FR2+ Recording Incident Data Data Management and ReviewRecording data in internal memory Recording data on a data cardM3860A/M3861A Heartstart FR2+ Reviewing data from internal memory Reviewing Incident DataReviewing data from a data card Data Management and Review M3860A/M3861A Heartstart FR2+ HeartStart Accessories Accessories for the HeartStart FR2+Suggested Additional Items Technical Specifications Waveform Parameters Category Nominal specificationsEnergy dose ECG Analysis System Controls and indicators Accessories Specifications Optional M3848A FR2+ rechargeable battery Optional M3854A data card Optional M3873A/M3874A FR2+ ECG assessment M3860A/M3861A Heartstart FR2+ Symboldescription Glossary of Symbols and ControlsSymbol Description Instructions for UseSymbol Accessories Symbols YYYY-MM M3863A standard battery M3848A FR2+ rechargeable battery M3849A charger For the M3848A FR2+ rechargeable battery Printed on recycled paper Electrical input Electrical output M3873A/M3874A FR2+ ECG assessment module Symboldescription M3855A charger For the M3864A training & administration pack Glossary of Terms Defibrillator pads with the HeartStart FR2+ See defibrillator pads Running HeartStart Event Review Web software Charger M3860A/M3861A Heartstart FR2+ Index ECG analysis system description D-2 Index M3860A/M3861A Heartstart FR2+ Index M3860A/M3861A Heartstart FR2+ Hilip s Me dical Sy ste ms Royal Philips Electronics Philips Medical Systems is part8 6 0 9 1 9 0 0