Page 63

stems

7-3

To replace a data card:

I M P O R T A N T : You must turn the FR2+ off (return it to standby mode) before you remove the data card, to ensure that no incident data are lost.

1.Press the On/Off button to turn off the FR2+. Never replace the data card unless the FR2+ is turned off.

2.Remove the loaded data card tray by grasping its handle and pulling it out of the port.

3.Remove the data card from the tray.

4.Give the data card to the appropriate person in your organization.

5.Because it helps seal the FR2+ against moisture, the data card tray should always be reinserted into the port of the FR2+. Either load a new data card into the tray and insert it, or insert the empty data card tray into the port.

Reviewing Incident Data

Reviewing data from internal memory

Summary information from the last incident that is stored in the internal memory of the HeartStart FR2+ can be displayed on its screen for review. To review this information:

1.Remove the data card if one is installed and unplug the pads connector.

2. Remove and reinstall the battery. (Make sure you are using the gray M3863A FR2 standard battery or the blue M3848A rechargeable battery, not the yellow training & administration pack.)

3. Select REVIEW INCIDENT from the menu. A new screen comes up.

4. Observe and record, if desired, the summary information displayed on the screen:

dical Sy

￿ how long the incident recorded by the FR2+ lasted, and

￿ how many shocks were delivered during the incident.

7

P hili ps Me

This information stays in the FR2+’s memory and can be displayed for review until the next time the FR2+ is used. At that time, the data from the new incident will be displayed.

Data Management and Review

Image 63
Contents S T R U C T I O N S F O R U S E A R T S T a R T F R 2 + D E F I B R I L L a T O RHeartStart FR2+ Defibrillator I C K R E F E R E N C E C a R D Intentionally left blank A R T S T a R T F R 2 + M3860A, M3861A Defibrillator Device Tracking Contents Clinical and Safety Considerations Setup and Advanced Mode FeaturesData Management and Review Appendices Accessories for the HeartStart FR2+Technical Specifications Glossary of Symbols and ControlsM3860A/M3861A Heartstart FR2+ What is the HeartStart FR2+? Introduction to the HeartStart FR2+Model M3860A Model M3861A How Does the HeartStart FR2+ Work? When Is the HeartStart FR2+ Used?How Is the HeartStart FR2+ Supplied? M3860A/M3861A Heartstart FR2+ Installing the Battery Preparing Your HeartStart FR2+ for UseOverview M3860A/M3861A Heartstart FR2+ Setting the Clock Running the Battery Insertion Selftest Press the upper Option button to activate the test Placing and Securing the HeartStart FR2+M3860A/M3861A Heartstart FR2+ Using Your HeartStart FR2+ Preparation ECG Analysis and Monitoring Shock Delivery ECG Display for Ongoing Observation M3860A/M3861A Heartstart FR2+ Maintenance Daily Monthly Maintenance task/response After Using the HeartStart FR2+Operator’s checklist Cleaning the HeartStart FR2+Date Scheduled frequency Operators ChecklistBattery insertion selftest TestingM3860A/M3861A Heartstart FR2+ Hili ps Me dical Sy stems Test Feature Test descriptionDevice history Periodic selftestsBattery History Recommended action during an emergency Troubleshooting GuideStatus indicator summary Status indicator MeaningStatus Indicator Flashing Black Hourglass Symptom Possible cause Recommended actionSymptom Possible cause Symptom General troubleshootingFR2+ has been physically damaged M3860A/M3861A Heartstart FR2+ Indications Clinical and Safety ConsiderationsSafety Considerations Clinical ConsiderationsGeneral dangers, warnings, and cautions Safety level Possible shock or fire hazard, or explosionSafety level Possible improper device performance Safety level Possible shock hazard Safety levels Possible ECG misinterpretationDefibrillation warnings and cautions Monitoring cautions Maintenance cautionsSafety level Possible fire or shock hazard Non-protocol parameters Setup and Advanced Mode FeaturesSetup Overview Parameter Settings Default DescriptionAutomatic protocol parameters NSA action Manual override parameters Parameter Settings DefaultUsing Setup Features Revising setup Reviewing current setupReceiving setup Reading setup Sending and Receiving Clock SettingsUsing Advanced Mode Features Using the manual analyze feature Using the manual charge feature M3860A only M3860A/M3861A Heartstart FR2+ Recording data on a data card Data Management and ReviewRecording Incident Data Recording data in internal memoryM3860A/M3861A Heartstart FR2+ Reviewing data from internal memory Reviewing Incident DataReviewing data from a data card Data Management and Review M3860A/M3861A Heartstart FR2+ HeartStart Accessories Accessories for the HeartStart FR2+Suggested Additional Items Technical Specifications Waveform Parameters Category Nominal specificationsEnergy dose ECG Analysis System Controls and indicators Accessories Specifications Optional M3848A FR2+ rechargeable battery Optional M3854A data card Optional M3873A/M3874A FR2+ ECG assessment M3860A/M3861A Heartstart FR2+ Instructions for Use Glossary of Symbols and ControlsSymboldescription Symbol DescriptionSymbol Accessories Symbols YYYY-MM M3863A standard battery M3848A FR2+ rechargeable battery M3849A charger For the M3848A FR2+ rechargeable battery Printed on recycled paper Electrical input Electrical output M3873A/M3874A FR2+ ECG assessment module Symboldescription M3855A charger For the M3864A training & administration pack Glossary of Terms Defibrillator pads with the HeartStart FR2+ See defibrillator pads Running HeartStart Event Review Web software Charger M3860A/M3861A Heartstart FR2+ Index ECG analysis system description D-2 Index M3860A/M3861A Heartstart FR2+ Index M3860A/M3861A Heartstart FR2+ Hilip s Me dical Sy ste ms Philips Medical Systems is part Royal Philips Electronics8 6 0 9 1 9 0 0