Philips M3861A, M3860A manual Reviewing data from a data card

Page 64

7-4

Reviewing data from a data card

If a data card is installed when the HeartStart FR2+ is turned on for use during an incident, the HeartStart FR2+ automatically records detailed information on the data card. To review this information on the HeartStart FR2+ screen:

1.Make sure the training & administration pack is not installed.

2. Make sure the data card is installed. Unplug the pads connector.

3. Remove and reinstall the battery.

4. Select REVIEW INCIDENT from the menu. A new screen comes up. This screen displays:

￿ELAPSED TIME — how long the incident recorded by the FR2+ lasted,*

￿SHOCKS DELIVERED — how many shocks were delivered during the incident, and

￿FIRST SHOCKS AT — the times at which the first three shocks were delivered.

N O T E : If the data card does not contain event data, only the summary information from FR2+ internal memory will be displayed when REVIEW INCIDENT is selected.

5.To review the events that occurred during the incident, select REVIEW

EVENTS. A new screen comes up. This and following screens, accessed by selecting NEXT EVENTS, display elapsed time information for critical activities in using the FR2+. These include:

￿ POWER ON — when the FR2+ was turned on,

￿PADS ON — when the defibrillator pads were connected,

￿SHOCK ADVISED — when a shock was advised,

￿ARMED — when the FR2+ charged for shock delivery,

￿SHOCKED — when a shock was delivered,

￿SHOCK ABORTED — when a shock was aborted,

￿PAUSE FOR CPR — when a pause occurred

￿POWER OFF — when the FR2+ was turned off

*The FR2+ displays elapsed time to a maximum of 99:59 minutes. If the elapsed time of use extends beyond this figure, the minutes are represented by “??” but the seconds are displayed. However, total elapsed time will be recorded on an installed data card for later review with HeartStart Event Review data management software.

P hili ps Me dical Sy stems

M3860A/M3861A HEARTSTART FR2+

Image 64
Contents A R T S T a R T F R 2 + D E F I B R I L L a T O R S T R U C T I O N S F O R U S EHeartStart FR2+ Defibrillator I C K R E F E R E N C E C a R D Intentionally left blank A R T S T a R T F R 2 + M3860A, M3861A Defibrillator Device Tracking Contents Setup and Advanced Mode Features Clinical and Safety ConsiderationsData Management and Review Accessories for the HeartStart FR2+ Technical SpecificationsGlossary of Symbols and Controls AppendicesM3860A/M3861A Heartstart FR2+ Introduction to the HeartStart FR2+ What is the HeartStart FR2+?Model M3860A Model M3861A When Is the HeartStart FR2+ Used? How Does the HeartStart FR2+ Work?How Is the HeartStart FR2+ Supplied? M3860A/M3861A Heartstart FR2+ Preparing Your HeartStart FR2+ for Use Installing the BatteryOverview M3860A/M3861A Heartstart FR2+ Setting the Clock Running the Battery Insertion Selftest Placing and Securing the HeartStart FR2+ Press the upper Option button to activate the testM3860A/M3861A Heartstart FR2+ Using Your HeartStart FR2+ Preparation ECG Analysis and Monitoring Shock Delivery ECG Display for Ongoing Observation M3860A/M3861A Heartstart FR2+ Maintenance After Using the HeartStart FR2+ Daily Monthly Maintenance task/responseCleaning the HeartStart FR2+ Operator’s checklistOperators Checklist Date Scheduled frequencyTesting Battery insertion selftestM3860A/M3861A Heartstart FR2+ Hili ps Me dical Sy stems Feature Test description TestPeriodic selftests Device historyBattery History Troubleshooting Guide Status indicator summaryStatus indicator Meaning Recommended action during an emergencySymptom Possible cause Recommended action Status Indicator Flashing Black HourglassSymptom Possible cause General troubleshooting SymptomFR2+ has been physically damaged M3860A/M3861A Heartstart FR2+ Clinical and Safety Considerations Safety ConsiderationsClinical Considerations IndicationsSafety level Possible shock or fire hazard, or explosion General dangers, warnings, and cautionsSafety level Possible improper device performance Safety levels Possible ECG misinterpretation Safety level Possible shock hazardDefibrillation warnings and cautions Maintenance cautions Monitoring cautionsSafety level Possible fire or shock hazard Setup and Advanced Mode Features Setup OverviewParameter Settings Default Description Non-protocol parametersAutomatic protocol parameters NSA action Parameter Settings Default Manual override parametersUsing Setup Features Reviewing current setup Revising setupReceiving setup Sending and Receiving Clock Settings Reading setupUsing Advanced Mode Features Using the manual analyze feature Using the manual charge feature M3860A only M3860A/M3861A Heartstart FR2+ Data Management and Review Recording Incident DataRecording data in internal memory Recording data on a data cardM3860A/M3861A Heartstart FR2+ Reviewing Incident Data Reviewing data from internal memoryReviewing data from a data card Data Management and Review M3860A/M3861A Heartstart FR2+ Accessories for the HeartStart FR2+ HeartStart AccessoriesSuggested Additional Items Technical Specifications Category Nominal specifications Waveform ParametersEnergy dose ECG Analysis System Controls and indicators Accessories Specifications Optional M3848A FR2+ rechargeable battery Optional M3854A data card Optional M3873A/M3874A FR2+ ECG assessment M3860A/M3861A Heartstart FR2+ Glossary of Symbols and Controls SymboldescriptionSymbol Description Instructions for UseSymbol Accessories Symbols YYYY-MM M3863A standard battery M3848A FR2+ rechargeable battery M3849A charger For the M3848A FR2+ rechargeable battery Printed on recycled paper Electrical input Electrical output M3873A/M3874A FR2+ ECG assessment module Symboldescription M3855A charger For the M3864A training & administration pack Glossary of Terms Defibrillator pads with the HeartStart FR2+ See defibrillator pads Running HeartStart Event Review Web software Charger M3860A/M3861A Heartstart FR2+ Index ECG analysis system description D-2 Index M3860A/M3861A Heartstart FR2+ Index M3860A/M3861A Heartstart FR2+ Hilip s Me dical Sy ste ms Royal Philips Electronics Philips Medical Systems is part8 6 0 9 1 9 0 0