Philips M3860A, M3861A manual Optional M3873A/M3874A FR2+ ECG assessment

Page 77

 

 

B-9

B

 

(Optional) M3873A/M3874A FR2+ ECG assessment

 

 

 

module

 

 

 

 

 

category

nominal specifications

 

 

 

 

Application

For use with the FR2+ M3860A with ECG display enabled and running version

 

 

1.5 software or higher (denoted by FR2+ on the front panel or rear label).

 

 

 

 

 

Length and Weight

100 inches (182 cm); ≤ 1 lb. (2.2 kg).

 

 

 

 

 

 

Operating Temperature and

32° to 122° F (0° to 50° C);

 

 

Humidity

0% to 95% relative humidity (non-condensing).

 

 

 

 

 

Storage Temperature and

32° to 109° F (0° to 43° C).

 

 

Humidity

0% to 75% relative humidity (non-condensing).

 

 

 

 

 

Patient Lead Wire

M3873A (AAMI):

M3874A (IEC):

 

Designation

positive lead − red

positive lead − green

 

 

negative lead − white

negative lead − red

 

 

reference lead − black.

reference lead − yellow

 

 

 

 

 

Typical (Lead II) Connection

Lead II vectors:

 

 

 

positive − left leg

 

 

 

negative − right arm

 

 

 

reference − left arm.

 

 

 

Other limb vectors can be obtained by different electrode positions.

 

 

 

 

Battery Type

3 V, 1 Ah, poly-carbonmonofluoride lithium (LiCFx). Non-replaceable disposable

 

 

primary cell.

 

 

 

 

 

 

Service Life

Typically, 5 years.

 

 

 

 

 

Performance with FR2+

Meets environmental specifications cited for FR2+ Defibrillator on page B1

 

Defibrillator

through B2.

 

 

 

 

 

 

P hili ps Me dical Sy stems

Technical Specifications

Image 77
Contents S T R U C T I O N S F O R U S E A R T S T a R T F R 2 + D E F I B R I L L a T O RHeartStart FR2+ Defibrillator I C K R E F E R E N C E C a R D Intentionally left blank A R T S T a R T F R 2 + M3860A, M3861A Defibrillator Device Tracking Contents Data Management and Review Clinical and Safety ConsiderationsSetup and Advanced Mode Features Technical Specifications Accessories for the HeartStart FR2+Glossary of Symbols and Controls AppendicesM3860A/M3861A Heartstart FR2+ Model M3860A Model M3861A What is the HeartStart FR2+?Introduction to the HeartStart FR2+ How Does the HeartStart FR2+ Work? When Is the HeartStart FR2+ Used?How Is the HeartStart FR2+ Supplied? M3860A/M3861A Heartstart FR2+ Overview Installing the BatteryPreparing Your HeartStart FR2+ for Use M3860A/M3861A Heartstart FR2+ Setting the Clock Running the Battery Insertion Selftest Press the upper Option button to activate the test Placing and Securing the HeartStart FR2+M3860A/M3861A Heartstart FR2+ Using Your HeartStart FR2+ Preparation ECG Analysis and Monitoring Shock Delivery ECG Display for Ongoing Observation M3860A/M3861A Heartstart FR2+ Maintenance Daily Monthly Maintenance task/response After Using the HeartStart FR2+Operator’s checklist Cleaning the HeartStart FR2+Date Scheduled frequency Operators ChecklistBattery insertion selftest TestingM3860A/M3861A Heartstart FR2+ Hili ps Me dical Sy stems Test Feature Test descriptionDevice history Periodic selftestsBattery History Status indicator summary Troubleshooting GuideStatus indicator Meaning Recommended action during an emergencyStatus Indicator Flashing Black Hourglass Symptom Possible cause Recommended actionSymptom Possible cause Symptom General troubleshootingFR2+ has been physically damaged M3860A/M3861A Heartstart FR2+ Safety Considerations Clinical and Safety ConsiderationsClinical Considerations IndicationsGeneral dangers, warnings, and cautions Safety level Possible shock or fire hazard, or explosionSafety level Possible improper device performance Defibrillation warnings and cautions Safety level Possible shock hazardSafety levels Possible ECG misinterpretation Monitoring cautions Maintenance cautionsSafety level Possible fire or shock hazard Setup Overview Setup and Advanced Mode FeaturesParameter Settings Default Description Non-protocol parametersAutomatic protocol parameters NSA action Manual override parameters Parameter Settings DefaultUsing Setup Features Revising setup Reviewing current setupReceiving setup Reading setup Sending and Receiving Clock SettingsUsing Advanced Mode Features Using the manual analyze feature Using the manual charge feature M3860A only M3860A/M3861A Heartstart FR2+ Recording Incident Data Data Management and ReviewRecording data in internal memory Recording data on a data cardM3860A/M3861A Heartstart FR2+ Reviewing data from internal memory Reviewing Incident DataReviewing data from a data card Data Management and Review M3860A/M3861A Heartstart FR2+ HeartStart Accessories Accessories for the HeartStart FR2+Suggested Additional Items Technical Specifications Waveform Parameters Category Nominal specificationsEnergy dose ECG Analysis System Controls and indicators Accessories Specifications Optional M3848A FR2+ rechargeable battery Optional M3854A data card Optional M3873A/M3874A FR2+ ECG assessment M3860A/M3861A Heartstart FR2+ Symboldescription Glossary of Symbols and ControlsSymbol Description Instructions for UseSymbol Accessories Symbols YYYY-MM M3863A standard battery M3848A FR2+ rechargeable battery M3849A charger For the M3848A FR2+ rechargeable battery Printed on recycled paper Electrical input Electrical output M3873A/M3874A FR2+ ECG assessment module Symboldescription M3855A charger For the M3864A training & administration pack Glossary of Terms Defibrillator pads with the HeartStart FR2+ See defibrillator pads Running HeartStart Event Review Web software Charger M3860A/M3861A Heartstart FR2+ Index ECG analysis system description D-2 Index M3860A/M3861A Heartstart FR2+ Index M3860A/M3861A Heartstart FR2+ Hilip s Me dical Sy ste ms 8 6 0 9 1 9 0 0 Philips Medical Systems is partRoyal Philips Electronics