Philips M3860A, M3861A manual FR2+ has been physically damaged

Page 41

4-15

P hili ps Me dical Sy stems

symptom

Status Indicator:

SOLID RED X

Audio signal:

NONE

Status Indicator:

SOLID RED X

Audio signal:

CHIRPING

Status Indicator:

NONE

possible cause

The battery is missing or completely depleted.

The training & administration pack is being used in the administration function (the solid red X is normal in this case) or has been left in the FR2+ by mistake.

A self-test detected a failure.

The training & administration pack is being used in the ADMINISTRATION function and more than 10 minutes have passed without user interaction (button press or pads change).

The training & administration pack is being used in the TRAINING function and more than 30 minutes have passed without user interaction (button press or pads change).

The FR2+ has been physically damaged.

recommended action

Install a new M3863A FR2 standard battery or a charged M3848A FR2+ rechargeable battery in the FR2+ and perform the battery insertion test (BIT).

Remove the training & administration pack and install a battery.

Remove and reinstall the battery and perform the battery insertion self-test. If it fails, install a new M3863A FR2 standard battery or a charged M3848A FR2+

rechargeable battery and repeat

 

the test. If it fails again, do not use

4

the FR2+.

To continue using the training & administration pack, press any button (except On/Off).

To return the FR2+ to standby mode, remove the Pack and install a battery.

Check for visible damage. Do not use the FR2+ if it appears to be damaged.

Remove and reinstall the battery to perform the battery insertion self-test. If it fails, install a new M3863A FR2 standard battery or a charged M3848A FR2+ rechargeable battery and repeat the test. If it fails again, do not use the FR2+.

Maintaining, Testing, and Troubleshooting Your HeartStart FR2+

Image 41
Contents S T R U C T I O N S F O R U S E A R T S T a R T F R 2 + D E F I B R I L L a T O RHeartStart FR2+ Defibrillator I C K R E F E R E N C E C a R D Intentionally left blank A R T S T a R T F R 2 + M3860A, M3861A Defibrillator Device Tracking Contents Data Management and Review Clinical and Safety ConsiderationsSetup and Advanced Mode Features Technical Specifications Accessories for the HeartStart FR2+Glossary of Symbols and Controls AppendicesM3860A/M3861A Heartstart FR2+ Model M3860A Model M3861A What is the HeartStart FR2+?Introduction to the HeartStart FR2+ How Does the HeartStart FR2+ Work? When Is the HeartStart FR2+ Used?How Is the HeartStart FR2+ Supplied? M3860A/M3861A Heartstart FR2+ Overview Installing the BatteryPreparing Your HeartStart FR2+ for Use M3860A/M3861A Heartstart FR2+ Setting the Clock Running the Battery Insertion Selftest Press the upper Option button to activate the test Placing and Securing the HeartStart FR2+M3860A/M3861A Heartstart FR2+ Using Your HeartStart FR2+ Preparation ECG Analysis and Monitoring Shock Delivery ECG Display for Ongoing Observation M3860A/M3861A Heartstart FR2+ Maintenance Daily Monthly Maintenance task/response After Using the HeartStart FR2+Operator’s checklist Cleaning the HeartStart FR2+Date Scheduled frequency Operators ChecklistBattery insertion selftest TestingM3860A/M3861A Heartstart FR2+ Hili ps Me dical Sy stems Test Feature Test descriptionDevice history Periodic selftestsBattery History Status indicator summary Troubleshooting GuideStatus indicator Meaning Recommended action during an emergency Status Indicator Flashing Black Hourglass Symptom Possible cause Recommended actionSymptom Possible cause Symptom General troubleshootingFR2+ has been physically damaged M3860A/M3861A Heartstart FR2+ Safety Considerations Clinical and Safety ConsiderationsClinical Considerations IndicationsGeneral dangers, warnings, and cautions Safety level Possible shock or fire hazard, or explosionSafety level Possible improper device performance Defibrillation warnings and cautions Safety level Possible shock hazardSafety levels Possible ECG misinterpretation Monitoring cautions Maintenance cautionsSafety level Possible fire or shock hazard Setup Overview Setup and Advanced Mode FeaturesParameter Settings Default Description Non-protocol parametersAutomatic protocol parameters NSA action Manual override parameters Parameter Settings DefaultUsing Setup Features Revising setup Reviewing current setupReceiving setup Reading setup Sending and Receiving Clock SettingsUsing Advanced Mode Features Using the manual analyze feature Using the manual charge feature M3860A only M3860A/M3861A Heartstart FR2+ Recording Incident Data Data Management and ReviewRecording data in internal memory Recording data on a data cardM3860A/M3861A Heartstart FR2+ Reviewing data from internal memory Reviewing Incident DataReviewing data from a data card Data Management and Review M3860A/M3861A Heartstart FR2+ HeartStart Accessories Accessories for the HeartStart FR2+Suggested Additional Items Technical Specifications Waveform Parameters Category Nominal specificationsEnergy dose ECG Analysis System Controls and indicators Accessories Specifications Optional M3848A FR2+ rechargeable battery Optional M3854A data card Optional M3873A/M3874A FR2+ ECG assessment M3860A/M3861A Heartstart FR2+ Symboldescription Glossary of Symbols and ControlsSymbol Description Instructions for UseSymbol Accessories Symbols YYYY-MM M3863A standard battery M3848A FR2+ rechargeable battery M3849A charger For the M3848A FR2+ rechargeable battery Printed on recycled paper Electrical input Electrical output M3873A/M3874A FR2+ ECG assessment module Symboldescription M3855A charger For the M3864A training & administration pack Glossary of Terms Defibrillator pads with the HeartStart FR2+ See defibrillator pads Running HeartStart Event Review Web software Charger M3860A/M3861A Heartstart FR2+ Index ECG analysis system description D-2 Index M3860A/M3861A Heartstart FR2+ Index M3860A/M3861A Heartstart FR2+ Hilip s Me dical Sy ste ms 8 6 0 9 1 9 0 0 Philips Medical Systems is partRoyal Philips Electronics