Philips M3860A, M3861A manual Technical Specifications

Page 69

B

 

B

Technical Specifications

 

 

The specifications for the HeartStart FR2+ provided in this chapter apply to

 

 

both the M3860A and M3861A, unless otherwise noted. Additional

 

 

information can be found in the Technical Reference Manual for the FR2

 

 

Defibrillator, located online at www.medical.philips.com.

 

 

HeartStart FR2+ Defibrillator Specifications

 

 

Physical

 

 

 

 

category

nominal specifications

 

 

 

 

Size

2.6" high x 8.6" wide x 8.6" deep (6.6 cm x 21.8 cm x 21.8 cm).

 

 

 

 

Weight

Approximately 4.7 lbs (2.1 kg) with M3863A FR2 standard battery installed.

 

 

Approximately 4.5 lbs (2 kg) with optional M3848A FR2+ rechargeable battery

 

 

installed.

 

 

 

 

 

Environmental

 

 

 

 

category

nominal specifications

 

 

 

 

Operating Temperature

32° to 122° F (0° to 50° C).

 

and Humidity

0% to 95% relative humidity (non-condensing).

 

 

 

 

Standby Temperature

32° to 109° F (0° to 43° C).

 

and Humidity

0% to 75% relative humidity (non-condensing).

 

 

Applies to HeartStart FR2+ with battery installed and stored with defibrillator

 

 

pads.

stemsSy

 

 

Altitude

Meets MIL-810E 500.3, Procedure II (-500 feet to 15,000 feet).

 

 

 

 

dicalMe

Shock/Drop Abuse

Meets MIL-STD-810E 516.4, Procedure IV (after a 1 meter drop to any edge,

Vibration

Meets MIL-STD-810E 514.4-17.

 

Tolerance

corner, or surface, in standby mode).

ps

 

 

Sealing

 

hiliP

With data card tray and battery installed, meets IEC 529 class IP54.

 

 

 

 

 

ESD

Meets EN 61000-4-2:1998 Severity Level 4.

 

 

 

 

EMI (Radiated)

Meets EN 60601-1-2 limits (1993), method EN 55011:1998 Group 1 Level B.

 

 

 

B-1

Image 69 Philips M3860A, M3861A manual Technical Specifications
Contents S T R U C T I O N S F O R U S E A R T S T a R T F R 2 + D E F I B R I L L a T O RHeartStart FR2+ Defibrillator I C K R E F E R E N C E C a R D Intentionally left blank A R T S T a R T F R 2 + M3860A, M3861A Defibrillator Device Tracking Contents Clinical and Safety Considerations Setup and Advanced Mode FeaturesData Management and Review Technical Specifications Accessories for the HeartStart FR2+Glossary of Symbols and Controls AppendicesM3860A/M3861A Heartstart FR2+ What is the HeartStart FR2+? Introduction to the HeartStart FR2+Model M3860A Model M3861A How Does the HeartStart FR2+ Work? When Is the HeartStart FR2+ Used?How Is the HeartStart FR2+ Supplied? M3860A/M3861A Heartstart FR2+ Installing the Battery Preparing Your HeartStart FR2+ for UseOverview M3860A/M3861A Heartstart FR2+ Setting the Clock Running the Battery Insertion Selftest Press the upper Option button to activate the test Placing and Securing the HeartStart FR2+M3860A/M3861A Heartstart FR2+ Using Your HeartStart FR2+ Preparation ECG Analysis and Monitoring Shock Delivery ECG Display for Ongoing Observation M3860A/M3861A Heartstart FR2+ Maintenance Daily Monthly Maintenance task/response After Using the HeartStart FR2+Operator’s checklist Cleaning the HeartStart FR2+Date Scheduled frequency Operators ChecklistBattery insertion selftest TestingM3860A/M3861A Heartstart FR2+ Hili ps Me dical Sy stems Test Feature Test descriptionDevice history Periodic selftestsBattery History Status indicator summary Troubleshooting GuideStatus indicator Meaning Recommended action during an emergencyStatus Indicator Flashing Black Hourglass Symptom Possible cause Recommended actionSymptom Possible cause Symptom General troubleshootingFR2+ has been physically damaged M3860A/M3861A Heartstart FR2+ Safety Considerations Clinical and Safety ConsiderationsClinical Considerations IndicationsGeneral dangers, warnings, and cautions Safety level Possible shock or fire hazard, or explosionSafety level Possible improper device performance Safety level Possible shock hazard Safety levels Possible ECG misinterpretationDefibrillation warnings and cautions Monitoring cautions Maintenance cautionsSafety level Possible fire or shock hazard Setup Overview Setup and Advanced Mode FeaturesParameter Settings Default Description Non-protocol parametersAutomatic protocol parameters NSA action Manual override parameters Parameter Settings DefaultUsing Setup Features Revising setup Reviewing current setupReceiving setup Reading setup Sending and Receiving Clock SettingsUsing Advanced Mode Features Using the manual analyze feature Using the manual charge feature M3860A only M3860A/M3861A Heartstart FR2+ Recording Incident Data Data Management and ReviewRecording data in internal memory Recording data on a data cardM3860A/M3861A Heartstart FR2+ Reviewing data from internal memory Reviewing Incident DataReviewing data from a data card Data Management and Review M3860A/M3861A Heartstart FR2+ HeartStart Accessories Accessories for the HeartStart FR2+Suggested Additional Items Technical Specifications Waveform Parameters Category Nominal specificationsEnergy dose ECG Analysis System Controls and indicators Accessories Specifications Optional M3848A FR2+ rechargeable battery Optional M3854A data card Optional M3873A/M3874A FR2+ ECG assessment M3860A/M3861A Heartstart FR2+ Symboldescription Glossary of Symbols and ControlsSymbol Description Instructions for UseSymbol Accessories Symbols YYYY-MM M3863A standard battery M3848A FR2+ rechargeable battery M3849A charger For the M3848A FR2+ rechargeable battery Printed on recycled paper Electrical input Electrical output M3873A/M3874A FR2+ ECG assessment module Symboldescription M3855A charger For the M3864A training & administration pack Glossary of Terms Defibrillator pads with the HeartStart FR2+ See defibrillator pads Running HeartStart Event Review Web software Charger M3860A/M3861A Heartstart FR2+ Index ECG analysis system description D-2 Index M3860A/M3861A Heartstart FR2+ Index M3860A/M3861A Heartstart FR2+ Hilip s Me dical Sy ste ms Philips Medical Systems is part Royal Philips Electronics8 6 0 9 1 9 0 0