Lifescan Blood Glucose Monitor System manual Important Control Solution Test Information

Page 21

Important Control Solution Test Information

Use only PocketScan™ Control Solution.

Check the expiration date on the control solution vial. Do not use if expired.

Shake the vial and discard the first drop of PocketScan Control Solution.

Use only for three months after first opening. Record the dis- card date (date opened plus three months) on the control solution vial. Discard after three months.

Store the control solution tightly closed at room temperature below 30°C (86°F). Do not refrigerate.

CAUTION: The control solution range printed on the test strip vial is for PocketScan Control Solution only. It is used to test meter and test strip performance. It is not a recommended range for your blood glucose level.

11

Image 21
Contents Owner’s Booklet Page Owner’s Booklet 5.8 Dear PocketScan System Owner Page Important Information Table of Contents Using the Meter Memory Viii Getting Started PocketScan Blood Glucose Meter Test Port Target Area PocketScan Test StripsImportant Test Strip Information Coding the Meter Checking the DisplayMatch the Code Numbers Enter the Code ModeCode the Meter Checking the System with PocketScan Control Solution Do a control solution test Important Control Solution Test Information Insert Test Strip Result Appears in 15 Seconds Apply Control SolutionAt UK 0800 Republic of Ireland 1800 Comparing control solution resultsTesting Your Blood Getting a Drop of Blood Insert a Lancet in the Penlet II Automatic Blood SamplerCock the Penlet II Sampler Wash Your Hands Lance Your FingerRemove the Lancet Step-by-Step Test Procedure Accurate Results in Just 15 Seconds Apply SampleExample Special MessagesUsing the Meter Memory Recall Test Results Data Downloading Exit the Memory ModeResults References Batteries MaintenanceSymbol appears on the display Display Messages and PROBLEM- Solving Guide Message What IT Means Result. At the end of the count Message What IT Means Message What IT Means Action That Blood Sample was Tests were performed within This period Select the appropriate unit of measurement Probable Cause What to do Calibration Plasma/Serum Test Time 15 secondsAutomatic Shutoff Two minutes after last user action Operating RangesChanging the Date Set the YearSet the Day Set the Month Changing the Time Set the HourSet the Minutes Select mmol/L or mg/dL Exit the Setting ModePage Test Page 1998 Visit the LifeScan website at