Lifescan Blood Glucose Monitor System manual Apply Control Solution, Result Appears in 15 Seconds

Page 23

If you decide not to perform a control solution test, the “ctl” sign will disappear if you press the C button again.

Note: Every time you perform a control solution test, you must mark the test with the “ctl” sign so that the test will be distinguished from a blood glucose test in the meter memory.

STEP 2

Apply Control Solution.

Shake the control solution vial well. Remove the cap. Squeeze the vial and dis- card the first drop. Apply the second drop to the white target area of the test strip.

STEP 3

Result Appears in 15 Seconds.

The meter will count down from “15” to “1” second and the control solution test result will appear. Compare the result with the range printed on the test strip vial. The result should fall within this range.

Solution

Normal

Glucose

Control

(Example)

13

Image 23
Contents Owner’s Booklet Page Owner’s Booklet 5.8 Dear PocketScan System Owner Page Important Information Table of Contents Using the Meter Memory Viii Getting Started PocketScan Blood Glucose Meter Test Port Target Area PocketScan Test StripsImportant Test Strip Information Coding the Meter Checking the DisplayMatch the Code Numbers Enter the Code ModeCode the Meter Checking the System with PocketScan Control Solution Do a control solution test Important Control Solution Test Information Insert Test Strip Result Appears in 15 Seconds Apply Control SolutionAt UK 0800 Republic of Ireland 1800 Comparing control solution resultsTesting Your Blood Cock the Penlet II Sampler Getting a Drop of BloodInsert a Lancet in the Penlet II Automatic Blood Sampler Remove the Lancet Wash Your HandsLance Your Finger Step-by-Step Test Procedure Accurate Results in Just 15 Seconds Apply SampleExample Special MessagesUsing the Meter Memory Recall Test Results Data Downloading Exit the Memory ModeResults References Batteries MaintenanceSymbol appears on the display Display Messages and PROBLEM- Solving Guide Message What IT Means Result. At the end of the count Message What IT Means Message What IT Means Action That Blood Sample was Tests were performed within This period Select the appropriate unit of measurement Probable Cause What to do Operating Ranges Test Time 15 secondsCalibration Plasma/Serum Automatic Shutoff Two minutes after last user actionSet the Day Changing the DateSet the Year Set the Month Set the Minutes Changing the TimeSet the Hour Select mmol/L or mg/dL Exit the Setting ModePage Test Page 1998 Visit the LifeScan website at