MODEL 54eA | SECTION 9.0 |
| CALIBRATION - FREE CHLORINE |
SECTION 9.0
CALIBRATION - FREE CHLORINE (498CL-01)
9.1 INTRODUCTION
This section covers the calibration of the
As Figure
The zero standard is necessary because chlorine sensors, even when no chlorine is in the sample, generate a small current called the residual current. The controller compensates for the residual current by subtracting it from the measured current before converting the result to a chlorine value. New sensors require zeroing before being placed in service, and sensors should be zeroed whenever the electrolyte solution is replaced. Either of the fol- lowing makes a good zero standard:
•Deionized water.
•Tap water known to contain no chlorine. Expose tap water to bright sunlight for at least 24 hours.
The purpose of the
•Take the grab sample from a point as close to the sensor as possible. Be sure that taking the sample does not alter the flow of the sample to the sensor. It is best to install the sample tap just downstream from the sensor.
•Chlorine solutions are unstable. Run the test immediately after taking the sample. Try to calibrate the sen- sor when the chlorine concentration is at the upper end of the normal operating range.
The Model
FIGURE 9-1. Sensor Current as a Function of Free Chlorine Concentration
60