MODEL 54eA | SECTION 5.0 |
| SOFTWARE CONFIGURATION |
TABLE | (continued) |
| |||
| ITEM |
|
| CHOICES | FACTORY SETTINGS |
CONFIGURE |
|
|
|
| |
A. Display options (Section 5.5) |
|
|
| ||
1. | Measurement | Oxygen, ozone, free chlorine, total chlorine, monochloramine | Oxygen | ||
2. | Sensor (Oxygen only) |
| Rosemount standard, Rosemount biopharm, or | Rosemount standard | |
|
|
|
| other steam sterilizable |
|
3. | Units (Oxygen only) |
|
| ppm, ppb, % saturation | ppm |
4. Temperature units |
|
| °C or °F | °C | |
5. | Output 1 |
|
| mA or % of full scale | mA |
6. | Output 2 |
|
| mA or % of full scale | mA |
7. | Language |
| English, Français, Español, Deutsch, Italiano | English | |
8. | Main display left |
|
| See section 5.5 | Sensor current |
9. | Main display right |
|
| See section 5.5 | Output 1 current |
10 Display contrast |
|
| 50 | ||
11. Test timeout |
|
| On or off | On | |
12. Timeout value |
|
| 1 to 60 min | 10 min | |
B Outputs (Section 5.6) |
|
|
|
| |
1. | Output 1 Control |
|
|
|
|
| a. Measurement |
| Oxygen, chlorine, ozone, pH, or temperature | Oxygen | |
| b. Control |
|
| Normal or PID | Normal |
2. | Output 1 Setup (normal) |
|
|
|
|
| a. Current |
|
| ||
| b. Dampening |
|
| 0 sec | |
| c. Hold mode |
|
| Hold last value or go to fixed value | Hold last value |
| d. Fixed hold value |
|
| 21 mA | |
| e. Fault value |
|
| 22 mA | |
3. | Output 1 Setup (PID) |
|
|
|
|
| a. Proportional |
|
| 0 to 299.9% | 100 % |
| b. Integral |
|
| 0 to 2999 sec | 0 sec |
| c. Derivative |
|
| 0 to 299.9% | 0/0% |
4. | Output 2 Control |
|
|
|
|
| a. Measurement |
| Oxygen, chlorine, ozone, pH, or temperature | Temperature | |
| b. Control |
|
| Normal or PID | Normal |
5. | Output 2 Setup (normal) |
|
| See output 1 | See output 1 |
6. | Output 2 Setup (PID) |
|
| See output 1 | See output 1 |
7. | Hold feature |
|
| Enable, disable, or 20 min timeout | Disable |
C. Alarms (Section 5.7) |
|
|
|
| |
1. | Alarm 1 Control |
|
|
|
|
| a. Activation method |
| Oxygen, chlorine, ozone, temperature, pH | Oxygen | |
| b. Control mode |
|
| Normal or TPC | Normal |
2. | Alarm 1 setup (normal) |
|
|
|
|
| a. Configuration |
|
| Low, high, or off | High |
| b. Hysteresis |
|
|
|
|
| if oxygen (ppm) |
|
| 0 to 20 ppm | 0 ppm |
| if oxygen (ppb) |
|
| 0 to 999 ppb | 0 ppb |
| if oxygen (% saturation) |
|
| 0 to 200% | 0% |
| if chlorine, monochloramine, or ozone | 0 ppm | |||
| if pH |
|
| 0 to 14.00 | 0 pH |
| if temperature |
|
| 0 to 10°C | 0.1°C |
| c. Delay time |
|
| 0 sec | |
| d. Relay fault |
|
| none, open, closed | None |
|
|
|
|
|
|
*
18 | Continued on the following page |
|