Emerson Process Management 53eA instruction manual Process Reaction Curve

Page 89

MODEL 54eA

SECTION 15.0

 

CALIBRATION - CONTROL

FIGURE 15-1. The Process Reaction Curve

6.When sufficient data have been collected, return the output signal to its original value using the sim- ulate test function. Maintain the controller in this manual mode until you are ready to initiate auto- matic PID control, after you have calculated the tuning constants.

Once these steps are completed, the resulting process reaction curve is used to obtain information about the overall dynamics of the system. It will be used to cal- culate the needed tuning parameters of the Model 54eA controller.

NOTE

The tuning procedure outlined below is adapted from "Instrumentation and Process Measurement and Control", by Norman A. Anderson, Chilton Co., Radnor, Pennsylvania, ©1980.

Information derived from the process reaction curve will be used with the following empirical formulas to predict the optimum settings for proportional and inte- gral tuning parameters.

Four quantities are determined from the process reac- tion curve for use in the formulas: time delay (D), time period (L), a ratio of these two (R), and plant gain (C).

A line is drawn on the process reaction curve tangent to the curve at point of maximum rise (slope) as shown in Figure 15-1. The Time Delay (D), or lag time, extends from "zero time" on the horizontal axis to the point where the tangent line intersects the time axis. The Response Time period (L), extends from the end of delay period to the time at which the tangent line intersects the 100% reaction completion line repre- senting the process stabilization value. The ratio (R) of the Response Time period to the Time Delay describes the dynamic behavior of the system.

In the example, the process Delay Time (D) was four seconds and the Response Time period (L) was 12 seconds, so:

R = L 12 seconds = 3

D4 seconds

The last parameter used in the equations is a plant gain (C). The plant gain is defined as a percent change in the controlled variable divided by the per- cent change in manipulated variable; in other words, the change in the measured variable (pH, conductivi- ty, temperature) divided by the percent change in the analog output signal.

The percent change in the controlled variable is defined as the change in the measured variable (pH,

83

Image 89
Contents Amperometric Hart Analyzer/Controller Model 54eAEssential Instructions Page Model 54eA ANALYZER/CONTROLLER 11.0 10.012.0 13.0Iii List of FiguresSection Title Features and Applications Section SpecificationsPower Specifications GeneralOutput Accuracy ± 0.05 mA Alarms Weight/Shipping Weight 5 lb/6 lb 2 kg/2.5 kgSpecifications Free Chlorine Specifications OxygenSpecifications Total Chlorine Specifications Ozone24 VDC 115/230 VAC, 50/60 Hz PowerOrdering Information Unpacking and Inspection Section InstallationInstallation Pipe mounting Pipe and Wall Mounting DimensionsPanel mounting Panel Mounting DimensionsGeneral Section WiringPOWER, ALARM, and Output Wiring 454EPH02 General Sensor WiringWiring Model 499A oxygen, chlorine, and ozone sensors 4054eA01 4054eA03 4054eA02 4054eA04 4054eA07 Section Display and Operation Alarm StatusGeneral Description DisplayProgram Settings List Section Software ConfigurationChoices Factory Settings Configure Choices Factory Settings Temperature compensation Section Menu Tree for the 54eA Controller Main Menu Main Alarm 2 setpoint Alarm setpointsRanging the Outputs MA 0.00 ppm 20 mA 20.00 ppm Output 1 12.00 mAOutput setpoints Alarm setpointsSimulated test Setpoint 1.000 ppm 4mA 0.000 ppm 20mA 10.00 ppmTest output Test alarm 1 OpenSimulated tests Display ConfigureLanguage English Display left Display right Out Display contrast Timeout OnTemp units C Output 1 mA Output 2 mA Output Measurement OutputsOutput 1 control Output 1 Setup Ctrl mode NormalOutput 1 Control Range 4-20 mA Dampen 0 sec Hold Last ValueUsing hold Hold setupAlways configure the control parameters Before making Changing Alarm ParametersChanges in the alarm setup Alarm Control SettingsAlarm 1 setup Alarm 1 controlAlarm 2 control Alarm Low Setpoint 0.000 ppmHigh Alarm Low AlarmFeed limit Disable Alarm FaultTimeout 3600 sec Feed limit timer setupInterval timer Alarm 4 setup Feed limit timerTimer Disable Timer Time activated Interval 24.0 hr Interval timer setupPH sensor Disable PH sensor EnableUsing the 499ACL-01 sensor PH comp Manual PH value 7.00 pHAutocal Standard Diagnostics OffGlass imp lo 20 MΩ Operate iso 7.00 pH Sensor iso 7.00 pH Temp coeffTemp comp Auto Temperature Compensation and Temperature UnitsTemp units C Temp comp Manual Temp units CNoise Reduction 60 HzMain sensor cal Barometric Pressure Bar meas AutoBar units mm Hg Noise rejection Main sensor cal Lock program Lock configSecurity Lock allAction Definitions Controller Mode PriorityCondition Definitions Introduction Section Calibration TemperatureTemperature Calibration Adjust temp + 025.1 CDissolved Oxygen Concentration Section Calibration Dissolved OxygenZeroing the Sensor Sensor must be in zero solutionCalibrating the Sensor in AIR Stabilizing . . . WaitSection Calibrating the Sensor Against a Standard Instrument Calibrate 8.32 ppmCalibrate main sensor Zero main sensor Adjust temperatureStandardize Calibrating Barometric PressureSensor Current as a Function of Free Chlorine Concentration Section Calibration Free Chlorine 499ACL-01Calibrate main sensor Full Scale Calibration Calibrate 1.100 ppmOutput trim Dual range calCalibrate low point Dual Slope CalibrationCalibrate high point Low point 2.000 ppmIntroduction Section Calibration Free Chlorine 498CL-01Ppm 100ppm 000 ppm Section Calibration Free Chlorine 498CL-01 Section Calibration Total Chlorine This page intentionally left blank Sensor Current as a Function of Monochloramine Concentration Section Calibration MonochloraminePpm 100ppm Sensor Current as a Function of Ozone Concentration Section Calibration OzonePpm 100ppm PH at 25C Standards Nominal pH SectionAutomatic TWO-POINT Calibration Autocal buffer10.02 pH Manual TWO-POINT Calibration Calibrate pointPt 04.00pH 25C Pt 4.00pH 25.0C 10.01 pHPt calibration Standardize pH pH slope Standardization ONE-POINT CalibrationPH slope 60.00 mV/pH Trimming the Outputs Section Calibration Current OutputsPID Control Code Section PID and TPC ControlProcess Reaction Curve Method Proportional Gain Plus Integral ResetControl Loop Adjustment and Tuning Process Reaction Curve Section Time Proportional Control TPC Mode Code Troubleshooting When a Fault Message is Showing Section TroubleshootingFault message Explanation See Section OverviewHigh input current Temperature error low or highCheck sensor zero PH low or high input voltageProblem See Section Troubleshooting When no Fault Message is Showing OxygenSense line open Failure factory and Failure eepromZero reading Is unstable Possible error warning during in-process calibrationZero current is too high Process readings are erratic Barometric pressure reading is too high or too lowReadings drift Sensor does not respond to changes in oxygen levelZero current is unstable ProblemSensor can be calibrated, but the current is too low Chlorine readings are too low Sensor does not respond to changes in chlorine levelProblem See Section Sensor does not respond to changes in monochloramine level Readings are too low Troubleshooting When no Fault Message is Showing Ozone Ozone readings are too low Sensor does not respond to changes in ozone levelTroubleshooting When no Fault Message is Showing pH 100 Controller will not accept manual slopeSensor does not respond to known pH changes 101 Process pH readings are noisyProblem Action Troubleshooting not Related to Measurement ProblemsSimulating Inputs Dissolved Oxygen Sensor Polarizing Voltage Resistance Expected currentSimulate Chlorine and Ozone 103 Simulating Inputs Other Amperometric Measurements104 Simulating pH input when the preamplifier is in the sensorVoltage mV PH at 25C 105 Simulating TemperatureSimulating temperature 106 Measuring Reference VoltageReplacement Parts Part Number Description Section MaintenanceGeneral Section Return of MaterialWarranty Specifications subject to change without notice