MODEL 54eA | SECTION 5.0 |
| SOFTWARE CONFIGURATION |
5.5 CHOOSING DISPLAY OPTIONS
1.The 54eA controller can be used with most amperometric sensors manufactured by Rosemount Analytical. The user must configure the analyzer to match the sensor being used.
2.The display menu also lets the user customize the third line in the display, change timeout values, choose a lan- guage other than English, and change the display contrast.
Output setpoints
Simulated tests
Configure
ExitEnter
3.Press any key to enter the main menu. Move the cursor to "Program" and press Enter (F4).
4.Move the cursor to "Configure" and press Enter (F4).
5. With the cursor on "Display", press Enter (F4).
Display
Outputs
Alarms
ExitEnter
Measure : Oxygen
Sensor: RMT Standard
Meas units: ppm
ExitEdit
Measure : Oxygen
Sensor: RMT Standard
Meas units: ppm
ExitEnter
6.A screen showing the present main measurement will appear. To
change the measurement, for example, from oxygen to ozone, press Edit (F4), then use the é key to scroll through the choices. Press Save (F4) to store the setting.
A screen will appear warning that if the measurement is changed, the analyzer will return to factory default settings. Press Cont (F3) to con- tinue. Press Abort (F1) to cancel the change.
7.If oxygen is being measured, the screen at left appears. Use the ê key to move the cursor to "Sensor" and press Edit (F4). Use the é or ê keys to scroll through the choices. Press Save (F4) to store the selection.
| RMT Standard | 499ADO or 499ATrDO | |
|
|
|
|
| RMT Biopharm | Hx438 or Gx448 | |
|
|
|
|
| SSDO Other | Steam sterilizable sensor from other manufacturers | |
|
|
|
|
Use the ê key to move the cursor to "Meas units" and press Edit (F4). Use the é or ê keys to scroll through the choices. Press Save (F4) to store the selection.
| ppm | (mg/L) use with any sensor except 499ATrDO | |
|
|
|
|
| ppb | (ug/L) use with 499ATrDO sensor | |
|
|
|
|
| % sat | (concentration as a percent of saturation value) | |
|
| use with any sensor except 499ATrDO | |
|
|
|
|
28