MODEL 54eA | SECTION 16.0 | ||
|
| TROUBLESHOOTING | |
16.5 TROUBLESHOOTING WHEN NO FAULT MESSAGE IS SHOWING - FREE CHLORINE | |||
|
|
|
|
| Problem | See Section | |
| Zero current was accepted, but the current is out of range | 16.5.1 |
|
| Error or warning message appears while zeroing the sensor (zero current is too high) | 16.5.1 |
|
| Zero current is unstable | 16.5.2 |
|
| Sensor can be calibrated, but sensitivity is out of range | 16.5.3 |
|
| Process readings are erratic | 16.5.4 |
|
| Readings drift | 16.5.5 |
|
| Sensor does not respond to changes in chlorine level | 16.5.6 |
|
|
|
|
|
| Chlorine reading spikes following rapid change in pH | 16.5.7 |
|
|
|
|
|
16.5.1 Zero current is too high
A.Is the sensor properly wired to the controller. See Section 3.3.
B.Is the zero solution
C.Has adequate time been allowed for the sensor to reach a minimum stable residual current? It may take sev- eral hours, sometimes as long as overnight, for a new sensor to stabilize.
D.Check the membrane for damage and replace it if necessary.
E.Is the cathode coated with copper? (Applies to
16.5.2 Zero current is unstable
A.Is the sensor properly wired to the analyzer? See Section 3.3. Verify that all wiring connections are tight.
B.Readings are often erratic when a new or rebuilt sensor is first placed in service. Readings usually stabilize after about an hour.
C.Is the conductivity of the zero solution correct for the sensor being used? For a
D.Is the space between the membrane and cathode filled with electrolyte solution and is the flow path between the electrolyte reservoir and membrane clear? Often the flow of electrolyte and be started by simply holding the sensor with the membrane end pointing down and sharply shaking the sensor a few times as though shak- ing down a clinical thermometer.
If you are using a
The
Verify that the sensor is filled with electrolyte solution
91