MODEL 54eA | SECTION 5.0 |
| SOFTWARE CONFIGURATION |
5.7 CHANGING ALARM PARAMETERS
1.This section describes how to configure the controller alarms. Alarms 1, 2, and 3 can be assigned to the main amperometric measurement (oxygen, ozone, free chlorine, or total chlorine), temperature, or pH (if pH was enabled). In addition, alarm 1, 2, or 3 can be configured as a feed limit timer or as an interval timer (see steps 10 and 11). Alarm 4 is always a fault alarm.
2.An alarms assigned to the main amperometric measurement, temperature, or pH can be configured as either a simple on/off alarm or as TPC (time proportional control) alarm.
Output setpoints
Simulated tests
Configure
ExitEnter
3.Press any key to enter the main menu. Move the cursor to "Program" and press Enter (F4).
4.Move the cursor to "Configure" and press Enter (F4).
Outputs Alarms pH
5. Move the cursor to "Alarms" and press Enter (F4).
Exit | Enter | ||
|
| 6. Nine menu headers relate to alarms. Alarms 1, 2 and 3, each have a | |
Alarm 1 control |
| ||
| control header and a setup header. Alarm 4 has only a setup header. | ||
Alarm 1 setup | The eighth menu header is for configuring the feed limit timer, and the | ||
ninth menu header is for configuring the interval timer. | |||
|
| ||
Alarm 2 control | Always configure the control parameters BEFORE making | ||
|
| changes in the alarm setup. | |
Exit | Enter | ||
|
| To access a header, move the cursor to the desired header and press | |
|
| Enter (F4). | |
|
| 7. Alarm Control Settings: | |
Activation method | |||
a. Move the cursor to the desired output control header. Press Enter | |||
Control mode | |||
(F4). | |||
|
| b. With the cursor on "Activation method" press Enter (F4). | |
Exit | Enter |