Emerson Process Management 53eA instruction manual Section Calibration Dissolved Oxygen

Page 55

MODEL 54eA

SECTION 7.0

 

CALIBRATION - DISSOLVED OXYGEN

SECTION 7.0

CALIBRATION - DISSOLVED OXYGEN

7.1 INTRODUCTION

As Figure 7-1 shows, oxygen sensors generate a current directly proportional to the concentration of dissolved oxygen in the sample. Calibrating the sensor requires exposing it to a solution containing no oxygen (zero stan- dard) and to a solution containing a known amount of oxygen (full-scale standard).

The zero standard is necessary because oxygen sensors, even when no oxygen is present in the sample, gener- ate a small current called the residual current. The analyzer compensates for the residual current by subtracting it from the measured current before converting the result to a dissolved oxygen value. New sensors require zeroing before being placed in service, and sensors should be zeroed whenever the electrolyte solution is replaced. The recommended zero standard is 5% sodium sulfite in water, although oxygen-free nitrogen can also be used.

The Model 499A TrDO sensor, used for the determination of trace (ppb) oxygen levels, has very low resid- ual current and does not normally require zeroing. The residual current in the 499A TrDO sensor is equivalent to less than 0.5 ppb oxygen.

The purpose of the full-scale standard is to establish the slope of the calibration curve. Because the solubility of atmospheric oxygen in water as a function of temperature and barometric pressure is well known, the natural choice for a full-scale standard is air-saturated water. However, air-saturated water is difficult to prepare and use, so the universal practice is to use air for calibration. From the point of view of the oxygen sensor, air and air-sat- urated water are identical. The equivalence comes about because the sensor really measures the chemical poten- tial of oxygen. Chemical potential is the force that causes oxygen molecules to diffuse from the sample into the sensor where they can be measured. It is also the force that causes oxygen molecules in air to dissolve in water and to continue to dissolve until the water is saturated with oxygen. Once the water is saturated, the chemical potential of oxygen in the two phases (air and water) is the same.

Oxygen sensors generate a current directly proportional to the rate at which oxygen molecules diffuse through a membrane stretched over the end of the sensor. The diffusion rate depends on the difference in chemical poten- tial between oxygen in the sensor and oxygen in the sample. An electrochemical reaction, which destroys any oxy- gen molecules entering the sensor, keeps the concentration (and the chemical potential) of oxygen inside the sen- sor equal to zero. Therefore, the chemical potential of oxygen in the sample alone determines the diffusion rate and the sensor current.

When the sensor is calibrated, the chemical potential of oxygen in the standard determines the sensor current. Whether the sensor is calibrated in air or air-saturated water is immaterial. The chemical potential of oxygen is the same in either phase. Normally, to make the calculation of solubility in common units (like ppm DO) simpler, it is convenient to use water-saturated air for calibration.

Automatic air calibration is standard. The user simply exposes the sensor to water-saturated air. The controller monitors the sensor current. When the current is stable, the controller stores the current and measures the baro- metric pressure and temperature. The temperature element is part of the dissolved oxygen sensor. The pressure sensor is inside theanalyzer. From the temperature, the

controller calculates the saturation vapor pressure of water. Next, it calculates the pressure of dry air by sub- tracting the vapor pressure from the barometric pressure. Using the fact that dry air always contains 20.95% oxy- gen, the analyzer calculates the partial pressure of oxy- gen. Once the analyzer knows the partial pressure of oxy- gen, it uses the Bunsen coefficient to calculate the equi- librium solubility of atmospheric oxygen in water at the prevailing temperature. At 25°C and 760 mm Hg, the equilibrium solubility is 8.24 ppm.

Often it is too difficult or messy to remove the sensor from

 

 

 

the process liquid for calibration. In this case, the sensor

 

 

 

can be calibrated against a measurement made with a

 

 

 

portable laboratory instrument. The laboratory instrument

 

 

 

 

FIGURE 7-1. Sensor Current as a Function of

 

typically uses a membrane-covered amperometric sensor

 

 

 

Dissolved Oxygen Concentration

 

that has been calibrated against water-saturated air.

 

 

 

 

 

 

 

 

49

Image 55
Contents Amperometric Hart Analyzer/Controller Model 54eAEssential Instructions Page Model 54eA ANALYZER/CONTROLLER 13.0 10.011.0 12.0Section Title List of FiguresIii Features and Applications Section SpecificationsWeight/Shipping Weight 5 lb/6 lb 2 kg/2.5 kg Specifications GeneralPower Output Accuracy ± 0.05 mA AlarmsSpecifications Ozone Specifications OxygenSpecifications Free Chlorine Specifications Total ChlorineOrdering Information 115/230 VAC, 50/60 Hz Power24 VDC Installation Section InstallationUnpacking and Inspection Pipe mounting Pipe and Wall Mounting DimensionsPanel mounting Panel Mounting DimensionsPOWER, ALARM, and Output Wiring Section WiringGeneral 454EPH02 Wiring Model 499A oxygen, chlorine, and ozone sensors Sensor WiringGeneral 4054eA01 4054eA03 4054eA02 4054eA04 4054eA07 Display Alarm StatusSection Display and Operation General DescriptionProgram Settings List Section Software ConfigurationChoices Factory Settings Configure Choices Factory Settings Temperature compensation Section Menu Tree for the 54eA Controller Main Menu Main Alarm 2 setpoint Alarm setpointsRanging the Outputs MA 0.00 ppm 20 mA 20.00 ppm Output 1 12.00 mASetpoint 1.000 ppm 4mA 0.000 ppm 20mA 10.00 ppm Alarm setpointsOutput setpoints Simulated testSimulated tests Test alarm 1 OpenTest output Display ConfigureTemp units C Output 1 mA Output 2 mA Display contrast Timeout OnLanguage English Display left Display right Out Output 1 control OutputsOutput Measurement Range 4-20 mA Dampen 0 sec Hold Last Value Ctrl mode NormalOutput 1 Setup Output 1 ControlUsing hold Hold setupAlarm Control Settings Changing Alarm ParametersAlways configure the control parameters Before making Changes in the alarm setupAlarm Low Setpoint 0.000 ppm Alarm 1 controlAlarm 1 setup Alarm 2 controlHigh Alarm Low AlarmFeed limit timer setup Alarm FaultFeed limit Disable Timeout 3600 secInterval timer setup Alarm 4 setup Feed limit timerInterval timer Timer Disable Timer Time activated Interval 24.0 hrPH comp Manual PH value 7.00 pH PH sensor EnablePH sensor Disable Using the 499ACL-01 sensorGlass imp lo 20 MΩ Diagnostics OffAutocal Standard Operate iso 7.00 pH Sensor iso 7.00 pH Temp coeffTemp comp Manual Temp units C Temperature Compensation and Temperature UnitsTemp comp Auto Temp units CNoise Reduction 60 HzMain sensor cal Bar units mm Hg Bar meas AutoBarometric Pressure Lock all Lock program Lock configNoise rejection Main sensor cal SecurityCondition Definitions Controller Mode PriorityAction Definitions Introduction Section Calibration TemperatureTemperature Calibration Adjust temp + 025.1 CDissolved Oxygen Concentration Section Calibration Dissolved OxygenZeroing the Sensor Sensor must be in zero solutionCalibrating the Sensor in AIR Stabilizing . . . WaitSection Calibrating the Sensor Against a Standard Instrument Calibrate 8.32 ppmCalibrating Barometric Pressure Zero main sensor Adjust temperatureCalibrate main sensor StandardizeSensor Current as a Function of Free Chlorine Concentration Section Calibration Free Chlorine 499ACL-01Calibrate main sensor Full Scale Calibration Calibrate 1.100 ppmDual Slope Calibration Dual range calOutput trim Calibrate low pointCalibrate high point Low point 2.000 ppmIntroduction Section Calibration Free Chlorine 498CL-01Ppm 100ppm 000 ppm Section Calibration Free Chlorine 498CL-01 Section Calibration Total Chlorine This page intentionally left blank Sensor Current as a Function of Monochloramine Concentration Section Calibration MonochloraminePpm 100ppm Sensor Current as a Function of Ozone Concentration Section Calibration OzonePpm 100ppm PH at 25C Standards Nominal pH SectionAutomatic TWO-POINT Calibration Autocal buffer10.02 pH Pt 04.00pH 25C Calibrate pointManual TWO-POINT Calibration Pt 4.00pH 25.0C 10.01 pHPt calibration Standardize pH pH slope Standardization ONE-POINT CalibrationPH slope 60.00 mV/pH Trimming the Outputs Section Calibration Current OutputsPID Control Code Section PID and TPC ControlControl Loop Adjustment and Tuning Proportional Gain Plus Integral ResetProcess Reaction Curve Method Process Reaction Curve Section Time Proportional Control TPC Mode Code Overview Section TroubleshootingTroubleshooting When a Fault Message is Showing Fault message Explanation See SectionPH low or high input voltage Temperature error low or highHigh input current Check sensor zeroFailure factory and Failure eeprom Troubleshooting When no Fault Message is Showing OxygenProblem See Section Sense line openZero current is too high Possible error warning during in-process calibrationZero reading Is unstable Sensor does not respond to changes in oxygen level Barometric pressure reading is too high or too lowProcess readings are erratic Readings driftZero current is unstable ProblemSensor can be calibrated, but the current is too low Chlorine readings are too low Sensor does not respond to changes in chlorine levelProblem See Section Sensor does not respond to changes in monochloramine level Readings are too low Troubleshooting When no Fault Message is Showing Ozone Ozone readings are too low Sensor does not respond to changes in ozone levelTroubleshooting When no Fault Message is Showing pH Sensor does not respond to known pH changes Controller will not accept manual slope100 101 Process pH readings are noisySensor Polarizing Voltage Resistance Expected current Troubleshooting not Related to Measurement ProblemsProblem Action Simulating Inputs Dissolved OxygenSimulate Chlorine and Ozone 103 Simulating Inputs Other Amperometric MeasurementsVoltage mV PH at 25C Simulating pH input when the preamplifier is in the sensor104 Simulating temperature Simulating Temperature105 106 Measuring Reference VoltageReplacement Parts Part Number Description Section MaintenanceGeneral Section Return of MaterialWarranty Specifications subject to change without notice