MODEL 54eA | SECTION 13.0 |
| CALIBRATION - pH |
SECTION 13.0
CALIBRATION - pH
13.1 INTRODUCTION
A new pH sensor must be calibrated before use. Regular recalibration is also necessary.
A pH measurement cell (pH sensor and the solution to be measured) can be pictured as a battery with an extreme- ly high internal resistance. The voltage of the battery depends on the pH of the solution. The pH meter, which is basically a voltmeter with a very high input impedance, measures the cell voltage and calculates pH using a con- version factor. The actual value of the
In automatic calibration the controller recognizes the buffer and uses
| pH at 25°C |
| Standard(s) |
|
|
| Note 1: NIST is National Institute of Standards, | |
| (nominal pH) |
|
|
|
|
| DIN is Deutsche Institute für Normung, JSI is | |
| 1.68 |
| NIST, DIN 19266, JSI 8802, BSI (see note 1) |
| Japan Standards Institute, and BSI is British | |||
| Standards Institute. | |||||||
| 3.56 |
| NIST, BSI |
|
|
| ||
|
|
|
| Note 2: pH 7 buffer is not a standard buffer. It is | ||||
| 3.78 |
| NIST |
|
|
| ||
|
|
|
| a popular commercial buffer in the United | ||||
| 4.01 |
| NIST, DIN 19266, JSI 8802, BSI |
|
|
| States. | |
|
|
|
|
|
| |||
| 6.86 |
| NIST, DIN 19266, JSI 8802, BSI |
|
|
|
|
|
| 7.00 |
| (see note 2) |
|
|
|
|
|
| 7.41 |
| NIST |
|
|
|
|
|
| 9.18 |
| NIST, DIN 19266, JSI 8802, BSI |
|
|
|
|
|
| 10.01 |
| NIST, JSI 8802, BSI |
|
|
|
|
|
12.45 |
| NIST, DIN 19266 |
|
|
|
|
| |
During automatic calibration, the controller also measures |
|
|
|
|
| |||
|
|
|
|
| ||||
noise and drift and does not accept calibration data until |
|
|
|
|
| |||
readings are stable. Calibration data will be accepted as soon |
|
|
|
|
| |||
as the pH reading is constant to within the |
|
|
|
|
| |||
0.02 pH units for 10 seconds. The stability settings can be |
|
|
|
|
| |||
changed. See Section 5.11. |
|
|
|
|
| |||
In manual calibration, the controller still monitors readings for |
|
|
|
|
| |||
stability; however, the buffer pH lookup feature is missing. |
|
|
|
|
| |||
The user has to enter the correct pH at the temperature the |
|
|
|
|
| |||
buffer is being used. |
|
|
|
|
|
| ||
Once the controller completes the calibration, it calculates |
|
|
|
|
| |||
the calibration slope and offset. The slope is reported as the |
|
|
|
|
| |||
slope at 25°C. Figure |
|
|
|
|
| |||
The controller can also be standardized. Standardization is |
|
|
|
|
| |||
| FIGURE |
| ||||||
the process of forcing the controller reading to match the |
|
| ||||||
|
|
|
|
| ||||
|
|
|
|
| ||||
|
|
|
|
| ||||
reading from a second pH instrument. Standardization is |
|
|
|
|
| |||
sometimes called a |
|
|
|
|
|
73