Emerson Process Management 53eA instruction manual Main sensor cal

Page 49

MODEL 54eA

SECTION 5.0

 

SOFTWARE CONFIGURATION

5.11MAIN SENSOR CALIBRATION PARAMETERS

1.Main sensor refers to the amperometric sensor.

Noise rejection

Main sensor cal

Security

ExitEnter

Stabilize : 0.050 ppm

Stabilize time: 10 sec

ExitEdit

Stabilize: 0.050 ppm

Stabilize time: 10 sec

Dual range cal : Disable

ExitEdit

Stabilize: 0.050 ppm

Stabilize time: 10 sec

Salinity : 0.0 o/oo

ExitEdit

2.Press any key to enter the main menu. Move the cursor to "Program" and press Enter (F4).

3.Move the cursor to "Configure" and press Enter (F4).

4.Move the cursor to "Main sensor cal" and press Enter (F4).

5.Use the é and ê keys to move through the list of items. To make a change press Edit (F4). Use the arrow keys to change settings to the desired value and press Save (F4). For allowed ranges, see Table 5-1.

The choices depend on the measurement being made.

Stabilize and Stabilize time: For the controller to accept calibration data, the concentration must remain within a specified range for a specified period of time. The default values are 0.05 ppm (50 ppb) and 10 seconds. Using a small stabilize value and a long stabilize time is the best protection against calibration while a reading is still changing.

Dual range cal: Free and total chlorine sensors from Rosemount Analytical (Model 499ACL-01 and 499ACL-02, respectively) become non-linear at high concentrations of chlorine. Dual range calibration allows the analyzer to correct for the non-linearity of the sensor. For more information see Sections 8.0 and 9.0.

Salinity: Salinity is used with oxygen measurements only. The solu- bility of oxygen in water depends on the concentration of dissolved salts in the water. Increasing the concentration decreases the solu- bility. If the salt concentration is greater than about 1000 ppm, the

accuracy of the measurement can be improved by applying a salini- ty correction. Enter the salinity as parts per thousand (o/oo). One per- cent is ten parts per thousand.

43

Image 49
Contents Amperometric Hart Analyzer/Controller Model 54eAEssential Instructions Page Model 54eA ANALYZER/CONTROLLER 11.0 10.012.0 13.0Section Title List of FiguresIii Features and Applications Section SpecificationsPower Specifications GeneralOutput Accuracy ± 0.05 mA Alarms Weight/Shipping Weight 5 lb/6 lb 2 kg/2.5 kgSpecifications Free Chlorine Specifications OxygenSpecifications Total Chlorine Specifications OzoneOrdering Information 115/230 VAC, 50/60 Hz Power24 VDC Installation Section InstallationUnpacking and Inspection Pipe mounting Pipe and Wall Mounting DimensionsPanel mounting Panel Mounting DimensionsPOWER, ALARM, and Output Wiring Section WiringGeneral 454EPH02 Wiring Model 499A oxygen, chlorine, and ozone sensors Sensor WiringGeneral 4054eA01 4054eA03 4054eA02 4054eA04 4054eA07 Section Display and Operation Alarm StatusGeneral Description DisplayProgram Settings List Section Software ConfigurationChoices Factory Settings Configure Choices Factory Settings Temperature compensation Section Menu Tree for the 54eA Controller Main Menu Main Alarm 2 setpoint Alarm setpointsRanging the Outputs MA 0.00 ppm 20 mA 20.00 ppm Output 1 12.00 mAOutput setpoints Alarm setpointsSimulated test Setpoint 1.000 ppm 4mA 0.000 ppm 20mA 10.00 ppmSimulated tests Test alarm 1 OpenTest output Display ConfigureTemp units C Output 1 mA Output 2 mA Display contrast Timeout OnLanguage English Display left Display right Out Output 1 control OutputsOutput Measurement Output 1 Setup Ctrl mode NormalOutput 1 Control Range 4-20 mA Dampen 0 sec Hold Last ValueUsing hold Hold setupAlways configure the control parameters Before making Changing Alarm ParametersChanges in the alarm setup Alarm Control SettingsAlarm 1 setup Alarm 1 controlAlarm 2 control Alarm Low Setpoint 0.000 ppmHigh Alarm Low AlarmFeed limit Disable Alarm FaultTimeout 3600 sec Feed limit timer setupInterval timer Alarm 4 setup Feed limit timerTimer Disable Timer Time activated Interval 24.0 hr Interval timer setupPH sensor Disable PH sensor EnableUsing the 499ACL-01 sensor PH comp Manual PH value 7.00 pHGlass imp lo 20 MΩ Diagnostics OffAutocal Standard Operate iso 7.00 pH Sensor iso 7.00 pH Temp coeffTemp comp Auto Temperature Compensation and Temperature UnitsTemp units C Temp comp Manual Temp units CNoise Reduction 60 HzMain sensor cal Bar units mm Hg Bar meas AutoBarometric Pressure Noise rejection Main sensor cal Lock program Lock configSecurity Lock allCondition Definitions Controller Mode PriorityAction Definitions Introduction Section Calibration TemperatureTemperature Calibration Adjust temp + 025.1 CDissolved Oxygen Concentration Section Calibration Dissolved OxygenZeroing the Sensor Sensor must be in zero solutionCalibrating the Sensor in AIR Stabilizing . . . WaitSection Calibrating the Sensor Against a Standard Instrument Calibrate 8.32 ppmCalibrate main sensor Zero main sensor Adjust temperatureStandardize Calibrating Barometric PressureSensor Current as a Function of Free Chlorine Concentration Section Calibration Free Chlorine 499ACL-01Calibrate main sensor Full Scale Calibration Calibrate 1.100 ppmOutput trim Dual range calCalibrate low point Dual Slope CalibrationCalibrate high point Low point 2.000 ppmIntroduction Section Calibration Free Chlorine 498CL-01Ppm 100ppm 000 ppm Section Calibration Free Chlorine 498CL-01 Section Calibration Total Chlorine This page intentionally left blank Sensor Current as a Function of Monochloramine Concentration Section Calibration MonochloraminePpm 100ppm Sensor Current as a Function of Ozone Concentration Section Calibration OzonePpm 100ppm PH at 25C Standards Nominal pH SectionAutomatic TWO-POINT Calibration Autocal buffer10.02 pH Pt 04.00pH 25C Calibrate pointManual TWO-POINT Calibration Pt 4.00pH 25.0C 10.01 pHPt calibration Standardize pH pH slope Standardization ONE-POINT CalibrationPH slope 60.00 mV/pH Trimming the Outputs Section Calibration Current OutputsPID Control Code Section PID and TPC ControlControl Loop Adjustment and Tuning Proportional Gain Plus Integral ResetProcess Reaction Curve Method Process Reaction Curve Section Time Proportional Control TPC Mode Code Troubleshooting When a Fault Message is Showing Section TroubleshootingFault message Explanation See Section OverviewHigh input current Temperature error low or highCheck sensor zero PH low or high input voltageProblem See Section Troubleshooting When no Fault Message is Showing OxygenSense line open Failure factory and Failure eepromZero current is too high Possible error warning during in-process calibrationZero reading Is unstable Process readings are erratic Barometric pressure reading is too high or too lowReadings drift Sensor does not respond to changes in oxygen levelZero current is unstable ProblemSensor can be calibrated, but the current is too low Chlorine readings are too low Sensor does not respond to changes in chlorine levelProblem See Section Sensor does not respond to changes in monochloramine level Readings are too low Troubleshooting When no Fault Message is Showing Ozone Ozone readings are too low Sensor does not respond to changes in ozone levelTroubleshooting When no Fault Message is Showing pH Sensor does not respond to known pH changes Controller will not accept manual slope100 101 Process pH readings are noisyProblem Action Troubleshooting not Related to Measurement ProblemsSimulating Inputs Dissolved Oxygen Sensor Polarizing Voltage Resistance Expected currentSimulate Chlorine and Ozone 103 Simulating Inputs Other Amperometric MeasurementsVoltage mV PH at 25C Simulating pH input when the preamplifier is in the sensor104 Simulating temperature Simulating Temperature105 106 Measuring Reference VoltageReplacement Parts Part Number Description Section MaintenanceGeneral Section Return of MaterialWarranty Specifications subject to change without notice