MODEL 54eA | SECTION 16.0 |
| TROUBLESHOOTING |
16.5.6 Sensor does not respond to changes in chlorine level.
A.Is the grab sample test accurate? Is the grab sample representative of the sample flowing to the sensor?
B.Is the pH compensation correct? (Applies only to the
C.Is the membrane clean? Clean the membrane and replace it if necessary. Check that the holes at the base of the cathode stem are open. Use a straightened paper clip to clear blockages. Replace the electrolyte solution.
D.Replace the sensor.
16.5.7 Chlorine readings spike following sudden changes in pH
Changes in pH alter the relative amounts of hypochlorous acid (HOCl) and hypochlorite ion
16.5.8 Chlorine readings are too low.
A.Was the sample tested as soon as it was taken? Chlorine solutions are unstable. Test the sample immediate- ly after collecting it. Avoid exposing the sample to sunlight.
B.Low readings can be caused by zeroing the sensor before the residual current has reached a stable minimum value. Residual current is the current the sensor generates even when no chlorine is in the sample. Because the residual current is subtracted from subsequent measured currents, zeroing before the current is a mini- mum can lead to low results.
Example: The true residual current for a free chlorine sensor is 4 nA, and the sensitivity is 350 nA/ppm. Assume the measured current is 200 nA. The true concentration is
C.Sensor response depends on flow. If the flow is too low, readings will be low and flow sensitive. Verify that the flow past the sensor equals or exceeds the minimum value. See the sensor instruction manual for recom- mended flows.
16.6 TROUBLESHOOTING WHEN NO FAULT MESSAGE IS SHOWING - TOTAL CHLORINE
Refer to the instruction manual for the TCL for a complete troubleshooting guide.
93