Operating the Pump Motor in the STOPPED Display Mode

5

 

 

Since tubes of different materials, sizes, ages, and durometers produce different flow rates, the CALIBRATE key allows you to calibrate each piece of tubing you use. Calibration accuracy data is shown in Table 9-1 in Chapter 9. Once the tubing is calibrated, you no longer need to be concerned about flow rate variances outlines in Table 9-3.

NOTE

Along with the usual pumping equipment (pump(s), pump motor), you will need calibrated beakers or cylinders to function as source and drain containers.

1.Before you initiate the calibration process, you must set the pump up for pumping in the same direction as you plan to use during normal pump operation (refer to Section 5.2). In addition, make sure the display shows STOPPED Display Mode (the motor is not running).

2.Set either the RPM or FLOW RATE setting of the pump motor:

2a. To set the RPM, press the RPM key. The display flashes the current RPM setting (xxx rpm). Use the UP and DOWN ARROW keys to adjust the setting.

2b. To set the FLOW RATE, press the TUBE ID key and use the UP and DOWN ARROW keys to select the inner diameter of the tube you are using. Press the FLOW RATE key. Use the UP and DOWN ARROW keys to adjust the setting.

3.When the display shows the RPM or FLOW RATE setting you desire, you are ready to initiate the calibration process.

4.Press the CALIBRATE key. The display should flash and show, “Prime or START ?”.

5.To prime the tube, press and hold down the PRIME key. As long as you hold down the PRIME key, fluid flows in the tube, and the display shows, “Priming...”.

6.When fluid has completely filled the tube, simply release the PRIME key. The pump motor stops running, and fluid stops flowing.

7.To dispense fluid for calibration, hold the tubing outlet over the drain container and press the START/STOP key. Fluid flows, and the display enters the Running Display Mode, showing the message, “Dispensing...”.

8.When you have allowed fluid to dispense for a minute or two, press the START/STOP key to stop the pump. Fluid stops dispensing, and the display returns to the STOPPED Display Mode. The display momentarily flashes the message, “Enter TOTAL VOL.” and then flashes the total volume dispensed (xxx ml or xxx Liters) alternating with the message “Start or CAL ?”.

9.To continue to dispense fluid, press the START/STOP key again. Fluid flows, and the display returns to the Running Display Mode, showing the message, “Dispensing...”.

5-7

Image 34
Contents User’s Guide Servicing North America Unpacking Instructions Table of Peristaltic Pump Motor Chapter Specifications Spare Parts and Accessories 10-1 Peristaltic Pump MotorPeristaltic Pump Motor Index RPMIndex III Introduction FeaturesDescription Introduction Front View Parts of the Pump MotorFront View Description FunctionLeft Side View Left Side ViewRight Side View Right Side ViewMembrane Keypad and Display Membrane Keypad DetailMembrane Keypad Details cont’d Flow Rate keyUP and Down Arrow Keys Scrolling Limits Rear ViewRear View Rear View cont’d Language ButtonParts of the Pump Motor Attaching the Pump Pump Motor Mounting Configurations Number of Pumps onGetting Started IntroductionAttaching a Single Pump Directly to the Motor Refer to Figures 3-1 throughStacking Two Pumps on One Side of the Motor Placing the Second Pump on Top of the First PumpTubing Life Selecting TubingLoading the Tubing See .2 to set the pumping direction Getting Started Stopped Display Mode Turning on the PowerUsing the ON/OFF Switch Introduction to the Two Display ModesRunning Display Mode Stopped vs. Running Display ModesOperating the Pump Motor Stopped Display ModeDisplaying Pump Messages in Five Different Languages English French Spanish German Italian Régler LA QTÉDefining the Direction of Pumping Selecting the Tubing Inner Diameter Left Pump Right PumpTube Inner Diameters English Metric Priming the Tubing Calibrating the TubingOperating the Pump Motor in the Stopped Display Mode Tips for Optimum Calibration Accuracy Measuring the Temperature of a Fluid Preparing for Pumping in a Specific ModeTotal Volume RPM Mode Running Display ModePumping in the RPM Mode Shows the range of the RPM numbers you can work with RPM Mode Cont’d Flow Rate Mode Pumping in the Flow Rate ModeShows the ranges allowed for the various tubing sizes Flow Rate Ranges cont’d Display Xxx mL/min Flow Rate Mode Flow Rate Mode Time Mode Pumping in the Time ModeTime Range Display Xxxxxx Time Mode Time Mode Total Volume Mode Block Diagram Pumping in the Total Volume ModePumping in the Total Volume Mode Total Volume RangeOperating the Pump Motor in the Running Display Mode Total Volume Mode Cont’d Other Pumping Options Preventing Tubing from Moving During Pump OperationAuxiliary Out Remote Monitor 500Ω LED Auxiliary In Remote Start/StopError Message Problem Troubleshooting GuideDisplay Error Messages SolutionError Message Running Display Mode and try to select the command againLocked Rotor? Overcurrent Press the Total Volume key Short Circuit? Error Message Other Pumping Problems Stopped Display Mode and refer to SectionProblem Solution Another pump, if necessary Troubleshooting Guide Technical Details Design SolutionMaintenance Theory of OperationTechnical Details Specifications Calibration AccuracyTubing Materials Average Flow RatesMotor SpeedAverage Tubing Life ⁄32 300 ⁄16 250 100 ML Per Revolution Specifications Spare Parts Spare Parts and AccessoriesAccessories Part Number DescriptionAccessories cont’d Peristatic Pump 10-4 WARRANTY/DISCLAIMER