Troubleshooting Guide | 7 |
|
|
Error Message | Problem | Solution |
OVERCURRENT!
(continued)
6.Make sure the tubing is not caught on the bottom roller. Reposition the roller slightly and reload the tubing, if necessary.
7.Make sure no part of the tubing is being constricted by a clamp. Loosen clamps if necessary.
8.Turn the unit’s power off and on, adjust the settings, and attempt to resume pumping.
Select a MODE! | You have attempted to start the |
| |
| pump, either while the display |
| shows the TUBE ID setting, or |
| after you had entered the |
| volume dispensed in the |
| calibration process. |
1.Select a pumping mode (RPM, FLOW RATE, TIME, or TOTAL VOLUME).
2.Press the key that corresponds to that mode.
3.Use the UP and DOWN ARROW keys to adjust the setting.
4.Press the START/STOP key to start the pump.
Set RPM or RATE! | You have selected a command |
| |
| that requires an RPM or FLOW |
| RATE setting to function properly. |
1.Press the RPM or FLOW RATE key.
2.Use the UP and DOWN ARROW keys to adjust the setting.
3.
SET TOTAL VOLUME | You have attempted to start the |
| |
| pump in the TOTAL VOLUME |
| Mode with the TOTAL VOLUME |
| set at 0 mL. |
1.Press the TOTAL VOLUME key
2.Use the UP and DOWN ARROW keys to adjust the TOTAL VOLUME setting.
3.Press the START/STOP key to start the pump.