7Troubleshooting Guide
Error Message | Problem | Solution |
Must STOP first! | You have attempted to calibrate |
| or change the direction of |
| pumping while the pump |
| is running. |
1.Press the START/STOP key to stop the pump. Verify that the display is in the STOPPED Display Mode.
2.For information on how to change the direction of pumping, refer to Section 5.2.
not applicable | You have attempted to perform | Review the instructions for | |
| a function that you can not | performing the required function | |
| perform at this time. | and proceed. | |
|
|
| |
not available | You have attempted to perform | Review the instructions for | |
| a function that you can not | performing the required function | |
| perform at this time. | and proceed. | |
|
|
|
|
OVERCURRENT! | The motor driver board has | 1. | The tubing may be causing a |
| sensed an overcurrent. |
| jam. Observe how the tubing |
| To recover from an overcurrent, |
| is loaded in the pump(s) and, |
|
| if necessary, reload it. | |
| first turn the power switch off, | 2. | Detach the pump(s) and check |
| and then back on. Then | ||
| attempt solutions 1 through 8. |
| for a mechanical jam. |
|
|
| Substitute another pump, if |
|
|
| necessary. |
|
| 3. | Check to see if the fuse is |
|
|
| blown. Replace if necessary. |
|
|
| Refer to Chapter 9 for fuse rat- |
|
|
| ings and chapter 8 for instruc- |
|
|
| tions on how to replace it. |
|
| 4. | Make sure you are using tub- |
|
|
| ing of the appropriate wall |
|
|
| thickness for the pump(s). |
|
|
| Refer to Table |
|
|
| |
|
| 5. | Make sure you are using tub- |
|
|
| ing of the appropriate durom- |
|
|
| eter for your pump(s). Try to |
|
|
| operate the motor with tubing |
|
|
| of 68 or smaller durometer. |